Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010231772), received between Nov 13, 2025 and Nov 13, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1

Patient problems

  • Metal Related Pathology1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 13, 2025Degraded; Degraded

    IT WAS REPORTED THAT, FOLLOWING A TOTAL HIP REPLACEMENT (THR) ON (B)(6) 2008, THE PATIENT HAD BEEN EXPERIENCING LEFT HIP PAIN AND WAS SCHEDULED FOR REVISION SURGERY ON (B)(6) 2023, DUE TO A BROKEN TOTAL HIP ARTHROPLASTY WITH SEVERE METALLOSIS. INTRAOPERATIVELY, THE RF INTERFIT ACET NH SZ 50 WAS FOUND TO BE WORN, CAUSING THE OXINIUM FEMORAL HEAD TO ARTICULATE DIRECTLY WITH THE INNER SURFACE OF THE ACETABULAR CUP. EXTENSIVE REMOVAL OF METALLOSIS WAS PERFORMED ON THE HIP, AND BOTH THE FEMORAL HEAD AND ACETABULAR SYSTEM WERE REPLACED WITH A MULTI-HOLE REVISION SYSTEM. THE PATIENT TOLERATED THE PROCEDURE WELL AND LEFT THE OPERATING ROOM FOR RECOVERY.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 23547169 ↗

Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices