ECHELON 71340413Recalls & reported problems
Smith & Nephew, Inc. · ECHELON 260MM POROUS BOWED FEMORAL COMPONENT STANDARD COLLAR SIZE 13 LEFT · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010328984), received between Jun 2, 2017 and Nov 3, 2020 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K211176, K983834); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Fracture2
- Appropriate Device Problem Term/Code Not Available1
Patient problems
- Bone Fracture(s)2
- Failure of Implant2
- Fall2
- Injury2
- Pain2
- Hip Fracture1
- Joint Dislocation1
- Joint Laxity1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryNov 3, 2020Break; Fracture; Break; Fracture
IT WAS REPORTED THAT, AFTER THE PATIENT HAD BEEN IMPLANTED AN ECHELON STEM ON (B)(6) 2014 DURING A REVISION SURGERY, THE PATIENT EXPERIENCED A DISLOCATION BECAUSE THE STEM FRACTURED. THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2020: ONLY THE FEMORAL HEAD, STEM AND ACCORD WIRES WERE EXPLANTED. THE PATIENT HAS HAD COMPLICATIONS SINCE THIS PROCEDURE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Bone Fracture(s); Failure of Implant; Hip Fracture; Joint Dislocation; Joint Laxity; Metal Related Pathology; Pain; Thromboembolism
MAUDE report 10781732 ↗InjuryFeb 11, 2019Fracture; Fracture
IT WAS REPORTED THAT A REVISION SURGERY NEEDS TO BE PERFORMED DUE TO THE FRACTURE OF THE STEM.
Patient problems: Injury
MAUDE report 8327746 ↗InjuryAug 7, 2017Break
IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION SURGERY DUE TO BREAKAGE OF FEMORAL STEM FOLLOWING A FALL. IT WAS REPORTED THAT BEFORE THE FALL PATIENT HAD SEVERE PAIN.
Patient problems: Failure of Implant; Fall; Injury; Pain
MAUDE report 6770295 ↗InjuryJun 2, 2017Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT PATIENT UNDERWENT A REVISION SURGERY WITH EXCHANGE OF FEMORAL STEM. THE PATIENT USED A WALKING STICK TO ASSIST HER, SHE WALKED WITH THE STICK INTO HER LAUNDRY & SLIPPED ON THE WET FLOOR. THIS RESULTED IN A FRACTURE OF THE LEFT FEMUR INCLUDING THE STEM FITTED IN 2000.
Patient problems: Bone Fracture(s); Fall
MAUDE report 6607655 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K211176, K983834. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page