Skip to content

RENOVATION 71369045Recalls & reported problems

Smith & Nephew, Inc. · CEMENT DRILL 4.5MM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010437617), received between Jun 18, 2020 and Apr 28, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Fracture” (5 reports).

1 other recall cover devices cleared under the same FDA submission (K093047); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture5
  • Break4

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Foreign Body In Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A REVISION THR SURGERY, ONE (1) CEMENT DRILL 4.5MM FRACTURED WHILE DRILLING THROUGH THE CEMENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21918491 ↗
  • MalfunctionOct 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR REVISION SURGERY, WHEN REMOVING CEMENT FROM THE FEMORAL CANAL, THE CEMENT DRILL 4.5MM BREAK FROM IMPACT ON THE CEMENT. IT IS UNKNOWN HOW WAS COMPLETED THE SURGERY. NO DELAY WAS REPORTED OR INJURY TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20371172 ↗
  • MalfunctionNov 1, 2023Break; Break

    IT WAS REPORTED THAT DURING A THR SURGERY, ONE (1) CEMENT DRILL 4.5MM WAS FOUND BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18050181 ↗
  • MalfunctionAug 4, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A REVISION SURGERY CONDUCTED TO REMOVE A COMPETITOR'S KNEE PROTHESIS, THE TIP OF A CEMENT DRILL 4.5MM FRACTURED WHILE USING THE DRILL ON THE TIBIA. THE SURGEON REMOVED THE DRILL TIP. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED AND IF THERE WAS A SURGICAL DELAY. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17456500 ↗
  • InjuryAug 29, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR 2 STAGE REVISION (1 STAGE) PROCEDURE, WHILE SURGEON WAS IN THE PROCESS OF REMOVING CEMENT FROM THE FEMORAL CANAL WITH THE CEMENT DRILL 4.5MM FROM THE RENOVATION SET, THE DRILL TIP SNAPPED OFF. THIS CAUSED THE DEVICE BECAME LODGED IN THE FEMORAL CANAL. AN ATTEMPT WAS MADE TO RETRIEVE THE DRILL BIT INITIALLY, THE FLUOROSCOPY SHOWED THE DRILL TIP WEDGED BETWEEN A REVISION KNEE STEM AND THE FEMORAL CORTEX. SURGEON COMMENTED THAT THIS ADVERSE EVENT WOULD BE ADDRESSED DURING THE REVISION OF THE KNEE PROSTHESIS BY REMOVING THE BROKEN DRILL TIP. SURGERY WAS RESUMED, WITH MORE THAN AN ONE HOUR DELAY. IT IS UNKNOWN HOW SURGERY WAS FINISHED. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Foreign Body In Patient

    MAUDE report 15316108 ↗
  • MalfunctionMay 18, 2022Break; Break

    IT WAS REPORTED THAT AFTER THR, THE CEMENT DRILL 4.5MM WAS FOUND TO BE BROKEN. NO CASE INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14432240 ↗
  • MalfunctionMay 18, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING SET UP FOR A THR SURGERY, THE CEMENT DRILL 4.5MM SNAPPED IN HALF. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. NO DELAYS AND NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14432235 ↗
  • MalfunctionDec 14, 2021Break; Break

    IT WAS REPORTED THAT A CEMENT DRILL 4.5MM ((B)(6)) BROKE INSIDE THE PATIENT, A RONGEUR 300MM WITH TEETH ((B)(6)) AND A RONGEUR 200MM WITH TEETH ((B)(6)) ARE BROKEN. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED OR IF THERE WAS ANY DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13002201 ↗
  • InjuryJun 18, 2020Break; Break

    IT WAS REPORTED THAT DURING THR REVISION SURGERY (IT IS UNKNOWN IF WAS A REVISION SURGERY OF SMITH AND NEPHEW DEVICES), WHILE USING THE DRILL TO DRILL THROUGH A CEMENT MANTLE IN THE FEMUR, THE TIP OF THE CEMENT DRILL BROKE OFF. THE PROCEDURE WAS FINISHED WITH A CHANGE IN THE SURGICAL TECHNIQUE. THERE WAS NO SURGICAL DELAY AND NO INJURY TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10166343 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093047. One submission often covers many models, so check the product and lot details in each record.

  • Z-0245-2022 · Jun 17, 2021 · Open, Classified
    Ilizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434

Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices

All APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices →