GENESIS II 71420146Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II OXINIUM FEMORAL SIZE 2 RIGHT · Class 2 · Implant
All models of GENESIS II (462)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010439413), received between Feb 16, 2021 and Mar 3, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Connection Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K021673, K962557); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
Patient problems
- Insufficient Information1
- Loss of Range of Motion1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 3, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, PATIENT HAD A TOTAL KNEE REPLACEMENT 3 YEAR AGO. PATIENT SUFFERED FROM A STIFF KNEE AND LIMITED RANGE OF MOTION 20 DEGREE - 90 DEGREE AND OVERSTUFFED PATELLA FEMORAL JOINT, THEREFORE PATIENT HAD A REVISION SURGERY, WHERE SURGEON REMOVED THE PATELLA, FEMORAL, TIBIAL INSERT AND TIBIAL BASEPLATE FROM PRIMARY SURGERY. SURGEON NOTICED THE PATELLA WAS NOT RESECTED ENOUGH, WHICH CAUSE OVER STUFFING. SURGEON ACHIEVED A RANGE OF MOTION FROM 0 DEGREES - 120 DEGREES RANGE OF MOTION. PATIENT OUTCOME IS UNKNOWN.
Patient problems: Loss of Range of Motion
MAUDE report 11408614 ↗MalfunctionFeb 16, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING SURGERY, THE GENESIS II OXINIUM FEM SIZE 2 RIGHT WAS NOT LOCKING PROPERLY AND WHILE REMOVING THE INSERT THE FEMUR WAS SCRATCHED. IT WAS REPORTED THAT THERE WAS A SIGNIFICANT DELAY IN THE SURGERY. THE PATIENT WAS INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 11330954 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K021673, K962557. One submission often covers many models, so check the product and lot details in each record.
- Z-0281-2025 · Oct 2, 2024 · Open, Classifiedsmith&nephew GENESIS II NONPOROUS POSTERIOR STABILIZED FEMORAL COMPONENT, SIZE 5 LEFT, REF 71421015; Uncoated knee femur prosthesis, metallic
- Z-1901-2024 · Apr 24, 2024 · Open, Classifiedsmith&nephew OXINIUM Femoral Head 12/14 Taper, REF 71343204, 32 MM OD +4, Sterile; and smith&nephew, 12/14 Taper, REF 71343604, 36 MM M / +4
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page