REFLECTION 71333325Recalls & reported problems
Smith & Nephew, Inc. · REFLECTION XLPE 28MM INNER DIAMETER 20 DEGREE 54-56MM OUTER DIAMETER SIZE F · Class 2 · Implant
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010455628), received between Jul 17, 2017 and Dec 3, 2025 — classified as 4 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Connection Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K002747, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Connection Problem2
- Difficult to Open or Close1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- Injury2
- No Known Impact Or Consequence To Patient2
- Bone Fracture(s)1
- Joint Dislocation1
- No Information1
- Periprosthetic fracture1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryDec 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER THR SURGERY HAD BEEN PERFORMED ON (B)(6) 2007, THE PATIENT EXPERIENCED A BONE BREAKAGE AROUND THE STEM. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY PERFORMED ON (B)(6) 2025, IN WHICH A COMPETITOR'S PLATE AND CABLE WARE USED TO FIX THE BREAKAGE. WHITE SMALL MATERIAL WAS FOUND IN THE SURGICAL FIELD. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Periprosthetic fracture
MAUDE report 23703496 ↗InjuryAug 23, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A THR PERFORMED ON (B)(6) 2008, PATIENT SUFFERED A FALL THAT CAUSED PAIN, A BONE FRACTURE AROUND THE STEM AND A DISLOCATION. THIS WAS ADDRESSED VIA REVISION SURGERY ON (B)(6) 2022 WHERE THE REFLECTION XLPE 28MM INNER DIAMETER 20 DEGREE 54-56MM OUTER DIAMETER SIZE F WAS EXPLANTED. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Bone Fracture(s); Joint Dislocation
MAUDE report 15278102 ↗InjuryJun 20, 2019Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING SURGERY THE LINER NOT LOCKED PROPERLY IN TO THE SHELL. THEY USED ANOTHER LINER THAT ALSO NOT LOCKED. THEY REMOVED REFLECTION SHELL 54 AND USED R3 SHELL 56 AND LINER. 30 MINUTES OF DELAY DUE TO THIS INCIDENT. A BACKUP DEVICE WAS AVAILABLE.
Patient problems: Injury
MAUDE report 8720155 ↗MalfunctionOct 3, 2018Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE REFLECTION MICROSTABLE LOCKING MECHANISM LINER FAILED TO LOCK INSIDE WITH THE REFLECTION CUP DURING TOTAL HIP REPLACEMENT PROCEDURE.THE SURGEON FOUND IT DIFFICULT ONCE DISLOCATION WAS DONE WITH THE ORIGINAL POLAR PROSTHESIS.THE SURGEON COULD NOT FIX THE LINER SINCE IT WAS NOT LOCKING PROPERLY. ALMOST ONE HOUR TWENTY MINUTES WAS SPENT ON THE TABLE, IMPLANTS WERE REMOVED FROM THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7931883 ↗MalfunctionSep 10, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT LINER IMPLANT DIDN'T LOCK WITH THE ACETABULAR CUP. THE SURGEON TRIED MANY TIMES TO GET IT SEATED IN THE CUP, WHICH WAS ALREADY IMPLANTED WITH A 6.5MM CANCELLOUS SCREW. THEN HE REMOVED THE SCREW, ACETABULAR CUP AND THE LINER. FINALLY COMPLETED THE PROCEDURE BY FIXING A REFLECTION ALL POLY CEMENTED CUP, BUT SURGERY HAD A DELAY GREATER TAN 30 MINUTES.
Patient problems: Injury
MAUDE report 7862646 ↗MalfunctionAug 27, 2018Difficult to Open or Close
IT WAS REPORTED THAT THE LINER COULD NOT BE LOCKED, FINALLY A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7820720 ↗InjuryJul 17, 2017Connection Problem; Connection Problem
THE LINER FAILED TO LOCK INSIDE. HENCE, THE CUP WAS REMOVED AND REAMED TO THE NEXT SIZE OF THE ACETABULAM. THE NEXT SIZE OF THE CUP WAS OPENED AND THE CORRESPONDING LINER & THEN THE PROCEDURE WAS COMPLETED. SURGERY WAS EXTENDED 30 TO 45 MINUTES.
Patient problems: No Information
MAUDE report 6717018 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K002747, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.