REFLECTION 71332535Recalls & reported problems
Smith & Nephew, Inc. · REFLECTION SPHERICAL HEAD SCREW 35MM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010463289), received between Dec 11, 2017 and May 31, 2022 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K211176, K990666); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Device Dislodged or Dislocated1
- Loose or Intermittent Connection1
- Material Deformation1
Patient problems
- Injury1
- Joint Dislocation1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
- Unspecified Infection1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 31, 2022Material Deformation; Material Deformation
IT WAS REPORTED THAT, DURING A THR SURGERY, THE SURGEON STRIPPED A REFLECTION SPHERICAL HEAD SCREW 35MM ((B)(4)), THIS OCCURRED WHILE USING INSTRUMENTS INSIDE THE PATIENT. THE STRIPPED SCREW WAS LEFT IMPLANTED IN THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING A SMITH AND NEPHEW BACK-UP DEVICE. THE PATIENT WAS NOT INJURED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14549731 ↗InjuryAug 17, 2018Device Dislodged or Dislocated
IT WAS REPORTED THAT PATIENT COMPLAINING OF A PAINFUL CLUNK WHEN INTERNALLY AND EXTERNALLY ROTATING LEG IN FLEXION. SURGEON BELIEVES CUP NOT MEDIALISED ENOUGH, IS PROUD AND A SCREW IS PERHAPS IRRITATING SOFT TISSUE. THE CUP ALSO RETROVERTED. REVISION SURGERY. - RANGE OF MOTION WAS TESTED WHILE PATIENT ASLEEP AND SURGEON REPORTED STABLE -THROUGH ORIGINAL INCISION -OPENED JOINT TO VIEW IMPLANTS. - SURGEON FELT ONCE OPEN THAT JOINT WAS DISLOCATING TOO EARLY WITH FLEXION AND INTERNAL ROTATION. HEAD REMOVED, LINER REMOVED, TWO SCREWS REMOVED AND THE CUP EXPLANTED. SURGEON REAMED DEEPER WITH A 58MM REAMER, TRIALLED A 58MM AND THEN INSERTED A 58MM THREE HOLE R3 CUP. A 40MM LATERALISED LINER 20 DEGREE HOOD WAS INSERTED AFTER TRIALLING AND 40MM PLUS 8MM HEAD INSERTED. SURGEON FELT THAT THIS COMBINATION GAVE BEST RANGE OF MOTION AND STABILITY.
Patient problems: Pain
MAUDE report 7795177 ↗InjuryFeb 10, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
PATIENT POSITIONED AND HIP DISLOCATED. LINER AND SCREWS REMOVED CUP WAS LOOSE. REMOVED CUP REAMED ACETABULUM AND IMPLANTED A 52MM MULTIHOLE CUP AND THREE SCREWS. A CONSTRAINED LINER INSERTED AND CHECKED TO BE LOCKED IN THE CUP. HIP REDUCED USING A 22MM STRYKER HEAD. HIP STABLE.
Patient problems: Joint Dislocation
MAUDE report 7262493 ↗InjuryDec 21, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO UKNOWN REASONS.
Patient problems: Injury
MAUDE report 7140050 ↗InjuryDec 11, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED TO REMOVE ALL THE PARTS DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 7105142 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K211176, K990666. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
- Z-2297-2016 · Jul 6, 2016 · TerminatedSPHERICAL HEAD REFLECTION(R) 6.5 MM CANCELLOUS SCREW, REF 71332525, QTY: (1), STERILE R, smith&nephew Total hip components are indicated for individuals undergoing primary or revis
Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.