Skip to content

GENESIS II 71440145Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II POSTERIOR STABILIZED CONSTRAINED HOUSING REAMER DOME · Class 2

All models of GENESIS II (462)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Smith & Nephew Genesis II posterior stabilized constrained housing reamer dome is a surgical instrument used to prepare the knee during total knee replacement. The 45 reports mostly describe the reamer dome becoming dull or snapping and breaking during surgery, and many were found broken during field inspections with no patient involved. Reports mention 3 injuries and no deaths, and no recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
45
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 03596010471338), received between May 18, 2020 and Apr 10, 2024 — classified as 3 injuries, 42 malfunctions.

The most frequently reported specific device problem is “Dull, Blunt” (19 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Dull, Blunt19
  • Break13
  • Material Deformation5
  • Defective Device4
  • Naturally Worn3
  • Degraded2
  • Fracture2
  • Use of Device Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions20
  • No Patient Involvement17
  • No Known Impact Or Consequence To Patient8
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE SURGEON WAS REAMING THE PS BOX ON A GENESIS II KNEE WHEN THE GII P/S CONS HOUSING REAM DOME SNAPPED. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH THE SAME REPORTED DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19076962 ↗
  • MalfunctionNov 7, 2023Fracture; Difficult or Delayed Separation; Fracture; Difficult or Delayed Separation

    IT WAS REPORTED THAT, DURING A TKA, BOTH GII P/S CONS HOUSING REAM DOME AND A GII PATELLAR REAMER SHAFT GOT STUCK TOGETHER, AS CONSEQUENCE THE GII P/S CONS HOUSING REAM DOME BROKE OFF. NO PIECES FELL INSIDE THE PATIENT. SURGERY WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, IMPLANTING A CR (CRUCIATE RETAINING) GEN II SYSTEM. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18083290 ↗
  • MalfunctionJul 28, 2022Break; Break

    IT WAS REPORTED THAT, ONE GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15126500 ↗
  • MalfunctionMay 23, 2022Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478787 ↗
  • MalfunctionMay 23, 2022Break; Break

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14478752 ↗
  • MalfunctionMay 23, 2022Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT THE GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. SINCE THIS WAS NOTICED DURING A FIELD INSPECTION, NO CASE WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14470584 ↗
  • MalfunctionMay 23, 2022Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14470292 ↗
  • MalfunctionMay 22, 2022Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14467111 ↗
  • MalfunctionMay 22, 2022Break; Break

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14467110 ↗
  • MalfunctionMay 22, 2022Break; Break

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14467108 ↗
  • MalfunctionMay 22, 2022Break; Break

    IT WAS REPORTED THAT, A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14467107 ↗
  • MalfunctionMay 22, 2022Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, DURING FIELD INSPECTION, THE GII P/S CONS HOUSING REAM DOME WAS FOUND TO BE BROKEN. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14467106 ↗
  • MalfunctionMay 20, 2022Detachment of Device or Device Component; Premature Separation; Detachment of Device or Device Component; Premature Separation

    IT WAS REPORTED THAT, DURING SET UP OR INSPECTION, A GII P/S CONS HOUSING REAM DOME WAS MISSING THE METAL PEGS TO LOCK ONTO REAMER, AND THE SHAFT WORN DOWN THE REAMER, THEREFORE, THE SHAFT DISENGAGED. SURGERY WAS NOT DELAYED. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14455085 ↗
  • MalfunctionApr 27, 2022Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE MIDDLE PIECE OF A GII P/S CONS HOUSING REAM DOME WAS FOUND BROKEN WHEN THE INSTRUMENT WAS REMOVED FROM THE TRAY. SURGERY WAS FINISHED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14225763 ↗
  • MalfunctionDec 29, 2021Break; Break

    IT WAS REPORTED THAT, THE GII P/S CONS HOUSING REAM DOME WAS BROKEN. IT IS UNKNOWN WHETHER THE EVENT HAPPENED DURING SURGERY AND IF THERE WAS PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 13110340 ↗
  • MalfunctionApr 23, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A GII P/S CONS HOUSING REAM DOME DID NOT STAY CONNECTED. IT IS UNKNOWN HOW SURGERY WAS COMPLETED AND IF IT WAS DELAYED, THEREFORE, PATIENT INVOLVEMENT IS NOT CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11716329 ↗
  • MalfunctionMar 24, 2021Break; Break

    IT WAS REPORTED THAT, BEFORE A TKA PROCEDURE, A REAMER HAD A CRACK IN TOP OF IT AND NEEDS REPLACED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH+NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11566386 ↗
  • InjuryFeb 22, 2021Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT DURING TKA PROCEDURE THREE GII P/S CONS HOUSING REAM DOMES WERE FOUND WORN AND NOT CUTTING THROUGH BONE ANYMORE. THERE WAS A CHANGE IN TECHNIQUE DUE TO THIS MALFUNCTION. A DELAY OF LESS THAN 30 MIN WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11357025 ↗
  • InjuryFeb 22, 2021Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT DURING TKA PROCEDURE THREE GII P/S CONS HOUSING REAM DOMES WERE FOUND WORN AND NOT CUTTING THROUGH BONE ANYMORE. THERE WAS A CHANGE IN TECHNIQUE DUE TO THIS MALFUNCTION. A DELAY OF LESS THAN 30 MIN WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11357022 ↗
  • InjuryFeb 22, 2021Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT DURING TKA PROCEDURE THREE GII P/S CONS HOUSING REAM DOMES WERE FOUND WORN AND NOT CUTTING THROUGH BONE ANYMORE. THERE WAS A CHANGE IN TECHNIQUE DUE TO THIS MALFUNCTION. A DELAY OF LESS THAN 30 MIN WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11357016 ↗
  • MalfunctionJul 19, 2020Degraded; Degraded

    IT WAS REPORTED THAT THE REAMER DOME WAS FOUND WORN OUT. THERE WAS NO DELAY NOR INJURY REPORTED. THE PROCEDURE WAS FINISHED WITH THE SAME DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10293709 ↗
  • MalfunctionJul 15, 2020Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT DURING TKA PROCEDURE WITH INSTRUMENT INSIDE THE PATIENT, THE GII P/S CONS HOUSING REAM DOME WAS FOUND TO BE WORN OUT. NO DELAY. PROCEDURE WAS FINISHED USING A S&N BACK UP DEVICE. NO INJURIES REPORTED AT THIS TIME. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10274513 ↗
  • MalfunctionJul 9, 2020Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE REAMER HAS BECOME DULL DUE TO WEAR AND TEAR OF USE. NO CASE INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 10254802 ↗
  • MalfunctionJul 9, 2020Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE REAMER HAS BECOME DULL DUE TO WEAR AND TEAR OF USE. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10254708 ↗
  • MalfunctionJul 8, 2020Dull, Blunt; Dull, Blunt

    IT WAS REPORTED THAT THE REAMER HAS BECOME DULL DUE TO WEAR AND TEAR OF USE. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Patient Involvement

    MAUDE report 10248513 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page