NA 71303612Recalls & reported problems
Smith & Nephew, Inc. · COBALT CHROME 12/14 TAPER FEMORAL HEAD 36MM +12 · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010480019), received between May 10, 2019 and Nov 15, 2022 — classified as 2 injuries.
The most frequently reported specific device problem is “Biocompatibility” (1 report).
1 other recall cover devices cleared under the same FDA submission (K022902, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Biocompatibility1
- Insufficient Device Problem Information1
Patient problems
- Foreign Body Reaction1
- Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryNov 15, 2022Biocompatibility; Biocompatibility
IT WAS REPORTED THAT, AFTER A RIGHT THR INDEX SURGERY PERFORMED ON (B)(6) 2015 TO ADDRESS OSTEOARTHRITIS SYMPTOMS, THE PATIENT EXPERIENCED AN UNSPECIFIED ADVERSE SOFT TISSUE REACTION TO PARTICULATE DEBRIS THAT MADE NECESSARY A REVISION SURGERY ON (B)(6) 2021. DURING THIS PROCEDURE, THE FEMORAL HEAD WAS EXPLANTED AND REPLACED WITH A SMITH AND NEPHEW COMPONENT. THIS INFORMATION WAS PROVIDED BY THE NATIONAL JOINT REGISTRY OF THE UNITED KINGDOM, AS PART OF A RETROSPECTIVE DATA COLLECTION OF PATIENTS WHO UNDERWENT A PRIMARY THR SURGERY WITH A HIP PROSTHESIS CONSTRUCT THAT INCLUDED A POLARSTEM AND THAT REQUIRED A REVISION SURGERY DUE TO SPECIFIC REASONS. AS SUCH, NO FURTHER INFORMATION WILL BE AVAILABLE.
Patient problems: Foreign Body Reaction
MAUDE report 15797679 ↗InjuryMay 10, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT WENT TO THE DOCTOR DUE TO LEFT HIP PAIN, RATED AS 8 OUT OF 10 BOTH AT REST AND WHEN ACTIVE. AN MRI OF HIS LEFT HIP WAS PERFORMED ON (B)(6) 2017 AND REVEALED A LARGE COMPLEX SOFT TISSUE FLUID COLLECTION WITH SURROUNDING SOFT TISSUE AND MUSCLE EDEMA, REACTIVE MARROW SIGNAL CHANGES IN THE SUPERIOR AND ANTERIOR LEFT ACETABULUM AS WELL AS POSTERIOR PROXIMAL LEFT FEMUR AND ATROPHY OF THE GLUTEUS, MEDIUS AND MINIMUS MUSCLES ON THE LEFT. DUE TO THE LARGE FLUID COLLECTION THERE WAS LOSS OF A NORMAL ANATOMY WITH THE DISTAL LEFT GLUTEAL TENDON INSERTION SHOULD INSERT. DUE TO THE MRI FINDINGS, SURGEON FELT THAT PATIENT LIKELY HAD TAPER CORROSION. SURGEON DETERMINED THAT THE PATIENT WOULD NEED HEAD EXCHANGE. HIS COBALT AND CHROMIUM LEVELS WERE FOUND TO BE 4.6 AND 2.6 RESPECTIVELY. ON (B)(6) 2017 A LEFT HIP REVISION SURGERY WAS PERFORMED FOR TRUNNIONOSIS, METAL DEBRIS LEFT HIP, AND NECROSIS OF SOFT TISSUE, INCLUDING ABDUCTOR, TROCHANTER OF THE LEFT TOTAL HIP. INTRAOPERATIVELY, SURGEON IDENTIFIED THE VASTUS LATERALIS, HOWEVER, THERE WAS NO SIGNIFICANT POSTERIOR STRUCTURES. THE PATIENT HAD ABOUT 2CM THICK LAYER OF WHITE NECROTIC TISSUE TROCHANTERIC BONE THROUGHOUT AND
Patient problems: Injury
MAUDE report 8600774 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K022902, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED devices
- PINNACLE 1217-25-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED735 reports
- NA 1365-52-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED683 reports
- APEX 1246-03-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED3 deaths613 reports
- PINNACLE 1217-20-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED2 deaths590 reports
- PINNACLE 1217-30-500Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 death484 reports
- NA 1365-53-000Depuy (Ireland) · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED404 reports
- PINNACLE ALTRX 1221-36-054Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED388 reports
- PINNACLE ALTRX 1221-32-050Depuy Orthopaedics, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED386 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.