Accord 71340007Recalls & reported problems
Smith & Nephew, Inc. · ACCORD 2.0MM COBALT CHROME CABLE WITH CLAMP · Class 2 · Implant
All models of Accord (14)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Accord 2.0mm cobalt chrome cable with clamp is used to secure bone and tissue during hip and other fixation surgeries. The 47 reports mostly describe the cable or its clamp breaking, fraying or unraveling while being tightened, and sometimes slipping out of place during surgery. 29 reports involved injuries and none involved deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 47
- Deaths reported
- 0
- Injuries reported
- 29
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 47 adverse event reports that list this exact device (DI 03596010485731), received between Apr 12, 2017 and Aug 4, 2025 — classified as 29 injuries, 18 malfunctions.
The most frequently reported specific device problem is “Break” (19 reports).
1 other recall cover devices cleared under the same FDA submission (K031162, K211176, K223762, K233949); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break19
- Insufficient Device Problem Information7
- Material Frayed6
- Mechanical Problem5
- Unintended Movement3
- Adverse Event Without Identified Device or Use Problem2
- Fracture2
- Activation, Positioning or Separation Problem1
Patient problems
- Injury10
- No Consequences Or Impact To Patient10
- No Clinical Signs, Symptoms or Conditions6
- No Known Impact Or Consequence To Patient6
- No Code Available3
- Unspecified Infection3
- Joint Dislocation2
- Bone Fracture(s)1
Recent adverse event reports
The 25 most recent of 47 reports, in the reporter's own words as published by the FDA.
InjuryAug 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A THR SURGERY HAD BEEN PERFORMED ON (B)(6) 2025, THE PATIENT EXPERIENCED DISCOMFORT. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2025, IN WHICH ONE (1) ACC 2.0MM COCR CABLE W/CLAMP WAS REMOVED. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Discomfort
MAUDE report 22710727 ↗MalfunctionSep 30, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, AN ACC 2.0MM COCR CABLE W/CLAMP TORE WHILE TENSIONING. ALL PIECES WERE REMOVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20338562 ↗InjuryAug 7, 2024Fracture; Device Slipped; Fracture; Device Slipped
IT WAS REPORTED THAT, DURING STEM PREPARATION IN A THR SURGERY, A FEMORAL FRACTURE OCCURRED WHILE USING A NON-SMITH AND NEPHEW PETER BREHM STEM, FOLLOWED BY ONE (1) ACC 2.0MM COCR CABLE W/CLAMP SLIPPING OUT OF THE CLAMP DESPITE THE TORQUE LOCK AS A CONSEQUENCE OF THE BONE FRACTURE. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, USING A COMPETITOR¿S SYNTHES CABLE DEVICE. A SAMPLE DEVICE SLIPPED FROM THE BONE INTO THE SOFT TISSUE CAUSING POTENTIAL INJURIES TO MUSCLE AND TO OTHER NON VERIFIABLE AREA SINCE IT WAS OVERLAID. ALSO, ADDITIONAL X-RAYS WERE USED. NO OTHER COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: Bone Fracture(s); Bruise/Contusion; Erosion; Hip Fracture; Insufficient Information
MAUDE report 19935338 ↗MalfunctionJul 5, 2024Unraveled Material; Unraveled Material
IT WAS REPORTED THAT, DURING THE EXCHANGE OF A COMPETITOR'S THR PRIMARY SYSTEM TO A REDAPT SYSTEM, AT THE MOMENT TENSIONING TWO (2) ACCORD 2.0MM COBALT CHROME CABLES, THE FIBERS BEGAN TO UNRAVEL. IT IS UNKNOWN IF THIS FAILURE REQUIRED FURTHER INTERVENTION. THE PROCEDURE WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT'S HEALTH CONDITION IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19680227 ↗MalfunctionNov 30, 2023Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, WHILE TIGHTENING, THE CLAMP OF AN ACC 2.0MM COCR CABLE W/CLAMP BROKE. NO PIECES FELL INSIDE THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18237578 ↗InjuryMar 2, 2022Material Frayed; Material Frayed
IT WAS REPORTED THAT, DURING TOTAL HIP REPLACEMENT, THE CABLE OF AN ACCORD 2.0MM COBALT CHROME CABLE W/CLAMP STARTED TO FRAY WHILE TENSIONING. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13655181 ↗MalfunctionApr 6, 2021Material Disintegration; Material Disintegration
IT WAS REPORTED THAT, DURING A THR WHEN THE ACC 2.0MM COCR CABLE W/CLAMP WAS CUT IT FRAYED AT THE END. THE PROCEDURE WAS COMPLETED WITH A DELAY LESS THAN OR EQUAL TO 30MIN USING A SMITH-NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11624909 ↗InjuryMar 30, 2021Break; Break
IT WAS REPORTED THAT, DURING PROCEDURE THE ACC 2.0MM COCR CABLE W/CLAMP WAS CUT OFF WHEN TENSIONER WAS APPLIED. IT HAD BEEN SHIFTED SEVERAL TIMES. INSTRUMENTS INSIDE THE PATIENT. IT CAN NOT DENY POSSIBILITY OF REMAINING IN THE BODY. THE PROCEDURE WAS COMPLETED WITH A DELAY LESS THAN OR EQUAL TO 30MIN WITH A SMITH-NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11593536 ↗InjuryOct 19, 2020Material Frayed; Material Frayed
IT WAS REPORTED THAT DURING SURGERY, THE 2.0 ACCORD CABLE FRAYED WHILE INSIDE THE PATIENT. ITEM WAS RETRIEVED FROM PATIENT AND PROCEDURE CONCLUDED WITH A DELAY OF 5 MINUTES AND WITHOUT AN INJURY. IT IS UNKNOWN IF A BACKUP DEVICE FROM SMITH AND NEPHEW WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10696478 ↗MalfunctionSep 30, 2020Break; Break
IT WAS REPORTED THAT THE CABLE BROKE DURING A REVISION SURGERY (NON S&N DEVICES). NO DELAY REPORTED. NO PATIENT INJURIES REPORTED. AN S&N BACKUP WAS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10604572 ↗MalfunctionJul 24, 2020Break; Break
IT WAS REPORTED THAT DURING SURGERY THE SURGEON USED AN ACCORD CABLE, THE CABLE BROKE DURING FASTENING PROCESS. THERE WAS NO INJURY TO PATIENT AND A SECOND IMPLANT WAS AVAILABLE. THERE WAS NO DELAY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10319274 ↗InjuryJun 18, 2020Break; Break
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION OF FEMORAL FRACTURE, TWO ACCORD CABLES WERE BROKE INSIDE THE PATIENT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10171577 ↗InjuryJun 18, 2020Break; Break
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION OF FEMORAL FRACTURE, TWO ACCORD CABLES WERE BROKE INSIDE THE PATIENT. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10171571 ↗MalfunctionDec 20, 2019Break; Break
IT WAS REPORTED THAT THE CABLE BROKE DURING SURGERY. THERE WAS NO DELAY NOR INJURY REPORTED. THE DEVICE BROKE INSIDE THE PATIENT BUT NO PIECES FELL INSIDE THE INCISION. A BACKUP DEVICE WAS USED TO COMPLETE THE SURGERY. THE SURGERY WAS SUCCESSFULLY COMPLETED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9503325 ↗MalfunctionSep 24, 2019Break; Break
IT WAS REPORTED THAT DURING A THR PROCEDURE, CABLE UNRAVELED WHILE TENSIONING AND BROKE. DELAY FROM (0-30) MINUTES REPORTED. NO DELAY. NO IMPACT OR INJURY TO PATIENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9114228 ↗InjurySep 3, 2019Break; Break
MalfunctionJul 3, 2019Break; Material Frayed; Break; Material Frayed
IT WAS REPORTED AS THE SURGEON BEGAN TO TENSION THE CABLE AFTER FOLLOWING THE PROTOCOL ON HOW TO PASS IT, THE CABLE BEGAN TO FRAY AND SPLIT. NOTHING VISUALLY WAS CAUSING THE PROBLEM. THE CABLE SCREW WAS UNLOCKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8761644 ↗MalfunctionJul 3, 2019Break; Material Frayed; Break; Material Frayed
IT WAS REPORTED AS THE SURGEON BEGAN TO TENSION THE CABLE AFTER FOLLOWING THE PROTOCOL ON HOW TO PASS IT, THE CABLE BEGAN TO FRAY AND SPLIT. NOTHING VISUALLY WAS CAUSING THE PROBLEM. THE CABLE SCREW WAS UNLOCKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8761631 ↗MalfunctionJun 26, 2019Break; Material Frayed; Break; Material Frayed
IT WAS REPORTED AS THE SURGEON BEGAN TO TENSION THE CABLE AFTER FOLLOWING THE PROTOCOL ON HOW TO PASS IT, THE CABLE BEGAN TO FRAY AND SPLIT. NOTHING VISUALLY WAS CAUSING THE PROBLEM. THE CABLE SCREW WAS UNLOCKED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8738108 ↗MalfunctionApr 7, 2019Break; Break
IT WAS REPORTED THAT DURING THAT DURING THE TREATMENT OF A INTERTROCHANTERIC COMMINUTED FRACTURE, THE CABLE SUDDENLY BROKE. DELAY OF LESS THAN 30MIN REPORTED. NO INJURY REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8489944 ↗MalfunctionOct 2, 2018Activation, Positioning or Separation Problem; Activation, Positioning or Separation Problem
IT WAS REPORTED THAT DEVICE RELEASED DURING PROCEDURE. DELAY OVER 2 HOURS, BACK UP AVAILABLE, NO INJURY TO PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7929013 ↗InjuryAug 23, 2018Loss of or Failure to Bond
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO THE PATIENT'S TROCHANTERIC REGION AND GREATER TROCHANTER WAS COMPLETELY UNUNITED. THE CABLES AND CLAW WERE REMOVED. THE TROCHANTERIC NONUNION WAS FIXED BY REAPPROXIMATING THE GREATER TROCHANTER IN THE APPROPRIATE POSITION AND PLACING A NEW CABLE CLAW CONSTRUCT IN THE APPROPRIATE POSITION.
Patient problems: Patient Problem/Medical Problem
MAUDE report 7808653 ↗InjuryAug 23, 2018Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO THE PATIENT'S TROCHANTERIC REGION AND GREATER TROCHANTER WAS COMPLETELY UNUNITED. THE CABLES AND CLAW WERE REMOVED. THE TROCHANTERIC NONUNION WAS FIXED BY REAPPROXIMATING THE GREATER TROCHANTER IN THE APPROPRIATE POSITION AND PLACING A NEW CABLE CLAW CONSTRUCT IN THE APPROPRIATE POSITION.
Patient problems: No Code Available
MAUDE report 7808643 ↗InjuryAug 23, 2018Loose or Intermittent Connection
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO THE PATIENT'S TROCHANTERIC REGION AND GREATER TROCHANTER WAS COMPLETELY UNUNITED. THE CABLES AND CLAW WERE REMOVED. THE TROCHANTERIC NONUNION WAS FIXED BY REAPPROXIMATING THE GREATER TROCHANTER IN THE APPROPRIATE POSITION AND PLACING A NEW CABLE CLAW CONSTRUCT IN THE APPROPRIATE POSITION.
Patient problems: No Code Available
MAUDE report 7808641 ↗MalfunctionAug 22, 2018Break; Material Separation
IT WAS REPORTED THAT METAL THREADS UNBRAIDED WHEN THE SURGEON HAS TIGHTENED AND THE CABLE BROKE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7806651 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K031162, K211176, K223762, K233949. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
Other CERCLAGE, FIXATION devices
- ZipFix 03.501.080Synthes GmbH · CERCLAGE, FIXATION236 reports
- CONTROL 1739-28-000Depuy Orthopaedics, Inc. · CERCLAGE, FIXATION52 reports
- Dall-Miles 6704-0-110Howmedica Osteonics Corp. · CERCLAGE, FIXATION1 recall1 report
- Tritium® Sternal Cable Plate System 86-27-12-S12Pioneer Surgical Technology, Inc. · CERCLAGE, FIXATION1 recall0 reports
- ZipFix 08.501.001.01SSynthes GmbH · CERCLAGE, FIXATION39 reports
- NA 298.801.01SSynthes GmbH · CERCLAGE, FIXATION33 reports
- ZipFix 08.501.001.05SSynthes GmbH · CERCLAGE, FIXATION29 reports
- NA 298.803SSynthes GmbH · CERCLAGE, FIXATION24 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page