Jet-X 71065458Recalls & reported problems

Smith & Nephew, Inc. · TiN 5MM X 45MM SHORT HALF PIN · Class 2 · Implant

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010487315), received between Jan 7, 2022 and Feb 7, 2022 — classified as 2 injuries.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Break2

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Rupture1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 7, 2022Break; Break

    IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, (2) JET-X TIN 5MM X 45MM SHORT HALF PINS BROKE OFF IN THE FEMUR. IT IS UNKNOWN IF/HOW SURGERY WAS COMPLETED, AND IF THERE WAS ANY DELAY. NO PATIENT INJURY REPORTED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13476730 ↗
  • InjuryJan 7, 2022Break; Break

    IT WAS REPORTED THAT DURING A TRAUMA SURGERY, THE TIN 5MM X 45MM SHORT HALF PIN BROKE INSIDE THE PATIENT'S BONE WHEN TIGHTENING. THE SURGEON USED THE SET OF THE BROKEN PIN TO REMOVE THE HALF PIN AND IT TOOK AROUND AN HOUR TO GET OUT THE BROKEN PIECE. THE PROCEDURE WAS FINISHED WITH A COMPETITOR'S DEVICE. PATIENT WAS NOT HARMED BEYOND THE ISSUE REPORTED.

    Patient problems: Rupture

    MAUDE report 13191902 ↗

Other PIN, FIXATION, THREADED devices

All PIN, FIXATION, THREADED devices →