GENESIS II 71421502Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II POSTERIOR STABILIZED HIGH FLEXION INSERT SIZE 1-2 11MM · Class 2 · Implant
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010492340), received between May 5, 2017 and Dec 29, 2020 — classified as 7 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K032295); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information2
- Connection Problem1
- Device-Device Incompatibility1
- Difficult to Insert1
- Failure To Adhere Or Bond1
- Loss of Osseointegration1
- Positioning Problem1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Failure of Implant1
- Inadequate Osseointegration1
- Injury1
- Insufficient Information1
- Joint Dislocation1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryDec 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO DISLOCATION BETWEEN THE FEMORAL COMPONENT AND THE INSERT. THE DOCTOR SAYS THAT THE INSERT HAD EXCESSIVE WEAR.
Patient problems: Failure of Implant; Joint Dislocation
MAUDE report 11091553 ↗InjuryDec 16, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT A REVISION SURGERY OF GEN II WAS PERFORMED DUE TO AN UNSPECIFIED REASON. PATIENT OUTCOME IS UNKNOWN. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information
MAUDE report 11018617 ↗MalfunctionDec 5, 2020Positioning Problem; Positioning Problem
IT WAS REPORTED DURING THE ARTHROSCOPY PROCEDURE THAT THE GENESIS PE COULD NOT BE INSERTED. THE INCIDENT OCCURRED INSIDE THE PATIENT. THE PROCEDURE HAD A DELAY BETWEEN 0-30 MIN AND IT WAS NECESSARY TO USE A BACKUP FROM SMITH & NEPHEW. ALTHOUGH THE SURGEON BLAMES THE DEVICE, HE WAS SATISFIED WITH THE SURGICAL OUTCOME. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10953269 ↗InjuryNov 21, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED ON (B)(6) 2019 DUE TO THE INFECTION AND LOOSENING. PRIMARY SURGERY WAS PERFORMED ON (B)(6) 2011 AND GII FEM COMPONENT(71420052), GII TIB BASE (71420180), LEGION INSERT (71453202), PATELLA (71420580). ALL OF THE IMPLANTED DEVICES WERE REVISED. THE REVISED DEVICES WILL NOT BE RETURNED. THE DOCTOR DOES NOT CONSIDER AS THE FAILURE OF DEVICE.
Patient problems: Unspecified Infection
MAUDE report 9356863 ↗InjuryNov 20, 2018Loss of Osseointegration; Loss of Osseointegration
IT WAS REPORTED THAT REVISION SURGERY WAS PERFORMED DUE TO LOOSENING.
Patient problems: Inadequate Osseointegration
MAUDE report 8092455 ↗InjuryFeb 12, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO GONARTHROSIS AND INSTABILITY IN KNEE.
Patient problems: Injury
MAUDE report 7264266 ↗InjuryAug 2, 2017Difficult to Insert
IT WAS REPORTED THAT DURING KNEE ARTHROPLASTY, THE INSERT COULD NOT BE IMPACTED INTO TIBIAL TRAY AFTER SEVERAL ATEMPTS. FINALY A SAME MODEL BACK-UP DEVICE WAS USED TO COMPLETE THE SURGERY WITHOUTH ISSUES, BUT CAUSING SIGNIFICANT SURGICAL DEALY.
Patient problems: No Code Available
MAUDE report 6759469 ↗InjuryMay 5, 2017Failure To Adhere Or Bond; Connection Problem; Device-Device Incompatibility; Failure To Adhere Or Bond; Device-Device Incompatibility
IT WAS REPORTED THAT A KNEE REPLACEMENT SURGERY WAS EXTENDED FOR ONE HOUR DUE TO MULTIPLE ATTEMPTS TO STEADY LOCK THE INSERT INTO THE BASEPLATE. A BACK-UP INSERT WAS GIVEN WHICH LOCKED IN SUCCESSFULLY.
Patient problems: No Information; No Known Impact Or Consequence To Patient
MAUDE report 6547008 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032295. One submission often covers many models, so check the product and lot details in each record.
- Z-0449-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, MOBILE BEARING ARTICULAR INSERT RIGHT, SIZE 5, 9 MM QTY: (1), REF 71621509, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0446-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 9 MM QTY: (1), REF 71421508, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0445-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0448-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0444-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0447-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 9 MM QTY: (1), REF 71421515, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page