GENESIS II 71421511Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II POSTERIOR STABILIZED HIGH FLEXION INSERT SIZE 3-4 15MM · Class 2 · Implant
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010492937), received between Mar 7, 2018 and Dec 6, 2024 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (K032295); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Dislodged or Dislocated2
- Adverse Event Without Identified Device or Use Problem1
- Delivered as Unsterile Product1
- Inadequacy of Device Shape and/or Size1
- Insufficient Device Problem Information1
- Material Fragmentation1
- Unsealed Device Packaging1
Patient problems
- Injury2
- Discomfort1
- Failure of Implant1
- Joint Dislocation1
- Joint Disorder1
- Limited Mobility Of The Implanted Joint1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 6, 2024Delivered as Unsterile Product; Unsealed Device Packaging; Delivered as Unsterile Product; Unsealed Device Packaging
IT WAS REPORTED THAT, DURING THE SET UP FOR A KNEE SURGERY, IT WAS NOTICED THAT THE OUTER PACKAGING ON ONE (1) GII PS HI FLEX ISRT SZ 3-4 15 WAS INTACT, AND THE INNER PACKAGING HAS BEEN OPENED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. SINCE INCIDENT OCCURRED BEFORE PROCEDURE; PATIENT WAS NOT YET INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20860680 ↗InjuryJul 15, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT, AFTER A TKA WAS PERFORMED ON AN (B)(6) 2022, DURING WHICH A GENESIS II BASEPLATE, A LEGION PS FEMUR, AND A GII PS HI FLEX ISRT SZ 3-4 15 WERE PLACED, THE PATIENT EXPERIENCED POLY DISSOCIATION. THE SURGEON STATED THAT THE LOCKING MECHANISM OF THE GENESIS II BASEPLATE IS VERY ROBUST; HOWEVER, HE ENSURED DURING THE SURGERY THAT THE POLY WAS PROPERLY LOCKED IN. IT IS UNKNOWN HOW THIS ADVERSE EVENT WAS TREATED. THE PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Joint Dislocation
MAUDE report 19741491 ↗InjuryJan 4, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER A SURGERY HAD BEEN PERFORMED ON (B)(6) 2010, PATIENT REPORTED DISCOMFORT. THE ADVERSE EVENT WAS ADDRESSED WITH A REVISION SURGERY ON (B)(6) 2021 TO REPLACE THE GII PS HI FLEX ISRT SZ 3-4 15. THE CONDITION OF THE PATIENT IS UNKNOWN.
Patient problems: Discomfort
MAUDE report 13159755 ↗InjuryNov 5, 2020Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO MEDIAL LAXITY/GAPPING. DURING THE REVISION SURGERY THE SURGEON SWAPPED THE 15 MM HIFELX INSERT FOR A 21 MM.
Patient problems: Joint Disorder
MAUDE report 10797203 ↗InjuryJan 14, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO POLY DISLOCATION. POLY WAS EXCHANGED AND REPLACED. PRIMARY SURGERY PERFORMED ON (B)(6) 2018.
Patient problems: Injury
MAUDE report 9586489 ↗InjuryDec 31, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE RIGHT TIBIAL BASEPLATE AND THE PS TIBIAL INSERT WERE REVISED DUE TO PAIN.
Patient problems: Injury; Pain
MAUDE report 8206209 ↗InjuryMar 7, 2018Material Fragmentation; Material Fragmentation
IT WAS REPORTED A REVISION SURGERY FOR POLY REPLACEMENT SINCE PATIENT WAS LOOSE IN FLEXION. WHEN SURGEON OPENED THE KNEE, HE FOUND A PIECE OF CHIPPED POLY ON POSTERIOR SIDE.
Patient problems: Failure of Implant; Limited Mobility Of The Implanted Joint
MAUDE report 7320479 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K032295. One submission often covers many models, so check the product and lot details in each record.
- Z-0447-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 9 MM QTY: (1), REF 71421515, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0445-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0449-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, MOBILE BEARING ARTICULAR INSERT RIGHT, SIZE 5, 9 MM QTY: (1), REF 71621509, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0446-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 9 MM QTY: (1), REF 71421508, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0448-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
- Z-0444-2014 · Aug 27, 2013 · TerminatedGENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page