Skip to content

GENESIS II 71421523Recalls & reported problems

Smith & Nephew, Inc. · GENESIS II POSTERIOR STABILIZED HIGH FLEXION INSERT SIZE 7-8 11MM · Class 2 · Implant

All models of GENESIS II (462)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010492982), received between Jul 27, 2018 and Oct 19, 2024 — classified as 8 injuries.

The most frequently reported specific device problem is “Fracture” (2 reports).

6 other recalls cover devices cleared under the same FDA submission (K032295); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2
  • Loosening of Implant Not Related to Bone-Ingrowth2
  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Insufficient Device Problem Information1
  • Loose or Intermittent Connection1
  • Mechanical Problem1
  • Naturally Worn1

Patient problems

  • Joint Laxity4
  • Injury3
  • Implant Pain2
  • Failure of Implant1
  • Insufficient Information1
  • Intraoperative Pain1
  • Joint effusion1
  • Joint locking1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 19, 2024Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A TKA SURGERY THAT HAD BEEN PERFORMED IN 2009, THE TENON OF ONE (1) GII PS HI FLEX ISRT SZ 7-8 11 BROKE. THE PATIENT EXPERIENCED RECURRENT JOINT BLOCKAGES, INSTABILITY AND SYNOVITIS, SO AN ARTHROSCOPY WAS THEN PERFORMED AND REVEALED A FRACTURE OF THE INLAY WITH REMOVAL OF THE PEG. THIS ADVERSE EVENT WAS ADDRESSED BY A REVISION SURGERY ON (B)(6) 2024 IN WHICH WITH REPLACEMENT OF THE INSERT. THE PATIENT HEALTH STATUS IS PHYSICALLY ACTIVE AND DOES NOT PARTICIPATE IN HIGH-RISK SPORTS. POSTOPERATIVE REHABILITATION AFTER THE REVISION SURGERY WENT SMOOTHLY.

    Patient problems: Failure of Implant; Implant Pain; Joint Laxity; Joint effusion; Joint locking; Swelling/ Edema; Synovitis

    MAUDE report 20489499 ↗
  • InjuryMar 28, 2023Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED, THE PATIENT EXPERIENCED INSERT DAMAGED, OSTEOLYSIS AND LOOSENING OF IMPLANTS. THIS ADVERSE EVENT WAS SOLVED THROUGH A REVISION SURGERY PERFORMED ON (B)(6)2023, IN WHICH THE GII PS HI FLEX ISRT SZ 7-8 11, LEGION PS OXIN FEM SZ6 RT AND GNS II RESURF PAT 35MM WERE EXPLANTED AND CHANGED. PATIENT STATUS IS UNKNOWN.

    Patient problems: Joint Laxity; Osteolysis

    MAUDE report 16632594 ↗
  • InjurySep 28, 2022Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT, AFTER A TKA PERFORMED ON 2011, THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2022 TO EXCHANGE THE GENESIS II POSTERIOR STABILIZED HIGH FLEXION INSERT SIZE 7-8 11MM. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Insufficient Information; Joint Laxity

    MAUDE report 15496219 ↗
  • InjuryFeb 28, 2022Break; Break

    IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON UNKNOWN DATE, THE PATIENT EXPERIENCED ANTERIOR KNEE PAIN AND INSTABILITY DUE TO THE PEG OF THE POLY HAD BROKEN OFF AND THERE WAS LOT OF WEAR. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION SURGERY TO EXCHANGE THE GII PS HI FLEX ISRT SZ 7-8 11 WITH A CONSTRAINED POLY ON (B)(6) 2022. CURRENT HEALTH STATUS OF PATIENT IS GOOD.

    Patient problems: Implant Pain; Joint Laxity

    MAUDE report 13619554 ↗
  • InjuryJul 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO A LOOSE PATELLA. PATELLA AND INSERT WERE EXCHANGED.

    Patient problems: Injury

    MAUDE report 8823649 ↗
  • InjuryOct 1, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT REVISION SURGERY WAS PERFORMED DUE TO ANTERIOR BASE LEFT KNEE PAIN.

    Patient problems: Intraoperative Pain

    MAUDE report 7924486 ↗
  • InjuryOct 1, 2018Fracture; Fracture

    IT WAS REPORTED THAT THE PATIENT FELL AND INJURED HIS RIGHT KNEE. HE DIDN'T SEEK MEDICAL ASSISTANCE FOR 12 MONTHS. WITHIN THAT TIME, THE TIBIAL BASE PLATE HAS PROGRESSIVELY GONE INTO VARUS CAUSING PAIN. DOCTOR BELIEVES THAT THE PATIENT FRACTURED UNDER THE TIBIAL TRAY IN THEIR FALL WHICH CAUSED THE TIBIAL TRAY TO FALL INTO VARUS.

    Patient problems: Injury; Pain

    MAUDE report 7921504 ↗
  • InjuryJul 27, 2018Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT WHEN TRYING TO IMPLANT THE ARTICULAR INSERT IT DIDN'T WANT TO LOCK IN PLACE CORRECTLY. AFTER NUMEROUS ATTEMPTS A NEW INSERT NEEDED TO ORDERED AND DELIVERED FROM S&N. DELAY GREATER THAN 30 MINUTES OCCURED DURING SURGERY.

    Patient problems: Injury

    MAUDE report 7726923 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K032295. One submission often covers many models, so check the product and lot details in each record.

  • Z-0447-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 9 MM QTY: (1), REF 71421515, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
  • Z-0445-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 13 MM QTY: (1), REF 71420820, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
  • Z-0444-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, POSTERIOR STABILIZED ARTICULAR INSERT, SIZE 1-2, 13 MM QTY: (1), REF 71420806, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
  • Z-0446-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 3-4, 9 MM QTY: (1), REF 71421508, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
  • Z-0448-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, PS HIGH FLEXION ARTICULAR INSERT, SIZE 5-6, 11 MM QTY: (1), REF 71421516, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic
  • Z-0449-2014 · Aug 27, 2013 · Terminated
    GENESIS(R) II, MOBILE BEARING ARTICULAR INSERT RIGHT, SIZE 5, 9 MM QTY: (1), REF 71621509, STERILE EO, Smith & Nephew, Inc. Memphis, TN 38116 USA Orthopedic

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page