GENESIS II 71441154Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II MINAMALLY INVASIVE DISTAL CUT FIRST ANTERIOR POSTERIOR CUTTING BLOCK SIZE 6 · Class 2
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Genesis II cutting block is a reusable guide used during knee replacement surgery. Reports mostly describe fixation spikes or pegs breaking off during use, with pieces sometimes falling into the surgical site and later removed. There are 23 reports naming its identifier, with no deaths and 3 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 23
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 03596010497222), received between Jul 29, 2019 and Jun 17, 2026 — classified as 3 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Fracture” (13 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture13
- Break5
- Component Missing1
- Crack1
- Device Damaged by Another Device1
- Dull, Blunt1
- Material Deformation1
- Material Separation1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Patient Involvement3
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
- Injury1
Recent adverse event reports
The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 17, 2026Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE SPIKE ON THE BACK OF ONE (1) GII MIS DCF AP CT BLK 6 BROKE OFF INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25537800 ↗MalfunctionMay 30, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING TKA SURGERY, THE PEGS OF THE GII MIS DCF AP CT BLK 6 BROKE OFF WHILE IMPACTING BLOCK. ALL PIECES WERE REMOVED FROM PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22119622 ↗MalfunctionMay 29, 2025Fracture; Device Damaged by Another Device; Fracture; Device Damaged by Another Device
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE MALLET IMPACT BROKE OFF THE FIXATION SPIKE OF ONE (1) GII MIS DCF AP CT BLK 6. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22106522 ↗MalfunctionJan 30, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, WHILE IMPACTING, THE SPIKE OF THE GII MIS DCF AP CT BLK 6 BROKE OFF. IT IS UNKNOWN IF ANY PIECE FELL INSIDE THE INCISION. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21274070 ↗MalfunctionJan 15, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, A SPIKE OF ONE (1) GII MIS DCF AP CT BLK 6 FRACTURED UPON REMOVAL. THE PROCEDURE WAS COMPLETED USING A S+N BACKUP DEVICE. NO PIECES FELL INTO THE PATIENT. THERE WAS NO SURGICAL DELAY OR INJURIES REPORTED AS A RESULT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21163854 ↗MalfunctionSep 2, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY ONE (1) GII MIS DCF AP CT BLK 6 BROKE OFF IN FEMORAL ANATOMY DURING NORMAL USE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20118204 ↗MalfunctionJun 7, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA, WHEN REMOVING THE GII MIS DCF AP CT BLK 6, ONE OF THE FIXATION SPIKES BROKE OFF. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19486127 ↗MalfunctionApr 5, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, WHILE IMPACTING ONE (1) GII MIS DCF AP CT BLK 6, ONE OF ITS PINS BROKE. THE BROKEN PIN WAS RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19053369 ↗MalfunctionApr 27, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, ONE OF THE SPIKES ON BACK OF THE GII MIS DCF AP CT BLK BROKE OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16825884 ↗MalfunctionJan 30, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, A GII MIS DCF AP CT BLK 6 BROKE INTO TWO PIECES. IT HAPPENED WHEN THE SURGEON REMOVED IT FROM BONE AFTER USING IT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16268141 ↗InjuryDec 5, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE SPIKE ON THE BACK OF THE GII MIS DCF AP CT BLK 6 DEVICE BROKE OFF INTO PATIENT AND WAS RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, WITH A S+N BACK UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15921679 ↗MalfunctionOct 11, 2022Fracture; Fracture
IT WAS REPORTED THAT A GII MIS DCF AP CT BLK 6 IS BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15578018 ↗InjurySep 12, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING TKA PROCEDURE, THE SPIKE ON BACK OF THE GII MIS DCF AP CT BLK 6 BROKE OFF INTO FEMUR. THE SPIKE WAS LEFT INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15399847 ↗MalfunctionMar 8, 2022Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THAT THE GII MIS DCF AP CT BLK 6 BROKE OFF FROM HITTING WITH MALLET THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13691429 ↗MalfunctionSep 15, 2021Material Separation; Material Separation
IT WAS REPORTED THAT A GENESIS II MINIMALLY INVASIVE DISTAL CUT FIRST ANTERIOR POSTERIOR CUTTING BLOCK SIZE 6 DISTAL PEG IS NO LONGER ATTACHED. NO CASE INVOLVED, THEREFORE NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12477950 ↗MalfunctionMar 29, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT, COMPONENT ARE UNFIT FOR USE. THE GII MIS DCF AP CT BLK 3 HAVE BURRS. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11586446 ↗MalfunctionFeb 23, 2021Crack; Crack
IT WAS REPORTED THAT THE CHAMFER OF CUTTING BLOCK CRACKED AT SLAPHAMMER ATTACHMENT POINT. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11364943 ↗MalfunctionOct 22, 2020Dull, Blunt; Dull, Blunt
IT WAS REPORTED THAT THE SPIKE OF THE GII MIS DCF AP CT BLK 6 HAS DEVELOPED AN ANGLE FROM WEAR AND TEAR. NO CASE OR PATIENT INVOLVED.
Patient problems: No Patient Involvement
MAUDE report 10721931 ↗MalfunctionSep 20, 2020Break; Break
IT WAS REPORTED THAT AFTER PROCEDURE THE SIZE 6 FEMORAL CUT BLOCK PEG WAS FOUND BROKEN. NO DELAY REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10558291 ↗MalfunctionSep 10, 2020Naturally Worn; Naturally Worn
IT WAS REPORTED THAT THE GII MIS DCF AP CT BLK 6 IS WORN FROM EXCESSIVE USE. THERE WAS NO CASE INVOLVED. NO PATIENT HARM OR SURGICAL DELAY WAS REPORTED.
Patient problems: No Patient Involvement
MAUDE report 10518437 ↗InjuryJun 19, 2020Break; Component Missing; Separation Failure; Break; Component Missing; Separation Failure
IT WAS REPORTED THAT AFTER PROCEDURE TKA THE SURGEON DID NOT NOTICE THE MISSING PEG FROM THE CUTTING BLOCK UNTIL POSTOPERATIVE X-RAY (DURING THE PROCEDURE THE SURGEON FOLLOWED PROPER TECHNIQUE IN PLACEMENT AND REMOVAL OF THE 4 IN 1 CUTTING BLOCK). THE PROCEDURE WAS COMPLETED WITHOUT DELAY.
Patient problems: Injury
MAUDE report 10176766 ↗MalfunctionNov 25, 2019Break; Break
IT WAS REPORTED THAT THERE IS A BROKEN LEGION CUTTING BLOCK. IT BROKE INSIDE OF THE BONE OF A PATIENT. NO INJURIES OR DELAYS REPORTED. S+N BACKUP DEVICE WAS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9370105 ↗MalfunctionJul 29, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY, PEG BROKE INSIDE PATIENT. SURGEON WAS ABLE TO EXTRACT IT WITH TWEESERS. USED A SECOND INSTRUMENT TO RESUME SURGERY. DELAY FROM 30 MINUTES TO 1 HOUR REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8839757 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.