Skip to content

REFLECTION 71323114Recalls & reported problems

Smith & Nephew, Inc. · REFLECTION CONSTRAINED LINER 0 DEG 26 INNER DIAMETER 50-52 OUTER DIAMETER SIZE E · Class 2

All models of REFLECTION (112)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010510884), received between May 18, 2017 and Mar 9, 2026 — classified as 7 injuries.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K033442, K211176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Degraded1
  • Device Damaged by Another Device1
  • Device Dislodged or Dislocated1
  • Fracture1
  • Unintended Movement1

Patient problems

  • Joint Dislocation4
  • Injury2
  • Discomfort1
  • Failure of Implant1
  • Loss of Range of Motion1
  • No Known Impact Or Consequence To Patient1
  • Pain1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 9, 2026Fracture

    IT WAS REPORTED THAT, AFTER A REVISION HIP SURGERY HAD BEEN PERFORMED IN 2017, THE PATIENT EXPERIENCED PAIN, DISCOMFORT AND LIMITED MOBILITY DUE TO A DISLOCATED HIP, AND IT APPEARS ON X-RAY THAT THE LOCKING RING OF THE CON LNR 0DEG 26ID 50-52OD SZ E HAS BROKEN, AND THE COCR 12/14 FEM HEAD 26 + 8 HAS MIGRATED SUPERIORLY. THE ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON, IN WHICH THE CON LNR 0DEG 26ID 50-52OD SZ E WAS REMOVED AND A CEMENTED POLARCUP WAS INSERTED INTO THE EXISTING SHELL. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Discomfort; Joint Dislocation; Loss of Range of Motion; Pain

    MAUDE report 24545599 ↗
  • InjuryMar 29, 2022Break; Degraded; Break; Degraded

    IT WAS REPORTED THAT, AFTER A THA SURGERY HAD BEEN PERFORMED ON 2004, THE PATIENT EXPERIENCED DISLOCATION OF HEAD FROM WITHIN CONSTRAINED LINER. THIS ADVERSE EVENT WAS SOLVED BY A REVISION SURGERY ON (B)(6) 2022, IN WHICH IT WAS FOUND THAT THE CON LNR 0DEG 26ID 50-52OD SZ E RING WAS BROKEN AND SEPARATED. IT WAS HARD TO TELL EXACTLY HOW MUCH WEAR THE LINER HAD AND WHETHER THIS WAS A FACTOR IN THE DISLOCATION OF THE THA. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.

    Patient problems: Failure of Implant

    MAUDE report 13945671 ↗
  • InjuryMar 1, 2022Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT, AFTER A LEFT THR SURGERY HAD BEEN PERFORMED ON AN UNSPECIFIED DATE, THE PATIENT EXPERIENCED DISLOCATION. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION SURGERY TO EXCHANGE THE CON LNR 0DEG 26ID 50-52OD SZ E, WITH COMPETITOR PRODUCTS. THE PATIENT CURRENT HEALTH STATUS IS UNKNOWN. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: Joint Dislocation

    MAUDE report 13635916 ↗
  • InjuryAug 9, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER A THR PERFORMED ON (B)(6) 2021, IT WAS REPORTED THAT A PATIENT EXPERIMENTED A DISLOCATION OF THE IMPLANT. A REVISION SURGERY TOOK PLACE ON (B)(6) 2021 TO EVALUATE THE BEST WAY TO FIX THE PROBLEM AND THE ACETABULAR CUP AND LINER WERE REPLACED WITH ANOTHER COMPANY'S IMPLANTS. NEW SMITH AND NEPHEW FEMORAL HEAD IMPLANTED. IT IS UNKNOWN HOW FEMORAL HEAD WAS ABLE TO DISLOCATE OUT OF CONSTRAINED LINER. PATIENT CONDITION IS UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 12294329 ↗
  • InjuryOct 8, 2020Unintended Movement; Unintended Movement

    IT WAS REPORTED THE FIRST REVISION TKA SURGERY WAS ON (B)(6) 2018. PATIENT SUFFERED DISLOCATION OF FEMORAL HEAD AND RETAINER RING FROM ACETABULAR SHELL AND CONSTRAINED LINER CONSTRUCT. PRIMARY SURGERY DATE UNKNOWN. SECOND REVISION (B)(6) 2020. NO INJURY REPORTED EXCEPT NEED FOR REVISION.

    Patient problems: Injury; Joint Dislocation

    MAUDE report 10653086 ↗
  • InjuryNov 18, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED FOR DISLOCATION CAUSED BY A BROKEN RING

    Patient problems: Injury

    MAUDE report 7043793 ↗
  • InjuryMay 18, 2017Device Damaged by Another Device; Device Damaged by Another Device

    IT WAS REPORTED A SURGERY WAS DELAYED BY MORE THAN ONE HOUR DUE TO THE LOCKING MECHANISM ON THE RING BREAKING WHEN IMPACTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6577038 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033442, K211176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0326-2025 · Oct 15, 2024 · Open, Classified
    Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.

Other Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices

All Prosthesis, hip, constrained, cemented or uncemented, metal/polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page