Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010535252), received between Dec 18, 2018 and Dec 19, 2025 — classified as 2 injuries, 7 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K113789); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Fracture3

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 19, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR, THE TRIAL REMOVAL HANDLE BROKE WHILE EXTRACTING TRIAL STEM FROM PATIENT'S FEMUR. NO PIECES FELL INTO PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23849909 ↗
  • MalfunctionOct 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, A REDAPT CASE, WHERE A TRIAL REMOVAL HANDLE BROKE DURING REGULAR USE-EXTRACTION OF FEMORAL STEM TRIAL. NO PIECES FELL INTO PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A COMPETITOR'S DEVICE., A HOOK HANDLE. PATIENT WAS ANESTHETIZED DURING THE DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20502443 ↗
  • InjuryAug 5, 2021Fracture; Fracture

    IT WAS REPORTED THAT AN EMPERION TRIAL REMOVAL HANDLE BROKE DURING SURGERY, WHILE INSTRUMENT WAS INSIDE THE PATIENT. ALL PIECES WERE RECOVERED. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO SURGICAL DELAY WAS REPORTED. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12285330 ↗
  • MalfunctionJul 12, 2021Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, A TRIAL REMOVAL HANDLE BROKE. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12155295 ↗
  • InjuryJan 12, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, TRIAL REMOVAL HANDLE BROKE. THE DEVICE BROKE DURING SURGERY, WHILE INSIDE THE PATIENT. ALL PIECES WERE RECOVERED. IT IS UNKNOWN HOW WAS THE PROCEDURE FINISHED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11156932 ↗
  • MalfunctionJan 4, 2021Break; Break

    IT WAS REPORTED THAT, DURING A THR PROCEDURE, THE TRIAL REMOVAL HANDLE BROKE OUTSIDE THE PATIENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING A SMITH+NEPHEW BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11114912 ↗
  • MalfunctionOct 30, 2020Break; Break

    IT WAS REPORTED THAT DURING TOTAL HIP REPLACEMENT THE TRIAL REMOVAL HANDLE BROKE IN CASE. NO DELAY. SURGERY WAS COMPLETED WITH ORIGINAL DEVICE. PATIENT WAS NOT AFFECTED BY INCIDENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10763277 ↗
  • MalfunctionJun 24, 2019Break; Break

    IT WAS REPORTED THAT HANDLE BROKE WHILE BACK SLAPPING IN AN EFFORT TO REMOVE A REDAPT STEM. NO DELAY AND BACK-UP DEVICE WAS AVAILABLE. UNKNOWN IF ANY INJURY OR PATIENT IMPACT WAS INVOLVED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8727951 ↗
  • MalfunctionDec 18, 2018Break; Break

    IT WAS REPORTED THAT WHEN SURGEON WAS REMOVING STEM, HANDLE BROKE. ALL PIECES WERE RETRIEVED, NO DELAY, NO INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8173820 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113789. One submission often covers many models, so check the product and lot details in each record.

  • Z-0945-2017 · Nov 15, 2016 · Terminated
    Modular SMF(TM) The Modular SMF(TM) and Modular REDAPT(TM) Hip Systems are indicated for individuals undergoing primary and revision surgery where other treatments or devices have

Other Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices

All Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page