Skip to content

LEGION 71421164Recalls & reported problems

Smith & Nephew, Inc. · LEGION OXINIUM CONSTRAINED FEMORAL SIZE 4 LEFT · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
5
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010543837), received between Oct 12, 2018 and Oct 14, 2024 — classified as 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

7 other recalls cover devices cleared under the same FDA submission (K043440, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 2, 2024Class IIOpen, Classified

    Complaint received indicating that a package contained a LGN OX CONSTRAINED FEM 3 LT implant instead of a LGN OX CONSTRAINED FEM 4 LT as described on the product label.

    Z-0278-2025 · Root cause: Packaging change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Appropriate Device Problem Term/Code Not Available1
  • Packaging Problem1
  • Wrong Label1

Patient problems

  • Injury2
  • Implant Pain1
  • Inflammation1
  • No Clinical Signs, Symptoms or Conditions1
  • Unspecified Infection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 14, 2024Packaging Problem; Wrong Label; Packaging Problem; Wrong Label

    IT WAS REPORTED THAT, DURING TKA REVISION SURGERY DUE TO IMPLANT LOOSENING, BOX AND STICKERS INDICATED THAT ONE (1) LGN OX CONSTRAINED FEM 4 LT WAS CONTAINED WITHIN THE PACKAGE, BUT INSTEAD IT CONTAINED ONE LGN OX CONSTRAINED FEM 3 LT (PART NUMBER 71421163; LOT NUMBER 23LM00479). THE DEVICE DIFFERENCE WAS NOTICED OUTSIDE THE PATIENT WHEN THE L-WEDGE BEING INSTALLED DID NOT FIT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS RESULT OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20445510 ↗
  • InjuryMay 11, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    LEGAL CASE: IT WAS REPORTED THAT, AFTER A PROSTHETIC RE-IMPLANT HAD OCCURRED ON (B)(6) 2018 WITH A LEGION KNEE SYSTEM, THE PATIENT STARTED EXPERIENCING PAIN. THE X-RAYS SHOWED THE FEMORAL AND THE TIBIAL STEMS WITH PHLOGOSIS. A REVISION SURGERY WAS NECESSARY TO REMOVE THE WHOLE SYSTEM AND REPLACED IT WITH CEMENT SPACERS ON (B)(6) 2021. THE OUTCOME OF THE PATIENT IS UNKNOWN.

    Patient problems: Implant Pain; Inflammation

    MAUDE report 11806625 ↗
  • InjurySep 16, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO INFECTION. ALL PARTS WERE REMOVED AND REPLACED WITH CEMENT MOULDS FROM HERAEUS. NO MORE INFORMATION AVAILABLE.

    Patient problems: Unspecified Infection

    MAUDE report 9073106 ↗
  • InjuryFeb 18, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO A FAILED PREVIOUS TOTAL KNEE ARTHROPLASTY.

    Patient problems: Injury

    MAUDE report 8348753 ↗
  • InjuryOct 12, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A LEFT KNEE ARTHROSCOPY WAS PERFORMED.

    Patient problems: Injury

    MAUDE report 7962370 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K043440, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-0277-2025 · Oct 2, 2024 · Open, Classified
    smith&nephew LEIGON OXINIUM CONSTRAINED NONPOROUS FEMORAL COMPONENT, SIZE 3 LEFT, REF 71421163; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POL
  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
  • Z-1951-2021 · May 19, 2021 · Terminated
    LEGION Posterior Stabilized OXINIUM Femoral Component, REF 71421216, Size 6, Left, Nonporous Femoral Component, A/P 65.5 MM, M/L 73 MM, For Cemented Use Only, ZR-2.5NB,Sterile. Kn
  • Z-2538-2019 · Jan 15, 2018 · Completed
    LEGION(TM) HEMI STEP TIBIAL WEDGE RK/HK, 5 MM, SIZE 3-4 LT.MDL/RT-LAT, REF 71423432 Product Usage: The constrained knee systems are designed for use in patients in primary and
  • Z-0816-2016 · Jan 18, 2016 · Terminated
    LEGION(TM) HEMI STEPPED, TIBIAL SCREW-ON WEDGE, REF 71421147, QTY 1, 5 MM, SIZE 5-6 LT-LAT / RT-MDL; Tibial knee prosthesis, augment.
  • Z-1820-2009 · Jan 22, 2009 · Terminated
    LEGION Pressfit Stem, Straight, 20 MM X 160 MM, REF 71424053, Sterile, Smith & Nephew, Memphis TX 38116. The device is used for Orthopedic Knee Implants.
  • Z-1819-2009 · Jan 22, 2009 · Terminated
    LEGION Pressfit Stem, Straight, 10 MM X 160 MM, REF 71424043, Sterile, Smith & Nephew, Memphis TX 38116. The device is used for Orthopedic Knee Implants.

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page