LEGION 71421168Recalls & reported problems
Smith & Nephew, Inc. · LEGION OXINIUM CONSTRAINED FEMORAL SIZE 8 LEFT · Class 2 · Implant
All models of LEGION (446)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010543875), received between Nov 2, 2018 and Dec 30, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
8 other recalls cover devices cleared under the same FDA submission (K043440, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information2
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Aseptic loosening1
- Insufficient Information1
- Joint Laxity1
- No Clinical Signs, Symptoms or Conditions1
- Pain1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 30, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT, AFTER A TKA SURGERY PERFORMED ON A CURRENTLY UNKNOWN DATE,THE PATIENT EXPERIENCED ASEPTIC LOOSENING OF THE LGN OX CONSTRAINED FEM 8 LT. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2025 IN WHICH ONE (1) LGN OX CONSTRAINED FEM 8 LT, ONE (1) LGN PS HIGH FLEX XLPE SZ 7-8 15MM, AND ONE (1) LGN PRESSFIT STEM 20MM X 160MM WERE REVISED. CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Aseptic loosening
MAUDE report 23926735 ↗InjuryOct 10, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT UNDERWENT A TKA LEFT REVISION SURGERY IN 2010 DUE TO ASEPTIC LOOSENING AND INSTABILITY. DURING THE REVISION SURGERY, LEGION/GII SMITH & NEPHEW COMPONENT(S) WERE IMPLANTED. 2,10 YEARS AFTER THIS PROCEDURE, THE PATIENT SUSTAINED AN UNSPECIFIED ADVERSE EVENT THAT REQUIRED A TKA RE-REVISION. IT IS UNKNOWN WHICH DEVICES WERE EXPLANTED/EXCHANGED DURING THIS PROCEDURE. THIS INFORMATION WAS COLLECTED RETROSPECTIVELY AS PART OF A POST-MARKET CLINICAL FOLLOW UP ACTIVITY AND IS PROVIDED IN AN ANONYMIZED FORMAT; THEREFORE, NO FURTHER INFORMATION AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20418519 ↗InjuryJan 19, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, THE PATIENT UNDERWENT A LEFT TKA REVISION DUE TO ASEPTIC LOOSENING AS PRIMARY DIAGNOSIS WITH BEARING WEAR, OSTEOLYSIS, INSTABILITY AND FAILED PATELLAR COMPONENT AS SECONDARY DIAGNOSES. APPROXIMATELY SIX (6) MONTHS AFTER THIS PROCEDURE, THE PATIENT EXPERIENCED INSTABILITY IN THE REVISED KNEE AND REQUIRED A SECOND REVISION. THE DATES IN WHICH THE INDEX SURGERY, FIRST REVISION AND SECOND REVISION TOOK PLACE ARE NOT KNOWN. IT IS ALSO UNKNOWN WHICH DEVICES WERE EXPLANTED AND/OR EXCHANGED DURING THE SECOND REVISION. THIS INCIDENT WAS IDENTIFIED IN A RETROSPECTIVE POST-MARKET CLINICAL FOLLOW UP ACTIVITY (PMCF) WHERE ANONYMIZED DATA SUMMARIZING TKA REVISION OUTCOMES WERE GATHERED; THEREFORE, ADDITIONAL INFORMATION IS NOT KNOWN.
Patient problems: Joint Laxity
MAUDE report 16210072 ↗MalfunctionMar 1, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING USE, OUTSIDE THE PATIENT, THE SCREW OF THE POSTERIOR WEDGE COULD NOT FULLY ENGAGE WITH THE LEGION OXINIUM FEMUR SCREW HOLE. BECAUSE OF THIS THE POSTERIOR WEDGE COULD NOT BE PROPERLY TIGHTENED AND WOULD STAY A BIT LOOSE. THE PROCEDURE FINISHED WITH A SMITH AND NEPHEW BACKUP DEVICE. SURGICAL DELAY LESS THAN 30 MINUTES. PATIENT INJURIES WERE NOT REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11391137 ↗InjuryNov 2, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT PATIENT HAD PAINFUL PATELLOFEMORAL CREPITATION ONE YEAR AFTER KNEE WAS REPLACED AND HAD INJECTED WITH 3CC XYLOCAINE, 3CC MARCAINE, 1CC DEPO-MEDROL.
Patient problems: Pain
MAUDE report 8035671 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K043440, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-0278-2025 · Oct 2, 2024 · Open, Classifiedsmith&nephew LEIGON OXINIUM CONSTRAINED NONPOROUS FEMORAL COMPONENT, SIZE 4 LEFT, REF 71421164; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POL
- Z-0277-2025 · Oct 2, 2024 · Open, Classifiedsmith&nephew LEIGON OXINIUM CONSTRAINED NONPOROUS FEMORAL COMPONENT, SIZE 3 LEFT, REF 71421163; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POL
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-1951-2021 · May 19, 2021 · TerminatedLEGION Posterior Stabilized OXINIUM Femoral Component, REF 71421216, Size 6, Left, Nonporous Femoral Component, A/P 65.5 MM, M/L 73 MM, For Cemented Use Only, ZR-2.5NB,Sterile. Kn
- Z-2538-2019 · Jan 15, 2018 · CompletedLEGION(TM) HEMI STEP TIBIAL WEDGE RK/HK, 5 MM, SIZE 3-4 LT.MDL/RT-LAT, REF 71423432 Product Usage: The constrained knee systems are designed for use in patients in primary and
- Z-0816-2016 · Jan 18, 2016 · TerminatedLEGION(TM) HEMI STEPPED, TIBIAL SCREW-ON WEDGE, REF 71421147, QTY 1, 5 MM, SIZE 5-6 LT-LAT / RT-MDL; Tibial knee prosthesis, augment.
- Z-1819-2009 · Jan 22, 2009 · TerminatedLEGION Pressfit Stem, Straight, 10 MM X 160 MM, REF 71424043, Sterile, Smith & Nephew, Memphis TX 38116. The device is used for Orthopedic Knee Implants.
- Z-1820-2009 · Jan 22, 2009 · TerminatedLEGION Pressfit Stem, Straight, 20 MM X 160 MM, REF 71424053, Sterile, Smith & Nephew, Memphis TX 38116. The device is used for Orthopedic Knee Implants.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page