LEGION 71421805Recalls & reported problems
Smith & Nephew, Inc. · LEGION SCREW-ON POSTERIOR FEMORAL WEDGE 5MM SIZE 5-6 LONG · Class 2 · Implant
All models of LEGION (446)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010545398), received between Jun 19, 2020 and Oct 13, 2020 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Incomplete or Inadequate Connection” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K041106, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incomplete or Inadequate Connection2
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryOct 13, 2020Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE LGN POS FEM WDG 5MM SZ5-6 LNG WOULD NOT TIGHTEN DOWN COMPLETELY EVEN WITH THE SCREW ALL THE WAY TIGHTENED. THIS RESULTED IN THE AUGMENT HAVING TO BE CEMENTED. THE PATIENT CONDITION IS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10673601 ↗InjuryOct 12, 2020Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE LGN POS FEM WDG 5MM SZ5-6 LNG WOULDN¿T TIGHTEN DOWN COMPLETELY EVEN WITH THE SCREW ALL THE WAY TIGHTENED. THIS RESULTED IN THE AUGMENT HAVING TO BE CEMENTED. THE PATIENT¿S CONDITION IS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10667415 ↗MalfunctionJun 19, 2020Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT DURING SURGERY WHEN TRYING TO ASSEMBLE THE 5 MM POSTERIOR AUGMENT THE SCREW WOULD NOT ENGAGE AND LOCK DOWN THE AUGMENT. IT WAS TRIED FOR AT LEAST 20 MINUTES. A BACK UP DEVICE WAS USED AND IT WENT DOWN EASILY. IT WAS FOUND THAT THE THAT THE SCREW WAS DIFFERENT IN THE SECOND PACKAGE. IT WAS BOTH SMALLER IN WIDTH AND LENGTH.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10173177 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K041106, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-0092-2018 · Sep 14, 2017 · Terminatedsmith&nephew LEGION L-WEDGE, 10 MM DIS X 5 MM POS, SIZE 4, SCREW-ON FEMORAL WEDGE, REF 71421733, STERILE R Product The Revision Knee System Components are indicated for rheumatoid
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page