Skip to content

LEGION 71424046Recalls & reported problems

Smith & Nephew, Inc. · LEGION PRESSFIT STEM 13MM X 160MM · Class 2 · Implant

All models of LEGION (446)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010546784), received between Sep 5, 2018 and Oct 25, 2023 — classified as 5 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K072531, K230653); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information3
  • Adverse Event Without Identified Device or Use Problem2
  • Appropriate Device Problem Term/Code Not Available1
  • Fitting Problem1

Patient problems

  • Unspecified Infection3
  • Injury2
  • Failure of Implant1
  • Loss of Range of Motion1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THREE (3) RIGHT REV FEMUR WITH 13X160 PRESSFIT STEM TRIALS WERE USED DURING TRIALING, THE SEATING WAS CORRECT, BUT DURING A TKA SURGERY PERFORMED ON (B)(6) 2023, UPON USAGE OF ACTUAL IMPLANTS IT SAT ROUGHLY 1CM PROUD AND WOULD NOT ADVANCE FURTHER. THE LGN PRESSFIT STEM 13MM X 160MM WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: Failure of Implant

    MAUDE report 18008255 ↗
  • InjuryFeb 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER TKR REVISION SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED RANGE OF MOTION ISSUES, PRESENTING LOOSE EXTENSION AND TIGHT IN FLEXION. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2023 TO ADDRESS THIS ADVERSE EVENT. THE FEMORAL COMPONENT, THE TIBIAL BASE PLATE, STEM AND THE TIBIAL INSERT WERE EXPLANTED. LEGION REVISION TIBIA AND FEMUR WERE IMPLANTED WITH FEMORAL DISTAL WEDGES AND A CONSTRAINED POLY INSERT. IT IS UNKNOWN CURRENT HEALTH STATUS OF PATIENT.

    Patient problems: Loss of Range of Motion

    MAUDE report 16325767 ↗
  • InjurySep 16, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO INFECTION. ALL PARTS WERE REMOVED AND REPLACED WITH CEMENT MOULDS FROM HERAEUS. NO MORE INFORMATION AVAILABLE.

    Patient problems: Unspecified Infection

    MAUDE report 9073111 ↗
  • MalfunctionMay 9, 2019Fitting Problem; Fitting Problem

    IT WAS REPORTED THAT DURING REVISION SURGERY THE CONSTRUCTED IMPLANTS WOULD NOT FIT PROPERLY IN PATIENT. MORE THAN 30 MINUTES OF DELAY WAS REPORTED AND BACKUP DEVICE WERE AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8597362 ↗
  • InjuryOct 29, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION WAS PERFORMED TO REMOVE THE IMPLANTS FROM THE PATIENT DUE TO INFECTION.

    Patient problems: Injury; Unspecified Infection

    MAUDE report 8017378 ↗
  • InjurySep 24, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A KNEE MANIPULATION UNDER ANESTHESIA WAS PERFORMED DUE TO STIFFNESS OF STUDY KNEE. NO SURGERY REQUIRED AND NO COMPLICATIONS.

    Patient problems: Injury

    MAUDE report 7900976 ↗
  • InjurySep 5, 2018Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT A REVISION SURGERY FO RIGHT KNEE WAS PERFORMED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 7846496 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K072531, K230653. One submission often covers many models, so check the product and lot details in each record.

  • Z-2621-2023 · Aug 31, 2023 · Open, Classified
    smith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint

Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices

All PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices →