LEGION 71434407Recalls & reported problems
Smith & Nephew, Inc. · LEGION 6 DEG VALGUS COLLET 6MM OFFSET SZ3-8 · Class 2
All models of LEGION (446)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010549150), received between Nov 3, 2020 and Apr 28, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam2
- Break1
- Material Deformation1
- Material Fragmentation1
- Positioning Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 28, 2025Material Fragmentation; Material Deformation; Positioning Problem; Material Fragmentation; Material Deformation; Positioning Problem
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) LEGION 6 DEG VALGUS COLLET 6MM OFFSET SZ3-8 WOULD NOT SEAT AROUND THE REVISION REAMER. THE PROCEDURE WAS RESUMED, AFTER A DELAY OF LESS THAN 30 MINUTES, USING A DIFFERENT SURGICAL TECHNIQUE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21922981 ↗MalfunctionFeb 16, 2022Break; Break
IT WAS REPORTED THAT DURING SET UP FOR TKA, THE LGN 6 DEG VAL COL 6MM OFF S3-8 NO LONGER SPINS/ SCREW MECHANISM IS BROKEN. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. AS THIS WAS NOTICED DURING SET UP FOR SURGERY, NO INJURY TO PATIENT OR DELAY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13540142 ↗MalfunctionJan 20, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, THE LGN 6 DEG VAL COL 6MM OFF S3-8 IS JAMMED AND UNABLE TO ROTATE PREVENTING THE SURGEON TO DIAL IN THE CORRECT ROTATION. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. THERE WAS NOT DELAY. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11205491 ↗MalfunctionNov 3, 2020Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A REVISION SURGERY THE VALGUS COLLET 6MM WAS SEIZED AND WOULD NOT ROTATE ON THE REAMER SHAFT. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACKUP DEVICE. THERE WAS A DELAY REPORTED OF LESS THAN OR EQUAL TO 30 MINUTES. NO PATIENT HARM WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10782147 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page