JOURNEY 74023241Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY ARTICULAR INSERT BCS STANDARD 3-4 LEFT 9MM · Class 2 · Implant
All models of JOURNEY (566)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010555069), received between Sep 6, 2018 and Dec 9, 2025 — classified as 8 injuries.
The most frequently reported specific device problem is “Fracture” (6 reports).
4 other recalls cover devices cleared under the same FDA submission (K042515, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture6
- Break1
- Insufficient Device Problem Information1
Patient problems
- Failure of Implant5
- Joint Laxity3
- Arthritis1
- Injury1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryDec 9, 2025Fracture; Fracture
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED IN 2007, PATIENT EXPERIENCED SUDDEN INSTABILITY FROM THE KNEE. THIS ADVERSE EVENT WAS ADDRESSED BY PERFORMING A REVISION SURGERY ON (B)(6) 2025, DURING WHICH, THE JOURN ART INS BCS STD 3-4 LT 9 WAS EXCHANGED DUE TO A BROKEN POST. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Joint Laxity
MAUDE report 23756727 ↗InjuryNov 25, 2024Break; Break
IT WAS REPORTED THAT, AFTER A TKA IN BOTH KNEES, THE PATIENT SUFFERED FROM BILATERAL INSTABILITY AND KNEE PAIN. THIS ADVERSE EVENT WAS SOLVED BY REVISION SURGERY OF THE LEFT SIDE ON (B)(6)2024 IN WHICH THE JOURN ART INS BCS STD 3-4 LT 9 WAS EXPLANTED DUE TO BROKEN AND WORN POST. IT WAS ALSO FOUND THAT THE PATIENT HAS PROGRESSION OF OSTEOARTHRITIS OF THE PATELLA. CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Arthritis; Failure of Implant; Joint Laxity
MAUDE report 20775590 ↗InjuryJul 12, 2024Fracture; Fracture
IT WAS REPORTED THAT AFTER A TKA THAT TOOK PLACE ON (B)(6) 2007, THE JOURN ART INS BCS STD 3-4 LT 9 BROKE. THE PART WAS REPLACED ON A REVISION SURGERY PERFORMED ON (B)(6) 2024. PATIENT STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 19731220 ↗InjuryJan 19, 2024Fracture; Fracture
IT WAS REPORTED THAT, AFTER A PRIMARY TKA WAS PERFORMED ON (B)(6) 2013, THE PATIENT EXPERIENCED AN IMPLANT FRACTURE: THE POST OF A JOURNEY ARTICULAR INSERT BCS STANDARD 3-4 LEFT 9MM BROKE OFF. THIS EVENT WAS TREATED BY PERFORMING A REVISION SURGERY ON (B)(6)2023, IN WHICH THE INSERT WAS EXCHANGED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 18544203 ↗InjuryFeb 21, 2023Fracture; Fracture
IT WAS REPORTED THAT, AFTER TKA SURGERY WAS PERFORMED ON (B)(6) 2006, THE PATIENT SUSTAINED JOINT AP INSTABILITY WITH A PEG FRACTURE OF THE JOURN ART INS BCS STD 3-4 LT 9. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2023. THE PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 16415094 ↗InjuryAug 10, 2022Fracture; Fracture
IT WAS REPORTED THAT, AFTER A TKA WAS PERFORMED ON (B)(6) 2008, PATIENT PRESENTED WITH ANTERIOR/POSTERIOR INSTABILITY. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2022 TO REPLACE THE JOURN ART INS BCS STD 3-4 LT 9 WITH A J2 CRDD OF THE SAME SIZE. UPON REVISION, THE POST OF THE JOURN ART INS BCS STD 3-4 LT 9 WAS FOUND TO BE FRACTURED. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.
Patient problems: Failure of Implant; Joint Laxity
MAUDE report 15203305 ↗InjurySep 30, 2019Fracture; Fracture
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO A BROKEN POLYETHYLENE POST.
Patient problems: Injury
MAUDE report 9135167 ↗InjurySep 6, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED A REVISION SURGERY IN WHICH THE BI-LATERAL PATELLA WAS REPLACED. AS A STANDARD THE SURGEON DOES A POLY CHANGE THEN AS WELL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7853095 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042515, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-0001-2019 · Jun 13, 2018 · Open, ClassifiedJourney BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 7
- Z-0002-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS OXINIUM Knee Femoral Components, Catalog Numbers: 74021110, 74021111, 74021112, 74021113, 74021114, 74021115, 74021116, 74021117, 74021118, 74021119, 74021120, 7402112
- Z-0003-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS Knee Conventional Polyethylene Inserts: 74023111, 74023112, 74023113, 74023114, 74023115, 74023116, 74023117, 74023118, 74023121, 74023122, 74023123, 74023124, 7402312
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page