The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010555281), received between May 21, 2018 and Dec 6, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (3 reports).
4 other recalls cover devices cleared under the same FDA submission (K042515, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Break1
- Material Fragmentation1
Patient problems
- Failure of Implant4
- Joint Dislocation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 6, 2024Fracture; Fracture
IT WAS REPORTED THAT, AFTER A TKR PERFORMED ON (B)(6) 2012, PATIENT WAS WALKING DOWNSTAIRS AND NOTICED INESTABILITY IN THE TKR. WHEN DOCTOR OPENED THE KNEE, THE JOURN ART INS BCS STD 7-8 RT10 WAS BROKEN. THIS ADVERSE EVENT WAS TREATED BY REMOVING THE BROKEN POST AND INSERTING A REV JOURNEY ART INS BCS STD 7-8 RT 9 ON AN UNKNOWN DATE. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 20873271 ↗InjuryApr 7, 2024Fracture; Fracture
IT WAS REPORTED THAT, AFTER A TKA PROCEDURE PERFORMED ON (B)(6) 2010, THE POST OF A JOURNEY ARTICULAR INSERT BCS STANDARD 7-8 RIGHT 10MM BROKE OFF. A REVISION SURGERY WAS CARRIED OUT ON (B)(6) 2024 TO ADDRESS THIS EVENT. FURTHER COMPLICATIONS WERE NOT REPORTED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 19057129 ↗InjuryJul 7, 2022Fracture; Fracture
IT WAS REPORTED THAT, AFTER A SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE PATIENT EXPERIENCED KNEE DISLOCATION, FURTHER INVESTIGATION REVEALED THE PATIENT HAD A BROKEN POLY COMPONENT. THIS ADVERSE EVENT WAS SOLVED BY REVISION SURGERY ON (B)(6) 2022. THE CURRENT HEALTH STATUS OF PATIENT IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 14952078 ↗InjuryMar 2, 2022Break; Break
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON UNKNOWN DATE, THE PATIENT EXPERIENCED A BREAK FROM THE POST. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION ON (B)(6) 2022 TO EXCHANGED AND EXPLANTED THE JOURN ART INS BCS STD 7-8 RT10 DUE TO BROKEN POST. CURRENT HEALTH STATUS OF PATIENT WAS REPORTED AS FIT AND HEALTHY.
Patient problems: Failure of Implant
MAUDE report 13654044 ↗InjuryMay 21, 2018Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE PATIENT DISLOCATED THE FEMORAL COMPONENT OVER POST IN HIGH FLEXION. THE SURGEON FOUND THE TIP OF INSERT HAD BROKEN OFF AND WAS FLOATING FREE IN KNEE. THE INSERT WAS REMOVED AND REPLACED.
Patient problems: Joint Dislocation
MAUDE report 7530751 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042515, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-0002-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS OXINIUM Knee Femoral Components, Catalog Numbers: 74021110, 74021111, 74021112, 74021113, 74021114, 74021115, 74021116, 74021117, 74021118, 74021119, 74021120, 7402112
- Z-0003-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS Knee Conventional Polyethylene Inserts: 74023111, 74023112, 74023113, 74023114, 74023115, 74023116, 74023117, 74023118, 74023121, 74023122, 74023123, 74023124, 7402312
- Z-0001-2019 · Jun 13, 2018 · Open, ClassifiedJourney BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 7
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.