JOURNEY 74023273Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY ARTICULAR INSERT BCS STANDARD 7-8 RIGHT 11MM · Class 2 · Implant
All models of JOURNEY (566)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010555298), received between Apr 3, 2019 and Feb 4, 2025 — classified as 8 injuries.
The most frequently reported specific device problem is “Fracture” (5 reports).
4 other recalls cover devices cleared under the same FDA submission (K042515, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture5
- Break2
- Adverse Event Without Identified Device or Use Problem1
Patient problems
- Failure of Implant5
- Injury2
- Pain2
- Joint Dislocation1
- Joint Laxity1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjuryFeb 4, 2025Fracture; Fracture
IT WAS REPORTED THAT, AFTER TKA SURGERY HAD BEEN PERFORMED ON 2011, THE PATIENT EXPERIENCED SEVERAL EPISODES OF GROSS KNEE INSTABILITY IN WHICH THE SURGEON DESCRIBED AS A KNEE DISLOCATION. THIS ADVERSE EVENT WAS TREATED BY REVISION SURGERY ON (B)(6) 2025, IN WHICH THE JOURN ART INS BCS STD 7-8 RT11 WAS EXCHANGED DUE TO BREAKAGE OF THE POST. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant; Joint Dislocation; Joint Laxity
MAUDE report 21304010 ↗InjuryJan 7, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT, AFTER PRIMARY TKA SURGERY HAD BEEN PERFORMED ON UNKNOWN DATE, THE PATIENT EXPERIENCED KNEE PAIN. THIS ADVERSE EVENT WAS SOLVED BY REVISION TKA SURGERY ON (B)(6) 2024, IN WHICH ONE (1) JOURN ART INS BCS STD 7-8 RT11 WAS EXPLANTED SINCE THE SURGEON WAS IN THE JOINT SPACE ALREADY. THE POLY IN QUESTION WAS NOT DAMAGED, NOR HAD ANY ISSUES PER THE SURGEON. THERE WERE NO ISSUES WITH THE S+N IMPLANTS, INFECTION OR LOOSENING, BUT THE IMPLANT HAD BEEN IN, OVER 15 YEARS. PATIENT LEFT SURGERY IN GOOD HEALTH. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Pain
MAUDE report 21096190 ↗InjuryJan 30, 2023Fracture; Fracture
IT WAS REPORTED THAT AFTER A TKA PROCEDURE HAD BEEN PERFORMED, THE PATIENT EXPERIENCED A BROKEN POST. THIS ADVERSE EVENT WAS ADDRESSED BY REVISION SURGERY ON (B)(6) 2023, TO EXPLANT THE JOURN ART INS BCS STD 7-8 RT11. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 16265140 ↗InjuryMay 10, 2021Break; Break
IT WAS REPORTED THAT, AFTER A PRIMARY TKA, A PATIENT COMPLAINED OF PAIN. A REVISION SURGERY WAS PERFORM ON (B)(6) 2021 TO EXCHANGE THE JOURNEY ARTICULAR INSERT BCS STD 7-8 RT11 DUE TO THE POST HAD BROKEN. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY. THE OUTCOME OF THE PATIENT IS UNKNOWN.
Patient problems: Failure of Implant; Pain
MAUDE report 11798606 ↗InjuryApr 19, 2021Break; Break
IT WAS REPORTED THAT, THERE WAS A REVISION SURGERY FOR A RIGHT JOURNEY BCS 7-8 11MM INSERT ON (B)(6) 2021 DUE TO A BROKEN POST.
Patient problems: Failure of Implant
MAUDE report 11687106 ↗InjuryFeb 15, 2021Fracture; Fracture
IT WAS REPORTED THAT, AFTER A TKA HAD BEEN PERFORMED, THE POST OF A JOURNEY ART INS BCS STD 7-8 RT11 BROKE. A REVISION SURGERY WAS MADE TO EXPLANT THE BROKEN IMPLANT. THE CURRENT STATUS OF THE PATIENT'S HEALTH IS UNKNOWN.
Patient problems: Failure of Implant
MAUDE report 11326534 ↗InjuryJun 12, 2020Fracture; Fracture
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO A BROKEN POST. PATIENT IS A FIT HEALTHY ACTIVE MAN. IMPLANT WAS REPLACED WITH A SIZE 7/8 11MM JOURNEY BCS INSERT. PATELLA WAS ALSO RESURFACED - SIZE 23MM JOURNEY BCS BICONVEX PATELLA. SURGEON DOES BLAME IMPLANT.
Patient problems: Injury
MAUDE report 10148992 ↗InjuryApr 3, 2019Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT WAS UNSTABLE, DURING THE REVISION THE POST WAS FOUND TO BE BROKEN. IT WAS REPLACED WITH A REVISION INSERT.
Patient problems: Injury
MAUDE report 8478886 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042515, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-0003-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS Knee Conventional Polyethylene Inserts: 74023111, 74023112, 74023113, 74023114, 74023115, 74023116, 74023117, 74023118, 74023121, 74023122, 74023123, 74023124, 7402312
- Z-0002-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS OXINIUM Knee Femoral Components, Catalog Numbers: 74021110, 74021111, 74021112, 74021113, 74021114, 74021115, 74021116, 74021117, 74021118, 74021119, 74021120, 7402112
- Z-0001-2019 · Jun 13, 2018 · Open, ClassifiedJourney BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 7
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page