JOURNEY 74023284Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY ARTICULAR INSERT BCS STANDARD 7-8 LEFT 13MM · Class 2 · Implant
All models of JOURNEY (566)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010555373), received between Jun 26, 2018 and May 21, 2021 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Break” (3 reports).
4 other recalls cover devices cleared under the same FDA submission (K042515, K230653); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break3
- Insufficient Device Problem Information2
- Fracture1
Patient problems
- Injury3
- Failure of Implant2
- Joint Dislocation1
- Joint Laxity1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryMay 21, 2021Break; Break
IT WAS REPORTED THAT, AFTER A TKA SURGERY HAD BEEN PERFORMED ON AN UNKNOWN DATE, THE PATIENT HAD A DISLOCATION AND THE JOURN ART INS BCS STD 7-8 LT13 POST WAS BROKEN. THE ADVERSE EVENT WAS ADDRESSED WITH A REVISION SURGERY ON (B)(6) 2021. THE CONDITION OF THE PATIENT IS UNKNOWN.
Patient problems: Failure of Implant; Joint Dislocation
MAUDE report 11863161 ↗InjuryNov 4, 2020Break; Fracture; Break; Fracture
IT WAS REPORTED THAT A PATIENT SHOWED INSTABILITY WAS BROUGHT TO THE OPERATING ROOM AND THE POST IN A 7-8 LEFT 15MM JOURNEY L BCS POLY WAS BROKEN OR WORN OFF . A NEW 15MM HI POST REVISION POLY WAS PUT IN.
Patient problems: Failure of Implant; Injury; Joint Laxity
MAUDE report 10789376 ↗MalfunctionJun 20, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE POST WAS FOUND BROKEN. ALL PIECES WERE REMOVED FROM THE PATIENT. NO CONFIRMATION OF DELAY HAS BEEN RECEIVED. BACKUP DEVICE AVAILABLE. NO INJURY OR IMPACT TO PATIENT REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8719986 ↗InjuryDec 18, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED. PATIENT PRESENTED WITH METALLOSIS, JOINT WAS INFLAMED, SOFT TISSUE HAD A BLACK COLOUR. THE SURGEON REVISED THE KNEE, TOOK OUT THE IMPLANTS AND REPLACED THEM.
Patient problems: Injury
MAUDE report 8175536 ↗InjuryJun 26, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED A REVISION SURGERY WAS PERFORMED. THE IMPLANT WAS REMOVED.
Patient problems: Injury
MAUDE report 7638673 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K042515, K230653. One submission often covers many models, so check the product and lot details in each record.
- Z-2621-2023 · Aug 31, 2023 · Open, Classifiedsmith&nephew LEGION OXINUM POSTERIOR STABILIZED NONPOROUS FERMOAL COMPONENT, SIZE 5N, LEFT, NARROW, REF 71421265; knee joint
- Z-0003-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS Knee Conventional Polyethylene Inserts: 74023111, 74023112, 74023113, 74023114, 74023115, 74023116, 74023117, 74023118, 74023121, 74023122, 74023123, 74023124, 7402312
- Z-0001-2019 · Jun 13, 2018 · Open, ClassifiedJourney BCS Knee CoCr Femoral Components, Catalog Numbers: 74021210, 74021211, 74021212, 74021213, 74021214, 74021215, 74021216, 74021217, 74021218, 74021219, 74021220, 74021221, 7
- Z-0002-2019 · Jun 13, 2018 · Open, ClassifiedJOURNEY BCS OXINIUM Knee Femoral Components, Catalog Numbers: 74021110, 74021111, 74021112, 74021113, 74021114, 74021115, 74021116, 74021117, 74021118, 74021119, 74021120, 7402112
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page