NA 71369716Recalls & reported problems
Smith & Nephew, Inc. · MI TRIAL FEMORAL HEAD 32MM +4 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010555625), received between Jun 19, 2019 and Jul 20, 2022 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break4
- Loose or Intermittent Connection3
- Naturally Worn3
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 20, 2022Fracture; Fracture
IT WAS REPORTED THAT ONE (1) TRIAL FEMORAL HEAD 32 M/+4 BROKE OFF AND IS UNABLE TO RECEIVE THE NECK TRIAL. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15067915 ↗MalfunctionDec 28, 2021Break; Break
IT WAS REPORTED THAT, DURING FIELD INSPECTION, IT WAS FOUND THAT THE BOTTOM PIECE OF A MI TRIAL FEM HEAD 36 +0, A MI TRIAL FEM HEAD 32 -3 AND A MI TRIAL FEM HEAD 32 +4 IS BROKEN SO THE NECK TRIALS CANNOT STAY ATTACHED TO THEM (CASE: (B)(4)). NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13094878 ↗MalfunctionApr 18, 2021Loose or Intermittent Connection; Naturally Worn; Loose or Intermittent Connection; Naturally Worn
IT WAS REPORTED THAT, THREE 32 +4 TRIAL HEADS ARE WORN AND WILL NOT ARTICULATE WITH THE FEMORAL STEM SECURELY. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11683513 ↗MalfunctionApr 18, 2021Loose or Intermittent Connection; Naturally Worn; Loose or Intermittent Connection; Naturally Worn
IT WAS REPORTED THAT, THREE 32 +4 TRIAL HEADS ARE WORN AND WILL NOT ARTICULATE WITH THE FEMORAL STEM SECURELY. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11683510 ↗MalfunctionApr 18, 2021Loose or Intermittent Connection; Naturally Worn; Loose or Intermittent Connection; Naturally Worn
IT WAS REPORTED THAT, THREE 32 +4 TRIAL HEADS ARE WORN AND WILL NOT ARTICULATE WITH THE FEMORAL STEM SECURELY. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11683509 ↗MalfunctionNov 28, 2019Break; Break
IT WAS REPORTED THAT DURING SURGERY, A TAB IN THE TRIAL HEAD SNAPPED OFF, AND IT WOULD NOT STAY ON THE STEM. THERE WERE NO DELAYS IN THE PROCEDURE AND A BACKUP DEVICE WAS AVAILABLE. NO PATIENT INJURIES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9390629 ↗MalfunctionJul 7, 2019Break; Break
IT WAS REPORTED THAT DURING A THR PROCEDURE, PLASTIC PIECE ON BASE OF FEMORAL TRIAL SNAPPED OFF WHEN TRIALING HIP. ALL PIECES REMOVED AND RECOVERED FROM PATIENT. NO INJURY OR IMPACT TO PATIENT. NO DELAY. BACKUP DEVICE AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8766049 ↗MalfunctionJun 19, 2019Break; Break
IT WAS REPORTED THAT THE ITEM BROKE DURING IMPACTION. NO INJURY REPORTED. SURGERY COMPLETED WITH A S&N BACKUP DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8713161 ↗
FDA clearance and classification
Cleared through 510(k) K121393 on Aug 7, 2012 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew, Inc. Total Knee System Instrumentation”. The FDA review took 90 days. FDA record ↗
The FDA regulates product code HRY (prosthesis, knee, femorotibial, semi-constrained, cemented, metal/polymer) under 21 CFR 888.3530 as a Class II device. 21 CFR 888.3530 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Common questions about the NA 71369716
Has the NA 71369716 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010555625), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the NA 71369716?
The FDA's MAUDE database lists 8 adverse event reports for this device, received between Jun 19, 2019 and Jul 20, 2022. In those reports, the most common specific device problems are “Break” (4), “Loose or Intermittent Connection” (3) and “Naturally Worn” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 71369716?
None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 71369716 FDA approved?
It was cleared through 510(k) K121393 on Aug 7, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 71369716, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand NA, model 71369716). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page