Skip to content

JOURNEY 74012418Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK SIZE 8 · Class 2

Quick summary

This Smith & Nephew Journey cutting block is a guide used during knee replacement surgery. Of its 90 reports, 11 involve injuries and no deaths are recorded. Reports mostly describe the block breaking or fracturing during surgery, with surgeons finishing the procedure using the same device or a backup. There were 4 reports in the last 12 months, down from 10 in the year before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
90
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 90 adverse event reports that list this exact device (DI 03596010556707), received between Feb 5, 2019 and Apr 28, 2026 — classified as 11 injuries, 79 malfunctions.

The most frequently reported specific device problem is “Break” (49 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 4 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break49
  • Fracture28
  • Detachment of Device or Device Component6
  • Loose or Intermittent Connection4
  • Material Fragmentation2
  • Material Separation2
  • Component Missing1
  • Connection Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions81
  • No Known Impact Or Consequence To Patient5
  • No Consequences Or Impact To Patient3
  • No Patient Involvement2
  • Insufficient Information1

Recent adverse event reports

The 25 most recent of 90 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 28, 2026Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE. ALL THE BROKEN PIECES WERE RETRIEVED FORM THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE AS THE FEMUR CUTS WERE ALREADY DONE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25017231 ↗
  • MalfunctionApr 14, 2026Break; Fracture

    IT WAS REPORTED THAT, DURING A TKA, A JOURNEY DCF AP FEM CUT BLK 8 WAS BROKEN. IT IS UNKNOWN HOW WAS THE PROCEDURE COMPLETED OR IF THERE WAS ANY SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24873592 ↗
  • MalfunctionFeb 26, 2026Fracture

    IT WAS REPORTED THAT, DURING A LEFT TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 FRACTURED AT THE FRONT WHILE BEING RETRIEVED USING A SLAP HAMMER FOLLOWING COMPLETION OF THE 4-IN-1 FEMORAL CUTS. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24454472 ↗
  • MalfunctionNov 4, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY, THE REMOVAL CAPTURE OF ONE (1) JOURNEY DCF AP FEM CUT BLK 8 WAS FOUND TO BE BROKEN. NO PIECES WERE LEFT INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED WITHOUT ANY SURGICAL DELAY USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23461665 ↗
  • MalfunctionSep 30, 2025Break; Fracture; Break; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, IT WAS NOTICED THAT ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE DURING IMPACTION. SLAP HAMMER DOES NOT WORK WITH BLOCK ANYMORE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23182052 ↗
  • MalfunctionSep 17, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, IT WAS NOTICED THAT ONE (1) JOURNEY DCF AP FEM CUT BLK 8 WAS DAMAGED DURING IMPACTING, IT FRACTURED. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY SURGICAL DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23073041 ↗
  • MalfunctionSep 9, 2025Detachment of Device or Device Component; Mechanics Altered; Detachment of Device or Device Component; Mechanics Altered

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE BOTTOMS SCREW AND SPRING OF ONE (1) JOURNEY DCF AP FEM CUT BLK 8 FELL LOOSE. NO METAL WAS LEFT IN THE PATIENT. THE PROCEDURE WAS RESUMED, WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23004514 ↗
  • MalfunctionAug 4, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, UPON REMOVING ONE (1) JOURNEY DCF AP FEM CUT BLK 8 WITH THE SLIDE HAMMER, THE WINGLETS ON THE CUTTING BLOCK THAT THE SLIDE HAMMER GRABS BROKE OFF. IT IS NOW UNABLE TO BE REMOVED VIA SLIDE HAMMER BECAUSE THE ATTACHMENT POINTS ON THE CUTTING BLOCK ARE GONE. ALL PIECES WERE ACCOUNTED FOR AND THIS INCIDENT DID NOT AFFECT THE SURGERY. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22710628 ↗
  • MalfunctionMay 20, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE IMPACTOR KNOB OF ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE OFF DURING IMPACTION. IT IS UNKNOWN IF ANY PIECE FELL INTO THE INCISION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22053783 ↗
  • MalfunctionApr 11, 2025Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 WAS CHIPPED AFTER IMPACT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21827505 ↗
  • MalfunctionMar 25, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA, THE CENTER CLEAT OF A JOURNEY DCF AP FEM CUT BLK 8 BROKE DURING REMOVAL. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21688245 ↗
  • MalfunctionMar 6, 2025Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 DISASSEMBLED DURING EXTRACTION FROM KNEE FOLLOWING CUT. THE PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21543037 ↗
  • MalfunctionMar 5, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING TKA SURGERY, THE PART OF THE JOURNEY DCF AP FEM CUT BLK 8 THAT SLAP HAMMER ATTACHES TO BROKE OFF WHILE SLAPPING THE FEMUR. IT DID NOT FALL INTO THE WOUND AND THE PATIENT WASN'T HARMED AND CASE WASN'T AFFECTED. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21531854 ↗
  • MalfunctionFeb 26, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE OFF NEAR THE KNOB. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21469251 ↗
  • MalfunctionJul 17, 2024Fracture; Fracture

    IT WAS REPORTED THAT, ON IMPACTION DURING TKA SURGERY, A PIECE OF A JOURNEY DCF AP FEM CUT BLK 8 AROUND THE DIAL, BROKE OFF. ALL PIECES WERE RETRIEVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19768338 ↗
  • MalfunctionMay 27, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, IT WAS NOTICED THAT, THE METAL FROM A JOURNEY DCF AP FEM CUT BLK 8 BROKE OFF DURING MALLEATING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19401019 ↗
  • MalfunctionMay 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE INTO TWO PIECES. ALL PIECES WERE REMOVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19376846 ↗
  • MalfunctionMay 15, 2024Break; Break

    IT WAS REPORTED THAT, DURING A TKA, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19322519 ↗
  • MalfunctionMay 14, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY, USING A JOURNEY DCF AP FEM CUT BLK 8, THE EXTRACTION RIM UNDER THE ADJUSTMENT DIAL BROKE DURING EXTRACTION. THE CASE WAS COMPLETED SUCCESSFULLY AS ALL CUTS WERE COMPLETED. ALL PIECES WERE REMOVED AND ACCOUNTED. IT CAME OUT WHEN THE BLOCK WAS REMOVED AND SEPARATED ON THE BACK TABLE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19310917 ↗
  • MalfunctionDec 19, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, IT WAS NOTICED THAT, METAL PIECE AROUND DIAL OF ONE (1) JOURNEY DCF AP FEM CUT BLK 8 WAS BROKEN OFF. NOTHING FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18357347 ↗
  • MalfunctionNov 18, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, THE METAL PIECE THAT GOES AROUND THE DIAL OF A JOURNEY DCF AP FEM CUT BLK 8 IS BROKEN OFF, ANY PIECE FELL INSIDE THE WOUND AND WAS AND REMOVED BY HAND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18166931 ↗
  • MalfunctionSep 4, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA, THE JOURNEY DCF AP FEM CUT BLK 8 BROKE WHERE EXTRACTOR MAKES CONTACT. NO PIECES FELL INTO PATIENT. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING THE SAME DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17675172 ↗
  • MalfunctionAug 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING TKA SURGERY, SURGEON IMPACTED A JOURNEY DCF AP FEM CUT BLK 8 ON THE DISTAL FEMUR AND OBSERVED THAT A SPRING CAME OUT LANDING ON THE TIBIA, THE SPRING WAS REMOVED AND RETAINED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17538079 ↗
  • MalfunctionJul 13, 2023Fracture; Fracture

    IT WAS REPORTED THAT, UPON REMOVAL ON A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 HAD THE METAL AROUND THE CENTER MECHANISM BROKE OFF. ALL PIECES WERE FOUND AND ACCOUNTED FOR. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17322342 ↗
  • MalfunctionJul 11, 2023Break; Break

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 8 BROKE WHILE IMPACTING IT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17302938 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.