JOURNEY 74012419Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK SIZE 9 · Class 2
All models of JOURNEY (566)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Journey femoral cutting block is used during knee replacement surgery to guide the saw cuts on the femur. The 38 reports mostly describe metal parts of the block breaking, cracking or chipping off during surgery, sometimes leaving small pieces near or in the patient;7 reports involved injuries and none involved deaths. Most surgeries were completed without delay, often with the same or a backup device. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 38
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 03596010556714), received between Feb 13, 2019 and Dec 11, 2025 — classified as 7 injuries, 31 malfunctions.
The most frequently reported specific device problem is “Break” (18 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break18
- Fracture13
- Crack2
- Detachment of Device or Device Component2
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Defective Device1
- Loose or Intermittent Connection1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Consequences Or Impact To Patient2
- No Information2
- No Known Impact Or Consequence To Patient2
- Foreign Body In Patient1
- Injury1
Recent adverse event reports
The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2025Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A PRIMARY TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 9 HAD A LOOSE PIECE THAT MAY FALL OFF COMPLETELY. THE PROCEDURE WAS RESUMED WITHOUT ANY DELAY USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23783306 ↗MalfunctionAug 4, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY DCF AP FEM CUT BLK 9 HAD A PIECE OF METAL FRACTURE OFF WHEN CUTTING. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22710600 ↗MalfunctionMay 11, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING TKA SURGERY WHILE CUTTING THROUGH THE CUTTING SLOTS, A SECTION OF ONE (1) JOURNEY DCF AP FEM CUT BLK 9 FELL OFF. IT WAS EASILY RETRIEVED OUTSIDE OF THE PATIENT, AND NO ADDITIONAL PIECES BROKE OFF. REMAINDER OF NECESSARY CUTS WERE EASILY COMPLETED WITHOUT THIS PIECE. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21999974 ↗MalfunctionJan 7, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING TKA SURGERY, WHILE CUTTING THE CHAMFERS WITH A JOURNEY DCF AP FEM CUT BLK 9, A METAL PIECE FELL OUT OF THE BLOCK. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21091981 ↗MalfunctionSep 2, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, THE JOURNEY DCF AP FEM CUT BLK 9 BROKE STRAIGHT THROUGH THE NUMBER 2 SLOT WHEN IMPACTING. IT IS UNKNOWN AT THIS TIME IF PIECES FELL INTO PATIENT. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20118489 ↗MalfunctionJul 17, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE ANTERIOR CUT SLOT OF A JOURNEY DCF AP FEM CUT BLK 9 SNAPPED OFF WHEN REMOVING SAW BLADE. BROKEN PIECE WAS REMOVED BY HAND, NO PIECES REMAINED IN PATIENT. THE PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19759299 ↗MalfunctionMay 29, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING TKA SURGERY, IT WAS NOTICED THAT, THE SMALL PIECE OF THE JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF DURING IMPACTING. NO PIECES FELL INSIDE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19420790 ↗MalfunctionFeb 22, 2024Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA, ONE (1) JOURNEY DCF AP FEM CUT BLK 9 BROKE, THE IMPACTION PART OF THE BLOCK CHIPPED OFF. THE PIECE WAS RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18763714 ↗MalfunctionFeb 12, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, A PIECE OF THE CHAMFER ON THE JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF AND FELL ON THE FLOOR. NO PIECES FELL INTO PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18692499 ↗MalfunctionAug 24, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE CHAMFER PART IN ONE (1) JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17614483 ↗MalfunctionAug 3, 2023Crack; Fracture; Crack; Fracture
IT WAS REPORTED THAT DURING A TKA, ONE (1) JOURNEY DCF AP FEM CUT BLK 9 WAS FOUND CRACKED AND CHIPPED. IT NO LONGER WORKS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17450432 ↗MalfunctionFeb 23, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, PART OF ONE (1) JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF WHILE IMPACTING IT ON THE DISTAL FEMUR. IT FELL ON THE TABLE AND WAS EASILY RETRIEVED. NO PARTS FELL IN THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16432471 ↗MalfunctionJul 23, 2022Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, THE SPEED PIN ON A JOURNEY FEMORAL CUTTING BLOCK SIZE 9 GOT STUCK IN THE PIN HOLE, CAUSING BURN MARKS AND DEBRIS. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15090397 ↗MalfunctionJul 13, 2022Break; Break
IT WAS REPORTED THAT, DURING A TKA PROCEDURE, A JOURNEY DCF AP FEM CUT BLK 9 BROKE. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15008827 ↗MalfunctionMar 26, 2022Break; Break
IT WAS REPORTED THAT A JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK SIZE 9 IS BROKEN. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13925098 ↗MalfunctionMar 1, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT A SCREW OF A JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK SIZE 9 WAS FOUND TO BE MISSING. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13636658 ↗MalfunctionJan 10, 2022Break; Break
IT WAS REPORTED THAT THE METAL PIECE OF A JOURNEY FEMORAL CUTTING BLOCK SIZE 9 BROKE OFF. AS THIS HAPPENED IN A BIOSKILLS LABORATORY/MARKETING DEMONSTRATION, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13214455 ↗InjuryJan 6, 2022Break; Break
IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, A PIECE OF BLOCK CRACKED OFF INSIDE THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13180946 ↗InjuryOct 19, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF WHILE CUTTING. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12662630 ↗MalfunctionOct 11, 2021Break; Break
IT WAS REPORTED THAT THE METAL PIECE OF A JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK SIZE 9 BROKE OFF. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605550 ↗InjurySep 28, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA, A PART OF A JOURNEY DCF AP FEM CUT BLK 9 BROKE OFF. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12539800 ↗MalfunctionAug 25, 2021Mechanics Altered; Mechanics Altered
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE JOURNEY DCF AP FEM CUT BLK 9 CAME OUT OF BLOCK AND NEEDS TO BE REPLACED. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. NO HEALTH CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12367685 ↗MalfunctionJul 20, 2021Fracture; Defective Device; Fracture; Defective Device
IT WAS REPORTED THAT THE JOURNEY DCF AP FEM CUT BLK 9 IS NOT PASSING CENTRAL PROCESSING DEPARTMENT STANDARDS. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED. THE INVESTIGATION FOUND THAT THE FLANGES ON THE ATTACHMENT POST FRACTURED OFF THE DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12196795 ↗MalfunctionJun 11, 2021Crack; Connection Problem; Crack; Connection Problem
IT WAS REPORTED THAT, DURING SURGERY, A JOURNEY DCF AP FEM CUT BLK 9 WAS CLOSED DOWN AND HAD SMALL CRACKS ON BOTH SIDES OF THE BLOCK, THIS CAUSED THAT THE CUTS WERE NOT PERFECT AND THE FEMORAL TRIAL WOULD NOT FIT ON PROPERLY. SURGERY WAS RESUMED, WITHOUT ANY SIGNIFICANT DELAY AND IT WAS NECESSARY TO CHANGE THE SURGICAL TECHNIQUE; THEREFORE, PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11988123 ↗InjuryMay 13, 2021Break; Break
IT WAS REPORTED THAT DURING A TKA SURGERY, THE JOURNEY DISTAL CUT FIRST ANTERIOR POSTERIOR FEMORAL CUTTING BLOCK BROKE, WHEN IT WAS BEING IMPACTED ON TO THE FEMUR INSIDE THE PATIENT. THE BROKEN PIECE WAS RECOVERED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED, WITHOUT DELAY USING A SMITH AND NEPHEW BACK-UP DEVICE. DESPITE THE FACT, THIS INCIDENT OCCURRED INSIDE THE PATIENT, NO HARM, INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11816097 ↗
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page