JOURNEY 74012831Recalls & reported problems

Smith & Nephew, Inc. · JOURNEY FEMORAL IMPLANT IMPACTOR · Class 2

Quick summary

This Smith & Nephew instrument, the Journey femoral implant impactor, is used to hammer a knee implant into place during total knee replacement surgery. Most of the 62 reports describe the tool breaking or cracking during the final hammering step, sometimes with the head coming off, though in the described cases the pieces were retrieved and surgery continued without delay. There are 4 reports of injury and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
62
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 62 adverse event reports that list this exact device (DI 03596010556905), received between Jun 9, 2019 and Jun 12, 2026 — classified as 4 injuries, 58 malfunctions.

The most frequently reported specific device problem is “Break” (28 reports).

4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break28
  • Fracture14
  • Loose or Intermittent Connection7
  • Crack2
  • Defective Device2
  • Detachment of Device or Device Component2
  • Material Fragmentation2
  • Material Separation2

Patient problems

  • No Clinical Signs, Symptoms or Conditions48
  • No Consequences Or Impact To Patient5
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient3
  • Foreign Body In Patient2
  • No Patient Involvement2
  • No Information1

Recent adverse event reports

The 25 most recent of 62 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2026Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) JOURNEY FEM IMPL IMPACTOR WAS BROKEN UPON THE FINAL IMPACTION AND ALL THE PIECES WERE RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25484860 ↗
  • MalfunctionMay 11, 2026Break

    IT WAS REPORTED THAT, DURING A TKR SURGERY, ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE DURING IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25145369 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT DURING A TKA SURGERY, THE JOURNEY FEM IMPL IMPACTOR BROKE UPON IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24722351 ↗
  • MalfunctionOct 31, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, A JOURNEY FEM IMPL IMPACTOR BROKE UPON IMPACTION. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23437881 ↗
  • MalfunctionJul 18, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE HEAD OF ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE OFF WHEN IMPACTING THE FEMUR ONTO THE PATIENT. ALL BROKEN PIECES WERE RETRIEVED BY TWEEZERS. THE PROCEDURE WAS COMPLETED AFTER A NON-SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22543520 ↗
  • MalfunctionDec 30, 2024Fracture; Detachment of Device or Device Component; Fracture; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A TKA PROCEDURE, ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE UPON IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21034894 ↗
  • MalfunctionNov 13, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, WHEN IMPACTING FEMORAL IMPLANT, THE HEAD OF THE JOURNEY FEM IMPL IMPACTOR BROKE OFF. NOTHING FELL INSIDE THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20678463 ↗
  • MalfunctionNov 4, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE OFF DURING INTENDED USE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO PATIENT HARM WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20605848 ↗
  • MalfunctionOct 24, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE HEAD OF ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE OFF. NOTHING FELL INTO THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20526917 ↗
  • MalfunctionAug 22, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, THE HEAD OF A JOURNEY FEM IMPL IMPACTOR CAME OFF. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20026857 ↗
  • MalfunctionApr 2, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING WHEN IMPACTING THE FEMORAL COMPONENT IN A TKA SURGERY, THE HEAD OF ONE (1) JOURNEY FEM IMPL IMPACTOR DETACHED FROM THE HANDLE WHILE IMPACTING. SCREW WAS PRESENT BUT BECAME LOOSE ON IMPACTING LEADING TO ULTIMATE FAILURE. IT WAS REMOVED WITH FORCEPS. SUCTION WAS USED TO ENSURE ANY POSSIBLE BROKEN PIECES WERE REMOVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19021156 ↗
  • MalfunctionFeb 16, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING TKA SURGERY, THE TIP OF A JOURNEY FEM IMPL IMPACTOR BROKE OFF DURING FEMUR IMPACTION, NOTHING FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18719237 ↗
  • MalfunctionFeb 5, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE OFF WHILE THE SURGEON WAS IMPACTING THE FEMUR IMPLANT. IT IS UNKNOW IF ANY PIECE FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18645940 ↗
  • MalfunctionAug 3, 2023Break; Break

    IT WAS REPORTED THAT DURING SURGERY, ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE. IT IS UNKNOWN HOW WAS COMPLETED THE PROCEDURE AND IF DELAY WAS PRESENTED. NO INJURY TO PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17450972 ↗
  • MalfunctionMay 23, 2023Solder Joint Fracture; Solder Joint Fracture

    IT WAS REPORTED THAT, DURING A TKA SURGERY, A JOURNEY FEM IMPL IMPACTOR SNAPPED AT THE NECK JUNCTION CLOSE TO THE BUMPER. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED DUE TO THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16989312 ↗
  • MalfunctionMay 9, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA SURGERY A JOURNEY FEM IMPL IMPACTOR WAS FRACTURED DURING IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. THE PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16895031 ↗
  • MalfunctionMar 20, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING A TKA, THE GREY PIECE AND SCREW BROKE OFF ONE (1) JOURNEY FEM IMPL IMPACTOR BROKE ON IMPACTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16579969 ↗
  • MalfunctionOct 4, 2022Break; Break

    IT WAS REPORTED THAT THE JOURNEY FEM IMPL IMPACTOR IS BROKEN. IT IS UNKNOWN IF THIS WAS NOTICED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED. UPON INVESTIGATION, IT WAS FOUND THAT THE HEAD OF THE IMPACTOR HAS BROKEN FROM THE REST OF THE DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 15535444 ↗
  • MalfunctionSep 30, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING TKA, THE JOURNEY FEM IMPL IMPACTOR SNAPPED AT THE NECK. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR SIGNIFICANT DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15514888 ↗
  • MalfunctionSep 27, 2022Fracture; Fracture

    IT WAS REPORTED THAT, A JOURNEY FEM IMPL IMPACTOR BROKE IN HALF. THE CONTEXT IN WHICH THIS OBSERVATION WAS MADE IS UNSPECIFIED. THEREFORE, PATIENT INVOLVEMENT IS YET TO BE CONFIRMED.

    Patient problems: Insufficient Information

    MAUDE report 15491718 ↗
  • MalfunctionMay 4, 2022Break; Break

    IT WAS REPORTED THAT THE JOURNEY FEM IMPL IMPACTOR WAS FOUND TO BE BROKEN. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14279543 ↗
  • MalfunctionApr 9, 2022Break; Break

    IT WAS REPORTED THAT, THE JOURNEY FEM IMPL IMPACTOR BROKE IN HALF. IT IS UNKNOWN WHETHER THIS PROBLEM WAS NOTICED IN A SURGICAL ENVIRONMENT OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14068430 ↗
  • MalfunctionApr 1, 2022Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT LITTLE PIECES OF PLASTIC OF A JOURNEY FEM IMPL IMPACTOR ARE STARTING TO COME OFF. IT IS UNKNOWN WHETHER THE EVENT HAPPENED RELATIVE TO THE SURGERY AND IF THERE WAS A PATIENT INVOLVEMENT.

    Patient problems: Insufficient Information

    MAUDE report 13975757 ↗
  • MalfunctionMar 29, 2022Break; Break

    IT WAS REPORTED THAT, DURING TKA SURGERY, A JOURNEY FEM IMPL IMPACTOR BROKE AT CASING. DEVICE OUTSIDE THE PATIENT AND NO PIECES FELL INSIDE. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13935779 ↗
  • MalfunctionMar 23, 2022Break; Break

    IT WAS REPORTED THAT THE GREY PLASTIC OF A JOURNEY FEM IMPL IMPACTOR IS CHIPPING OFF. AS THIS WAS NOTICED DURING SET UP FOR A TKA, THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT IMPACTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13870341 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.

  • Z-0878-2020 · Dec 16, 2019 · Open, Classified
    LEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
  • Z-0200-2019 · Sep 18, 2018 · Terminated
    Smith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
  • Z-2163-2018 · May 4, 2018 · Terminated
    smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
  • Z-1902-2018 · Apr 9, 2018 · Terminated
    smith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a

Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.