JOURNEY 74018911Recalls & reported problems
Smith & Nephew, Inc. · JOURNEY ARTICULAR INSERT ASSEMBLY TOOL · Class 2
All models of JOURNEY (565)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew JOURNEY articulating insert assembly tool is used during total knee replacement surgery. The 79 reports mostly describe the tool breaking or failing to connect during surgery, with the procedure usually continuing on another device; 77 malfunctions, 2 injuries, and no deaths are mentioned. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 79
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 79 adverse event reports that list this exact device (DI 03596010557292), received between Feb 19, 2019 and Aug 6, 2026 — classified as 2 injuries, 77 malfunctions.
The most frequently reported specific device problem is “Break” (44 reports).
4 other recalls cover devices cleared under the same FDA submission (K121393); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break44
- Connection Problem8
- Fracture8
- Incomplete or Inadequate Connection4
- Unstable4
- Detachment of Device or Device Component3
- Loose or Intermittent Connection3
- Mechanical Jam2
Patient problems
- No Clinical Signs, Symptoms or Conditions54
- No Consequences Or Impact To Patient11
- No Known Impact Or Consequence To Patient9
- No Patient Involvement7
- Injury1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 79 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 6, 2026Break
It was reported that, during a TKA surgery, the JOURNEY ART ISRT ASSM TOOL was broken. The procedure was resumed, without any delay, using the same device. No further complications were reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50160813 ↗MalfunctionJun 25, 2026Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, A PIECE OF ONE (1) JOURNEY ART ISRT ASSM TOOL BROKE OFF WHILE TRYING TO INSERT THE POLY INTO THE FEMUR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25626110 ↗MalfunctionNov 12, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE RESISTANCE SPRING OF ONE (1) JOURNEY ART ISRT ASSM TOOL BROKE OFF ON ONE SIDE. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23539149 ↗MalfunctionJun 12, 2025Positioning Failure; Detachment of Device or Device Component; Positioning Failure; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING TKA SURGERY, A JOURNEY ART ISRT ASSM TOOL WOULD NOT ENGAGE THE INSERT AS DESIGNED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE. DURING INVESTIGATION IT WAS FOUND THAT THE 0.1858" X .290"DOWEL PIN APPEARS TO BE LOOSE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22199051 ↗MalfunctionSep 24, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE SPRING OF THE JOURNEY ART ISRT ASSM TOOL BROKE, IT WONT CLAMP TO ENGAGE POLY AND THE HANDLE WONT COMPRESS. NO PIECES FELL INTO PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20296156 ↗MalfunctionDec 6, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE JOURNEY ART ISRT ASSM TOOL BROKE OUT OF PATIENT IN THE PROCESS OF INSERT PLACEMENT. THE PROCEDURE WAS RESUMED, AFTER NO DELAY, BUT IT IS UNKNOWN HOW. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18275266 ↗MalfunctionOct 4, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING INSERTION, WHILE PERFORMING A TKA, THE SPRING ON A JOURNEY ART ISRT ASSM TOOL BROKE. HOWEVER, ALL PIECES WERE RECOVERED. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15537938 ↗MalfunctionOct 4, 2022Fracture; Fracture
IT WAS REPORTED THAT, AFTER A TKA SURGERY, WHILE WASHING A JOURNEY ART ISRT ASSM TOOL THE DEVICE BROKE AND THE PIECE THAT PROVIDES TENSION GOT DETACHED. SINCE THE INCIDENT OCCURRED AFTER THE PROCEDURE, THE PATIENT WAS NO LONGER INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15535463 ↗MalfunctionJun 29, 2022Fracture; Fracture
IT WAS REPORTED THAT THE JOURNEY ART ISRT ASSM TOOL BROKE. AS THIS WAS NOTICED UPON FIELD INSPECTION, NO PATIENT WAS INVOLVED. UPON INVESTIGATION, IT WAS FOUND THAT THE THE BASEPLATE HOOK (AT THE TIP) IS CHIPPED ON THE END.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14861064 ↗MalfunctionMay 23, 2022Break; Break
IT WAS REPORTED THAT THE POLY GUN OF A JOURNEY ARTICULAR INSERT ASSEMBLY TOOL WAS FOUND BROKEN ON THE INNER TENSION PIECE. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14481168 ↗MalfunctionMay 14, 2022Break; Corroded; Detachment of Device or Device Component; Break; Corroded; Detachment of Device or Device Component
IT WAS REPORTED THAT, ONE OF THE SPRING LOADED TABS OF THE JOURNEY ART ISRT ASSM TOOL DETACHED AND THE RIVET IS RUSTED. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14403007 ↗MalfunctionOct 11, 2021Connection Problem; Mechanics Altered; Connection Problem; Mechanics Altered
IT WAS REPORTED THAT, A JOURNEY ARTICULAR INSERT ASSEMBLY TOOL WILL NOT GRAB, IT DOESN'T WORK PROPERLY. THE CONTEXT IN WHICH THIS OBSERVATION WAS MADE IS UNSPECIFIED, FOR WHICH PATIENT INVOLVEMENT IS YET TO BE CONFIRMED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12605507 ↗MalfunctionJun 14, 2021Break; Component Missing; Break; Component Missing
IT WAS REPORTED THAT A R3 OFFSET IMPACTOR HAD A PART MISSING. NO FURTHER DETAILS WERE GIVEN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11991681 ↗MalfunctionJun 4, 2021Mechanical Problem; Connection Problem; Mechanical Problem; Connection Problem
IT WAS REPORTED THAT, DURING SURGERY, IT WAS NOTICED THAT A JOURNEY FEM IMPACT BUMPER LT ((B)(4)) HAD A PIECE MISSING A LUG ON BOTTOM AND CRACKED. THE GROOVES THAT SLIDE ONTO THE JRNY II TKA FEM TRIAL SZ 3 LT ((B)(4)) WERE STRIPPED AND IT DID NOT LOAD PROPERLY, AND ALSO A JOURNEY ART ISRT ASSM TOOL ((B)(4)) HAD NO TENSION AND DID NOT WORK TO SEAT POLY. DEVICES OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK UP DEVICES. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11942601 ↗MalfunctionJun 4, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY WHILE THE INSTRUMENT WAS OUTSIDE OF THE PATIENT, THE JOURNEY ART ISRT ASSM TOOL WAS BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY AND IT WAS UNKNOWN HOW THE PROCEDURE WAS COMPLETED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11942467 ↗MalfunctionJun 3, 2021Break; Break
IT WAS REPORTED THAT, DURING STERILIZATION PROCESS, IT WAS FOUND THAT THE INNER METAL CLASP IN THE HANDLE OF THE JOURNEY ART ISRT ASSM TOOL WAS BROKE. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11934740 ↗MalfunctionMay 27, 2021Break; Break
IT WAS REPORTED THAT, THREE (3) JOURNEY ART ISRT ASSM TOOL HAVE BROKEN SPRINGS. NO CASE INVOLVED; THEREFORE, NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11894116 ↗MalfunctionMay 27, 2021Break; Break
IT WAS REPORTED THAT, THREE (3) JOURNEY ART ISRT ASSM TOOL HAVE BROKEN SPRINGS. NO CASE INVOLVED; THEREFORE, NO PATIENT INJURIES OR OTHER COMPLICATIONS WERE REPORTED. NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11894098 ↗MalfunctionMay 20, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING TKA SURGERY, THE JOURNEY ART INSERT ASSM TOOL DID NOT ENGAGE POLY. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11860170 ↗MalfunctionMay 20, 2021Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE BARS THAT ACTS AS SPRING MECHANISM ON A JOURNEY ART ISRT ASSM TOOL BROKE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED, WITHOUT DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11860154 ↗MalfunctionMay 13, 2021Break; Break
IT WAS REPORTED THAT, A JOURNEY ART ISRT ASSM TOOL WAS FOUND BROKEN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821616 ↗MalfunctionMay 13, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT, DURING A TKA, THE JRNY II BCS XLPE ART ISRT SZ 1-2 RT 9MM ((B)(4)) COULD NOT BE INSERTED WITH THE JOURNEY ART ISRT ASSM TOOL ((B)(4)). INSTRUMENT INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING BOTH S+N BACK-UP DEVICES. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821468 ↗MalfunctionMay 13, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT DURING TKA, THE JOURNEY ART ISRT ASSM TOOL WAS NOT WORKING PROPERLY, SINCE IT DID NOT SNAP THE INSERT IN PLACE ONTO THE TIBIAL BASEPLATE. NO SURGICAL DELAY WAS REPORTED. IT IS UNKNOWN IF THE PROCEDURE WAS FINISHED WITH THE SAME DEVICE OR A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11817255 ↗MalfunctionMay 10, 2021Break; Break
IT WAS REPORTED THAT, A JOURNEY ART ISRT ASSM TOOL IS BROKEN. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11803356 ↗MalfunctionMay 10, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT, DURING A TKA, A JOURNEY ART ISRT ASSM TOOL DID NOT WORK PROPERLY TO HELP SEAT POLY INTO BASEPLATE. INSTRUMENT OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11798554 ↗
FDA clearance and classification
Cleared through 510(k) K121393 on Aug 7, 2012 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew, Inc. Total Knee System Instrumentation”. The FDA review took 90 days. FDA record ↗
The FDA regulates product code HRY (prosthesis, knee, femorotibial, semi-constrained, cemented, metal/polymer) under 21 CFR 888.3530 as a Class II device. 21 CFR 888.3530 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K121393. One submission often covers many models, so check the product and lot details in each record.
- Z-0878-2020 · Dec 16, 2019 · Open, ClassifiedLEGION AP Cutting Block Size 4, REF 71434409; LEGION AP Cutting Block Size 6, REF 71434411; LEGION AP Cutting Block Size 7, REF 71434412; LEGION AP Cutting Block Size 8, REF 714344
- Z-0200-2019 · Sep 18, 2018 · TerminatedSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to mak
- Z-2163-2018 · May 4, 2018 · Terminatedsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
- Z-1902-2018 · Apr 9, 2018 · Terminatedsmith&nephew 5o (5 degree) GENESIS(R) II VALGUS BUSHING, REF CATALOG NUMBER 71440014 The GENESIS II Knee System Components are indicated for rheumatoid arthritis; post-traumatic a
Common questions about the JOURNEY 74018911
Has the JOURNEY 74018911 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010557292), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the JOURNEY 74018911?
The FDA's MAUDE database lists 79 adverse event reports for this device, received between Feb 19, 2019 and Aug 6, 2026. In those reports, the most common specific device problems are “Break” (44), “Connection Problem” (8) and “Fracture” (8); the most common patient problem is “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the JOURNEY 74018911?
None of the reports are classified as deaths, 2 as injuries and 77 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the JOURNEY 74018911 FDA approved?
It was cleared through 510(k) K121393 on Aug 7, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the JOURNEY 74018911, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand JOURNEY, model 74018911). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- R3 71363077Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER461 reports
- NA 71364009Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER174 reports
- JOURNEY 74012416Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012415Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER163 reports
- JOURNEY 74012414Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER129 reports
- GENESIS II 71440194Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER114 reports
- GENESIS II 71440842Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER113 reports
- JOURNEY 74018901Smith & Nephew, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER106 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page