TRIGEN 71674074Recalls & reported problems
Smith & Nephew, Inc. · 3.2MM T-HANDLE TROCAR · Class 2
All models of TRIGEN (380)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010560704), received between Jun 16, 2020 and Mar 23, 2023 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break4
- Fracture1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Foreign Body In Patient1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 23, 2023Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE DOCTOR FOUND IT NECESSARY TO HAMMER A HONEYCOMB, IN VIEW OF CALCIFICATION IN THE MEDULLARY CANAL, AND TO HELP THE CANNULA ENTRY, BUT THE "T" OF THE HONEYCOMB BROKE. THE BROKEN PARTS WERE NOT REMOVED, AS SURGEON'S REQUEST. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 16607575 ↗MalfunctionApr 27, 2022Fracture; Fracture
IT WAS REPORTED THAT, A 3.2MM T-HANDLE TROCAR WAS SNAPPED IN HALF. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14225471 ↗MalfunctionNov 11, 2021Break; Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY, THE 3.2MM T-HANDLE TROCAR BROKE. INCIDENT OCCURRED WHILE INSTRUMENT WAS OUTSIDE THE PATIENT AND NO PIECES FELL INTO THE INCISION. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12800979 ↗MalfunctionApr 9, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE HANDLE BROKE OF FROM THE BODY OF THE T-HANDLE TROCAR, WHILE WAS BEING USED OUTSIDE THE PATIENT. THE PROCEDURE WAS FINISHED USING A BACKUP DEVICE FROM SMITH AND NEPHEW, WITHOUT SURGICAL DELAY AND NO INJURY TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11642772 ↗MalfunctionNov 6, 2020Break; Break
IT WAS REPORTED THAT DURING INSPECTION PROCESS, THE 3.2MM T-HANDLE TROCAR WAS FOUND INSIDE THE TRAY BROKEN AT THE BASE OF THE HANDLE. THERE WAS NO CASE INVOLVED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10804834 ↗MalfunctionJun 16, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT, DURING CLEANING PROCESS, HANDLE WELD CAME APART AND SEPARATED. THE SHAFT CAME APART FROM THE HANDLE. NO ISSUES DURING CASE SINCE THE EVENT OCCURRED AFTER PROCEDURE. THEREFORE, NO DELAY AND NO BACKUP WAS NEEDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10160242 ↗
FDA clearance and classification
Cleared through 510(k) K120871 on Jun 7, 2012 (decision: Substantially Equivalent). Submitted by Smith & Nephew, Inc. as “Smith & Nephew, Inc. External Fixation System Instrumentation”. The FDA review took 77 days. FDA record ↗
The FDA regulates product code JDW (pin, fixation, threaded) under 21 CFR 888.3040 as a Class II device. 21 CFR 888.3040 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the TRIGEN 71674074
Has the TRIGEN 71674074 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03596010560704), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the TRIGEN 71674074?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Jun 16, 2020 and Mar 23, 2023. In those reports, the most common specific device problems are “Break” (4), “Fracture” (1) and “Material Fragmentation” (1); the most common patient problem is “Foreign Body In Patient” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the TRIGEN 71674074?
None of the reports are classified as deaths, 0 as injuries and 6 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the TRIGEN 71674074 FDA approved?
It was cleared through 510(k) K120871 on Jun 7, 2012. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the TRIGEN 71674074, and is it still sold?
The FDA's device database lists Smith & Nephew, Inc. as the company responsible for it (brand TRIGEN, model 71674074). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other PIN, FIXATION, THREADED devices
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall2 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- ORTHOFIX MODULSYSTEM 1Orthofix Srl · Pin, Fixation, Threaded1 recall0 reports
- APEX 5038-1-080Stryker GmbH · PIN, FIXATION, THREADED48 reports
- Agee Digit Widget® DWD-232Hand Biomechanics Lab Inc · Pin, Fixation, Threaded32 reports
- SPATIAL FRAME 71070720Smith & Nephew, Inc. · PIN, FIXATION, THREADED15 reports
- APEX 5038-2-110Stryker GmbH · PIN, FIXATION, THREADED11 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page