EMPERION 71290902Recalls & reported problems
Smith & Nephew, Inc. · EMPERION STEM 9 STANDARD OFFSET 140MM · Class 2 · Implant
All models of EMPERION (14)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010561275), received between Jun 21, 2017 and Aug 27, 2024 — classified as 5 injuries.
The most frequently reported specific device problem is “Fracture” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture4
- Insufficient Device Problem Information1
- Loosening of Implant Not Related to Bone-Ingrowth1
Patient problems
- Injury3
- Failure of Implant2
- Bone Fracture(s)1
- Deformity/ Disfigurement1
- Implant Pain1
- Loss of Range of Motion1
- Pain1
- Radiolucent Lines1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryAug 27, 2024Fracture; Fracture
IT WAS REPORTED THAT AFTER UNDERGOING A TOTAL HIP REPLACEMENT (THR) IN 2014 DUE TO UNKNOWN REASONS PATIENT SUFFERED BREAKAGE OF AN EMPERION STEM. THIS ADVERSE EVENT WAS ADDRESSED BY CONDUCTING A REVISION SURGERY ON (B)(6) 2023, DURING WHICH THE EMP STEM 9 SO 140MM, HEAD AND LINER WERE EXCHANGED. DURING PROCEDURE, THEY ENCOUNTERED SIGNIFICANT SCLEROSIS ON THE DISTAL ASPECT OF THE SLEEVE, SO A DECISION WAS MADE TO START PREPARING A TROCHANTERIC OSTEOTOMY. HOWEVER, THERE HAD ALREADY BEEN A SMALL CRACK IN THE LATERAL ASPECT OF THE GREATER TROCHANTER. PATIENT WAS ON TOE TOUCH WEIGHTBEARING. SHE HAD STRICT ABDUCTOR PRECAUTIONS FOR SIX TO TWELVE WEEKS WHICH DEPENDED ON THE FOUR WEEK POSTOPERATIVE X-RAY. IN ADDITION, SEVEN WEEKS POST-OP PATIENT SHOWED ALIGNMENT AND APPEARANCE OF THE HARDWARE UNCHANGED AND WITHOUT ACUTE OSSEOUS COMPLICATION.
Patient problems: Failure of Implant; Radiolucent Lines
MAUDE report 20076040 ↗InjuryApr 26, 2022Fracture; Loosening of Implant Not Related to Bone-Ingrowth; Fracture; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT, AFTER A RIGHT THR ON (B)(6) 2015 DUE TO RIGHT HIP OSTEOARTHRITIS WITH LIMITED RANGE OF MOTION, THE PATIENT EXPERIENCED PAIN, GROIN TENDERNESS, LIMITED INTERNAL AND EXTERNAL ROTATION. CT SCAN SHOWED PROSTHETIC FRACTURE OF STEM AND X-RAYS CONFIRMED BROKEN STEM DISPLACED, FRACTURED, AND ANGULATED AT THE LEVEL OF THE GREATER TROCHANTER/INTERTROCHANTERIC REGION. THE PATIENT UNDERWENT REVISION SURGERY ON RIGHT HIP ON (B)(6) 2021. IT IS UNKNOWN THE STATUS OF THE PATIENT.
Patient problems: Deformity/ Disfigurement; Failure of Implant; Implant Pain; Loss of Range of Motion; Pain
MAUDE report 14212753 ↗InjuryDec 31, 2018Fracture; Fracture
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO PERIPROSTHETIC FRACTURE/IMPLANT FAILURE. DUE TO WHERE THE FRACTURE OF THE PROSTHESIS WAS, COULDN'T USE EMPERION REMOVAL INSTRUMENTS AND REMOVAL OF STEM WAS DIFFICULT AS THE DISTAL STEM WAS WELL FIXED. THEN CUT BASE OF SLEEVE HOWEVER WAS UNABLE TO REMOVE IT. THEN TRIED THE TREPHINES OVER THE STEM AND STILL UNABLE TO FREE THE STEM. SO AN EXTENDED TROCHANTERIC OSTEOTOMY WAS PERFORMED, STRUT GRAFT AND THE LONGEST ACCORD TROCHANTERIC GRIP PLATE WAS USED OVER THE RESTORATION HIP SYSTEM WITH 9 CABLES IN TOTAL. THE LINER WAS NOT REPLACED.
Patient problems: Injury
MAUDE report 8206177 ↗InjuryAug 31, 2017Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE REVISION WAS PERFORMED DUE TO BREAKAGE OF THE PROXIMAL EMPERION STEM.
Patient problems: Bone Fracture(s); Injury
MAUDE report 6833607 ↗InjuryJun 21, 2017Fracture; Fracture
IT WAS REPORTED IN A PUBLICATION THAT A PATIENT HAD A REVISION DUE TO FRACTURE OF THE STEM. NO FURTHER INFORMATION WAS AVAILABLE.
Patient problems: Injury
MAUDE report 6656103 ↗
Other Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate devices
- Novation 130-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls156 reports
- Novation 130-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls79 reports
- CORAIL 3L92511Depuy (Ireland) · Prosthesis, hip, semi-constrained, uncemented, metal/polymer, non-porous, calicum-phosphate158 reports
- Acumatch A Series 138-36-29Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate3 recalls1 report
- Novation 136-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls43 reports
- Novation 130-32-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls33 reports
- Novation 132-32-51Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls26 reports
- Novation 136-36-53Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Non-Porous, Calicum-Phosphate2 recalls22 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page