Skip to content

ANTHOLOGY 71356006Recalls & reported problems

Smith & Nephew, Inc. · ANTHOLOGY STANDARD OFFSET POROUS SIZE 6 · Class 3 · Implant

All models of ANTHOLOGY (69)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010567765), received between Jun 9, 2020 and Oct 24, 2024 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Biocompatibility” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Biocompatibility1
  • Delivered as Unsterile Product1
  • Unstable1

Patient problems

  • Infiltration into Tissue1
  • Injury1
  • Joint Disorder1
  • No Clinical Signs, Symptoms or Conditions1
  • Synovitis1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 24, 2024Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT, DURING THR SURGERY, IT WAS NOTICED THAT THE BOX OF ONE (1) ANTHOLOGY SO POROUS SIZE 6 HAD A HOLE IN IT, AND DID NOT KNOW IF IT WAS STILL STERILE. A BACK UP STEM WAS AVAILABLE IN THE ROOM, THEN THEY WERE FINE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20530014 ↗
  • InjuryMay 19, 2022Biocompatibility; Biocompatibility

    IT WAS REPORTED THAT, AFTER A LEFT BHR-THA WAS PERFORMED ON (B)(6) 2010 DUE TO DEGENERATIVE ARTHRITIS, THE PATIENT EXPERIENCED INCREASING COBALT AND CHROMIUM LEVELS. AN MRI SHOWED FLUID COLLECTION WITH SYNOVIAL THICKENING, CONSISTENT WITH ADVERSE LOCAL TISSUE REACTION. THIS ADVERSE EVENT WAS TREATED WITH A LEFT REVISION SURGERY ON (B)(6) 2021. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Infiltration into Tissue; Synovitis

    MAUDE report 14438332 ↗
  • InjuryJun 9, 2020Unstable; Unstable

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED ON THE LEFT HIP DUE TO FAIL THA, WITH GROSSLY LOOSE AND UNSTABLE FEMORAL COMPONENT.

    Patient problems: Injury; Joint Disorder

    MAUDE report 10133521 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page