Skip to content

ANTHOLOGY 71356106Recalls & reported problems

Smith & Nephew, Inc. · ANTHOLOGY HIGH OFFSET POROUS SIZE 6 · Class 3 · Implant

All models of ANTHOLOGY (69)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010567888), received between May 21, 2019 and Feb 12, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information2
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1

Patient problems

  • Failure of Implant2
  • High metal ion levels1
  • Injury1
  • Unspecified Infection1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 12, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, DURING A FOLLOW-UP OF A PATIENT WHO HAD A REVISION SURGERY DUE TO ELEVATED CHROMIUM-COBALT ION LEVELS, IT WAS FOUND THAT, THEY TESTED POSITIVE FOR TITANIUM IN THEIR BLOOD. AND THE ANTHOLOGY HO POROUS SZ 6 THAT WAS IMPLANTED ON (B)(6) 2010 WAS ALSO FOUND TO HAVE A DENT. FURTHER INFORMATION ON HOW THIS ADVERSE EVENT WILL BE RESOLVED HAS NOT BEEN MADE AVAILABLE.

    Patient problems: Failure of Implant; High metal ion levels

    MAUDE report 18692493 ↗
  • InjuryMay 4, 2021Fracture; Fracture

    IT WAS REPORTED THAT, AFTER A THA SURGERY HAD BEEN PERFORMED ON (B)(6) 2009, THE NECK OF AN ANTHOLOGY HO POROUS SZ 6 STEM FRACTURED. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021 TO EXPLANT THE STEM. THE PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Failure of Implant

    MAUDE report 11773667 ↗
  • InjuryDec 4, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED DUE TO A BILATERAL PULMONARY EMBOLISM. THE EVENT WAS TREATED WITH MEDICATION THERAPY AND THE PATIENT WAS DISCHARGED THE NEXT DAY.

    Patient problems: Injury

    MAUDE report 9413561 ↗
  • InjuryMay 21, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT RIGHT HIP REVISION SURGERY WAS PERFORMED DUE TO SEPTIC THA STATUS AND INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 8628224 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Ceramic/Metal, Cemented Or Uncemented devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page