Skip to content

BIRMINGHAM HIP RESURFACING (BHR) 90128275Recalls & reported problems

Smith & Nephew, Inc. · OFFSET CUP INTRODUCER · Class 3

All models of BIRMINGHAM HIP RESURFACING (BHR) (47)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010591531), received between Dec 31, 2019 and Nov 6, 2023 — classified as 1 injury, 6 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

29 other recalls cover devices cleared under the same FDA submission (P040033); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Adverse Event Without Identified Device or Use Problem1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Known Impact Or Consequence To Patient2
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 6, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR, THE THREADED ROD IN ONE (1) OFFSET CUP INTRODUCER BROKE DURING IMPACTION OF THE BHR SHELL. NO PIECES WERE NOTICED BY THE SURGEON IN THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18075632 ↗
  • MalfunctionMar 10, 2022Break; Break

    IT WAS REPORTED THAT WHILE THE FINAL IMPACTION WAS TAKING PLACE DURING THR, THE OFFSET CUP INTRODUCER BROKE OUTSIDE THE PATIENT. IT IS UNKNOWN HOW THIS PROCEDURE FINISHED. NO INJURY TO PATIENT OR DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13722574 ↗
  • MalfunctionJan 6, 2022Break; Break

    IT WAS REPORTED THAT, DURING BHR SURGERY, AN OFFSET CUP INTRODUCER BROKE OFF. IT IS UNKNOWN IF A SURGICAL DELAY OCCURRED DUE TO THIS INCIDENT, AND WHETHER A BACK UP INTRODUCER WAS AVAILABLE OR NOT TO CONCLUDE THE PROCEDURE. IT IS UNKNOWN IF THERE WAS A PATIENT HARM AS A CONSEQUENCE OF THIS FAILURE.

    Patient problems: Insufficient Information

    MAUDE report 13176922 ↗
  • MalfunctionMar 12, 2021Break; Break

    IT WAS REPORTED THAT, THE OFFSET CUP INTRODUCER WAS BROKEN. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11468414 ↗
  • MalfunctionFeb 4, 2021Break; Break

    IT WAS REPORTED THAT THE ITEM WAS FOUND BROKEN WHILE OUTSIDE THE PATIENT. NO DELAY REPORTED. NO PATIENT INJURIES REPORTED. AN S&N BACKUP WAS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11282126 ↗
  • InjuryOct 26, 2020Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING THR SURGERY, THE THREADED BOLT OF THE LOCKING MECHANISM OF THE BHR OFFSET CUP IMPACTOR BROKE INSIDE THE PATIENT DURING IMPACTION OF THE CUP. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK UP DEVICE, WITH NO SURGICAL DELAY AND NO INJURY TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10733063 ↗
  • MalfunctionDec 31, 2019Break

    IT WAS REPORTED THAT DURING IMPACT OF THE BHR CUP, THE INSTRUMENT BROKE DUE TO FATIGUE FROM MULTIPLE BLOWS WITH A MALLET. BACKUP DEVICE WAS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9535238 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P040033. One submission often covers many models, so check the product and lot details in each record.

  • Z-2242-2021 · Jul 14, 2021 · Terminated
    smith&nephew BHR SQUARE HEADED NAIL, REF CATALOG NUMBER 999908. For use during orthopedic surgery.
  • Z-2268-2018 · May 11, 2018 · Terminated
    Birmingham Hip Resurfacing Acetabular Cup with Impactor: (a) Size 56 mm, REF 74120156 (b) Size 64 mm, REF 74120164 (c) Size 54 mm, REF 74122154 (d) Size 58 mm, REF 74122158 (e) Siz
  • Z-2267-2018 · May 11, 2018 · Terminated
    Birmingham Hip Resurfacing Femoral Head: (a) Size 58 mm, REF 74121158 (b) Size 56 mm, REF 74123156
  • Z-2745-2015 · Jun 3, 2015 · Terminated
    Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING FEMORAL HEAD, STERILE R, Rx only, REF Numbers: 74121138 (38 MM), 74123140 (40 MM), 74121142 (42 MM), 74123144 (44 MM), 74121146 (46 M
  • Z-2747-2015 · Jun 3, 2015 · Terminated
    Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING DYSPLASIA CUP W/IMPACTOR, REF Numbers: 74120246 (46 MM), 74122248 (48 MM), 74120250 (50 MM), 74122252 (52 MM), 74120254 (54 MM) Pro
  • Z-2746-2015 · Jun 3, 2015 · Terminated
    Smith & Nephew BIRMINGHAM HIP (TM) RESURFACING ACETABULAR CUP W/IMPACTOR, REF Numbers: 74120144 (44 MM), 74120146 (46 MM), 74122146 (46 MM), 74122148 (48 MM), 74120148 (48 MM), 74
  • Z-0023-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 56 mm cup Ref.: 74120156. Hip implant component.
  • Z-0017-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 44 mm cup Ref.: 74120144. Hip implant component.
  • Z-0029-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Dysplasia Cup, 50 mm cup Ref.: 74120250. Hip implant component.
  • Z-0019-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 48 mm cup Ref.: 74120148. Hip implant component.
  • Z-0032-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 58 mm cup Ref.: 74120258. Hip implant component.
  • Z-0033-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 62 mm cup Ref.: 74120262. Hip implant component.
  • Z-0022-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 54 mm cup Ref.: 74120154. Hip implant component.
  • Z-0024-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup, 58 mm cup Ref.: 74120158. Hip implant component.
  • Z-0016-2008 · Jul 12, 2007 · Terminated
    Smith + Nephew, Birmingham Hip Resurfacing, Acetabular Cup 46 mm cup Ref.: 74120146. Hip implant component.

…and 14 more.

Other Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing devices

All Prosthesis, Hip, Semi-Constrained, Metal/Metal, Resurfacing devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page