R3 71332760Recalls & reported problems
Smith & Nephew, Inc. · R3 0 DEGREE XLPE ACETABULAR LINER 36MM INNER DIAMETER X OUTER DIAMETER 60MM · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010598028), received between Nov 13, 2018 and Jan 9, 2026 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (K113848, K211176); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product2
- Adverse Event Without Identified Device or Use Problem1
- Patient Device Interaction Problem1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Unspecified Infection2
- Injury1
- Joint Dislocation1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 9, 2026Delivered as Unsterile Product
IT WAS REPORTED THAT, DURING A THR SURGERY, A PACKAGING ISSUE WAS IDENTIFIED WITH ONE (1) R3 0 DEG XLPE ACET LNR 36MM X 60MM. THE OUTER BOX WAS CELLOPHANE-WRAPPED; HOWEVER, INSIDE THE BOX, THE DOUBLE BLISTER CONTAINING THE DEVICE WAS FOUND TO BE OPEN. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK- UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24028642 ↗MalfunctionMar 3, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT DURING SET UP OR INSPECTION, WHEN OPENED THE ONE (1) R3 0 DEG XLPE ACET LNR 36MM X 60MM BOX, IT WAS NOTICED A RIP IN THE PACKAGING ALL THE WAY TO THE LINER. THE LINER WAS NOT STERILE. THE PROCEDURE WAS COMPLETED AFTER A NON-SIGNIFICANT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21503761 ↗InjuryAug 17, 2021Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT, AFTER A THR SURGERY HAD BEEN PERFORMED ON (B)(6) 2016 THE PATIENT EXPERIENCED INFECTION. A REVISION SURGERY WAS PERFORMED ON (B)(6) 2021. THE PATIENT HAD A REVISION OF THE RIGHT HIP VIA POSTERIOR APPROACH WITH WOUND DEBRIDED AND IRRIGATED. WOUND SAMPLES SENT TO PATHOLOGY. FEMORAL HEAD REMOVED WITH BONE PUNCH, AND LINER REMOVED WITH LINER REMOVAL TOOL. WOUND IRRIGATED WITH BETADINE. NEW LINER AND HEAD OF THE SAME PREVIOUS SIZING INSERTED TO RETAINED SHELL AND STEM AS PER STANDARD TECHNIQUE WITH LINER IMPACTOR AND FEMORAL HEAD IMPACTOR. DRAINS INSERTED AND WOUND CLOSED IN LAYERS AS PER STANDARD TECHNIQUE.
Patient problems: Unspecified Infection
MAUDE report 12331685 ↗InjuryApr 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT RIGHT HIP REVISION SURGERY WAS PERFORMED DUE TO RECURRENT DISLOCATIONS.
Patient problems: Joint Dislocation
MAUDE report 8485195 ↗InjuryNov 13, 2018Scratched Material; Scratched Material
IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED DUE TO INFECTION. THE SURFACE OF REMOVED LINER AND OX HEAD WAS DAMAGED AND FOUND THAT THERE HAVE BEEN LOT OF SCRATCHES. NO DELAY OR BACK UP REPORTED.
Patient problems: Injury; Unspecified Infection
MAUDE report 8065062 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K113848, K211176. One submission often covers many models, so check the product and lot details in each record.
- Z-0326-2025 · Oct 15, 2024 · Open, ClassifiedTandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
- Z-1929-2013 · May 29, 2013 · TerminatedR3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 60 MM OD, QTY: (1), STERILE H2O2, REF 71335760. Orthopaedic.
- Z-1928-2013 · May 29, 2013 · TerminatedR3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 62 MM OD, QTY: (1), STERILE H2O2, REF 71335762. Orthopaedic.
Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices
- NA 1365-51-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented753 reports
- NA 1365-50-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented412 reports
- ARTICUL/EZE 1365-21-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 recall280 reports
- ARTICUL/EZE 1365-11-000Depuy (Ireland) · Prosthesis, hip, semi-constrained, metal/polymer, cemented1 death262 reports
- LFIT 6260-9-236Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, cemented256 reports
- PINNACLE GRIPTION 1217-32-052Depuy Orthopaedics, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented2 deaths239 reports
- ELEOS™ 25002111EOnkos Surgical, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, cemented3 deaths230 reports
- Novation 130-36-52Exactech, Inc. · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented2 recalls124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.