Skip to content

R3 71335758Recalls & reported problems

Smith & Nephew, Inc. · R3 20 DEGREE XLPE ACETABULAR LINER 36MM INNER DIAMETER X OUTER DIAMETER 58MM · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03596010598363), received between Aug 10, 2018 and Jan 8, 2025 — classified as 10 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (2 reports).

3 other recalls cover devices cleared under the same FDA submission (K113848, K211176); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem5
  • Device Dislodged or Dislocated2
  • Insufficient Device Problem Information2
  • Appropriate Device Problem Term/Code Not Available1
  • Incomplete or Inadequate Connection1

Patient problems

  • Joint Dislocation3
  • Pain2
  • Post Operative Wound Infection2
  • Bone Fracture(s)1
  • Failure of Implant1
  • Inadequate Osseointegration1
  • Injury1
  • Muscle Weakness/Atrophy1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A THR SURGERY WAS PERFORMED ON (B)(6) 2024, THE PATIENT EXPERIENCED A DISLOCATION. THIS ADVERSE EVENT WAS ADDRESSED BY A REVISION SURGERY ON (B)(6) 2024, IN WHICH THEY REPLACED ONE (1) R3 20 DEG XLPE ACET LNR 36MM X 58MM AND ONE (1) OXINIUM FEM HD 12/14 36 MM +0 WITH A 40X58 POLY AND SZ40 +8 HEAD. THE CURRENT HEALTH STATUS OF THE PATIENT IS UNKNOWN.

    Patient problems: Joint Dislocation

    MAUDE report 21107967 ↗
  • InjurySep 25, 2024Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT AFTER A THA SURGERY HAD BEEN PERFORMED ON (B)(6) 2023, THE PATIENT EXPERIENCED RECURRENT DISLOCATIONS BETWEEN THE FEMORAL HEAD AND THE POLY LINER. THIS ADVERSE EVENT WAS ADDRESSED BY A REVISION SURGERY ON (B)(6) 2024, IN WHICH A R3 20 DEG XLPE ACET LNR 36MM X 58MM AND A COCR 12/14 FEM HEAD 36MM +0 WERE EXPLANTED AND EXCHANGED BY A CONSTRAINED LINER TO ASSIST. THE PATIENT WAS DISCHARGED FROM THE HOSPITAL.

    Patient problems: Joint Dislocation

    MAUDE report 20302172 ↗
  • InjuryJan 4, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER A THR SURGERY WAS PERFORMED ON (B)(6) 2022, THE PATIENT EXPERIENCED DISLOCATION AND MUSCLE WEAKNESS. THIS ADVERSE EVENT WAS TREATED BY A REVISION SURGERY ON (B)(6) 2022, IN WHICH A R3 20 DEG XLPE ACET LNR 36MM X 5 WAS EXPLANTED AND A CONSTRAINED LINER IMPLANTED. PATIENT'S CURRENT HEALTH STATUS IS UNKNOWN.

    Patient problems: Joint Dislocation; Muscle Weakness/Atrophy

    MAUDE report 16098357 ↗
  • InjuryApr 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER THA A REVISION SURGERY WAS NECESSARY AND THE DEVICES IMPLANTED WERE REMOVED IN ORDER TO IMPLANT NEW ONES AT A DIFFERENT ANGLE. THE PATIENT WAS NOT INJURED.

    Patient problems: Failure of Implant

    MAUDE report 11722556 ↗
  • InjuryFeb 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT, AFTER THE SECOND SURGERY DUE TO THE FRACTURE ON (B)(6) 2020 THE PATIENT GOT INFECTED THEREFORE IT WAS DONE A REVISION SURGERY ON (B)(6) 2021. ALL IMPLANTS WERE REMOVED. ( OXINIUM FEM HD 12/14 36 MM -3, REDAPT SLVLS MONO STEM 190MM SZ 19 SO, R3 3 HOLE ACET SHELL 58MM, R3 20 DEG XLPE ACET LNR 36MM X 58MM, TWO REF SPHER HEAD SCREW 20MM, THREE ACC 2.0MM COCR CABLE W/CLAMP.) THOSE IMPLANTS WERE REPLACED, EXCEPTED FOR THE FLEXIBLE OSTEOTOMES USED TO REMOVE THE REDAPT SLVLS MONO STEM 190MM SZ 19 SO AND ALSO WAS IMPLANTED FOUR CC 2.0MM COCR CABLE W/CLAMP. PATIENT OUTCOME IS UNKNOWN.

    Patient problems: Post Operative Wound Infection

    MAUDE report 11263355 ↗
  • MalfunctionDec 30, 2020Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING A THR SURGERY ,THE LINER DID NOT SIT DOWN ALL THE WAY. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A SMITH & NEPHEW BACK-UP DEVICE. NO OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11096990 ↗
  • InjuryJul 3, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD A REVISION SURGERY OF PRIMARY THR (B)(6) 2020. IT WAS INFORMED THAT THE PATIENT HAD ANOTHER REVISION SURGERY IN (B)(6) 2020, BECAUSE IT PRESENTED A PERIPROSTHETIC FRACTURING FEMUR. THE SURGEON REMOVED THE FEMORAL PROSTHESIS AND THE FEMORAL HEAD. ALSO, IT WAS KNOWN THAT THE STEM WAS SUBSIDED. THE SURGEON FIXED THE PROBLEM WITH THREE CABLES. THE SURGEON DID NOT BLAME THE DEVICES IN THE DEVELOPMENT OF THE PROBLEM. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: Bone Fracture(s); Inadequate Osseointegration

    MAUDE report 10230703 ↗
  • InjuryDec 30, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT WAS RE ADMITTED ON THE HOSPITAL DUE TO PAIN, HE WAS TREATED WITH MEDICATION.

    Patient problems: Pain

    MAUDE report 9530978 ↗
  • InjuryDec 30, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT WAS RE ADMITTED ON THE HOSPITAL DUE TO PAIN, HE WAS TREATED WITH MEDICATION.

    Patient problems: Pain

    MAUDE report 9530870 ↗
  • InjuryMar 25, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED ON THE PATIENT DUE TO DISLOCATION. NO ADDITIONAL DETAILS OF THE REVISION SURGERY REPORTED. THREE EXPLANTS REPORTED. NO DELAY OR ADDITIONAL INJURIES REPORTED. A BACK UP DEVICE WAS AVAILABLE.

    Patient problems: Injury

    MAUDE report 8448781 ↗
  • InjuryAug 10, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED AT FIRST STAGE REVISION LEFT HIP, WHERE THE PATIENT HAD INFECTION. A REVISION HIP WAS DONE ORIGINALLY ON THE (B)(6) 2017 AFTER A PERIPROSTHETIC FRACTURE.

    Patient problems: Post Operative Wound Infection

    MAUDE report 7773325 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K113848, K211176. One submission often covers many models, so check the product and lot details in each record.

  • Z-0326-2025 · Oct 15, 2024 · Open, Classified
    Tandem Intl Bipolar Cocr Shell/UHMWPE Liner, TI-6AL-4V Retaining Ring, 28MM I.D. 46 MM O.D., REF 71324046, sterile, femoral head bipolar component of the Tandem Bipolar Hip System.
  • Z-1928-2013 · May 29, 2013 · Terminated
    R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 62 MM OD, QTY: (1), STERILE H2O2, REF 71335762. Orthopaedic.
  • Z-1929-2013 · May 29, 2013 · Terminated
    R3 (TM) XLPE ACETABULAR LINER UHMWPE, 20o OVERHANG, 36 MM ID, 60 MM OD, QTY: (1), STERILE H2O2, REF 71335760. Orthopaedic.

Other Prosthesis, hip, semi-constrained, metal/polymer, cemented devices

All Prosthesis, hip, semi-constrained, metal/polymer, cemented devices →

Sources and updates