Jet-X 71067322Recalls & reported problems
Smith & Nephew, Inc. · JET-X 10MM RATCHET WRENCH · Class 2
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010607768), received between Feb 28, 2021 and Feb 17, 2022 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K093047); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
- Material Deformation2
- Mechanical Jam1
- Naturally Worn1
- Physical Resistance/Sticking1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 17, 2022Break; Naturally Worn; Break; Naturally Worn
IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, A JET-X 10MM RATCHET WRENCH WAS OBSERVED TO BE BROKEN AND TO NO LONGER WORK DUE TO WEAR AND TEAR. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13549229 ↗MalfunctionJun 2, 2021Physical Resistance/Sticking; Physical Resistance/Sticking
IT WAS REPORTED THAT A 10MM RATCHET WRENCH WON'T RATCHET. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11918388 ↗MalfunctionMay 19, 2021Material Deformation; Mechanical Jam; Material Deformation; Mechanical Jam
IT WAS REPORTED THAT DURING STERILIZATION PROCESS, IT WAS NOTICED THAT THE 10MM RATCHET WRENCH SIDE WOULD NOT RATCHET. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11850308 ↗MalfunctionApr 11, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT, DURING A TRAUMA SURGERY, IT WAS FOUND THAT THE RATCHET HAS BURRS AND NEEDS TO BE REPLACED. INSTRUMENTS WERE OUTSIDE THE PATIENT. THERE WAS NO DELAY AND PROCEDURE WAS COMPLETED WITH A BACKUP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11646843 ↗MalfunctionFeb 28, 2021Break; Break
IT WAS REPORTED THAT DURING SET UP INSPECTION PROCESS WAS IDENTIFIED THAT A GEAR OF THE 10MM RATCHET WRENCH WAS BROKEN. WHEN TIGHTENING THE GOLD SIDE DID SKIP A GEAR TOOTH INSIDE THE RATCHETING MECHANISM. THERE WAS NO PATIENT INVOLVED WHEN THIS ISSUE WAS NOTICED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11390985 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093047. One submission often covers many models, so check the product and lot details in each record.
- Z-0245-2022 · Jun 17, 2021 · Open, ClassifiedIlizarov Wire Tensioner, Model 71070341, UDI (10)08033201842072(10)21HP02550(11)210831(21E434
Other APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT devices
- PRECICE® Plating System LP4.5-DLF31Nuvasive Specialized Orthopedics, Inc. · Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component3 recalls0 reports
- R3 71364451Smith & Nephew, Inc. · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT145 reports
- DFS HA BONE SCREWS HAA60-15040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15050Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAB60-12040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-13040Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-18030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
- DFS HA BONE SCREWS HAA60-15030Biomet Orthopedics, LLC · APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT2 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.