Medicrea International · PEDICLE SCREWDRIVER SHAFT · Class 1

All models of PASS LP (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03613720160175), received between May 1, 2024 and Feb 16, 2026 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 16, 2026Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL SCREW RE MOVAL PROCEDURE AT THE L4 LEVEL. IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE WHILE BEING USED TO FACILITATE THE REMOVAL OF AN IMPLANTED SCREW. THE INSTRUMENT WAS APPLIED TO THE TOP OF THE SCREW WITH DOWNWARD PRESSURE TO BACK THE SCREW OUT, WHICH REQUIRED A GREAT DEAL OF FORCE, AND THE TIP BROKE AS A RESULT. NO FRAGMENTS REMAINED IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24342665 ↗
  • MalfunctionFeb 11, 2026Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL SCREW REMOVAL PROCEDURE AT THE L4 LEVEL. IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE WHILE BEING USED TO FACILITATE THE REMOVAL OF AN IMPLANTED SCREW. THE INSTRUMENT WAS APPLIED TO THE TOP OF THE SCREW WITH DOWNWARD PRESSURE TO BACK THE SCREW OUT, WHICH REQUIRED A GREAT DEAL OF FORCE, AND THE TIP BROKE AS A RESULT. NO FRAGMENTS REMAINED IN THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24312876 ↗
  • MalfunctionJan 19, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR REMOVAL OF PEDICLE SCREWS. IT WAS REPORTED THAT, WHEN ATTEMPTING TO REMOVE IMPLANTED SCREWS, THE DISTAL TIP TORQUED AND RIPPED OFF. THE TIPS OF THE DRIVERS WERE BROKEN. THERE WAS NO FRAGMENT OF BROKEN INSTRUMENT REMAINS IN PATIENT BODY. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED AS A RESULT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21187838 ↗
  • MalfunctionJan 19, 2025Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR REMOVAL OF PEDICLE SCREWS. IT WAS REPORTED THAT, WHEN ATTEMPTING TO REMOVE IMPLANTED SCREWS, THE DISTAL TIP TORQUED AND RIPPED OFF. THE TIPS OF THE DRIVERS WERE BROKEN. THERE WAS NO FRAGMENT OF BROKEN INSTRUMENT REMAINS IN PATIENT BODY. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS REPORTED. NO ADDITIONAL TREATMENT OR SURGERY PERFORMED AS A RESULT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21187836 ↗
  • MalfunctionMay 21, 2024Material Deformation; Material Deformation

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REP ORTED THAT THE T20 SCREW NEAR BARREL, ATTACHING DRIVER PIECES TOGETHER IS STUCK/STRIPPED. IT CANNOT BE SCREWED BACK INTO TO STABILIZE THE DRIVER PIECES. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19356884 ↗
  • MalfunctionMay 1, 2024Break; Break

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. PRE-OP DIAGNOSIS OF PATIENT WAS PROXIMAL JUNCTIONAL KYPHOSIS. IT WAS REPORTED THAT THE DEVICE WAS BROKEN. THE PEDICLE SCREWDRIVER WAS BEING USED TO EXPLANT PREVIOUSLY IMPLANTED SCREWS BY ENGAGING THE BASE OF THE SCREW, AND WITH DOWNWARD PRESSURE, BACK THE SCREWS OUT. WHEN ATTEMPTING TO ENGAGE AND REMOVE THE SCREW, THE DISTAL TIP OF THE DRIVER SHEARED OFF. A SECOND DRIVER WAS SUBSEQUENTLY USED TO COMPLETE THE EXPLANTATION.  THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219296 ↗

Other Screwdriver devices

All Screwdriver devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page