Skip to content

IB3D™ PL Spinal System A24000000Recalls & reported problems

Medicrea International · INSERTER A24000000 IB3D IMPLANT · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 03613720286677), received between Nov 24, 2025 and Jan 16, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Mechanics Altered” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K241164); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 9 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 29, 2025Class IIOpen, Classified

    Due to complaints, orthopedic rotation tool handle may untread from the shaft resulting in the inability to rotate implant.

    Z-1038-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanics Altered5
  • Material Twisted/Bent3
  • Mechanical Problem3
  • Break2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 16, 2026Mechanics Altered

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT, THE INSERTERS WERE TWISTING. THERE WAS NO PATIENT INVOLVEMENT REPORTED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24091143 ↗
  • MalfunctionJan 16, 2026Mechanics Altered

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REP ORTED THAT, THE INSERTERS WERE TWISTING. THERE WAS NO PATIENT INVOLVEMENT REPORTED. NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24091128 ↗
  • MalfunctionNov 24, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR TRANSFORAMINAL LUMBAR INTE RBODY FUSION. IT WAS REPORTED THAT THE INSTRUMENT HANDLES WERE TWISTING DURING ROUTINE CHECKING IN THE STERILE PROCESSING DEPARTMENT, AND THE INSTRUMENT BROKE. THERE WAS NO PATIENT INVOLVED. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629499 ↗
  • MalfunctionNov 24, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3-4 TRANSFORAM INAL LUMBAR INTERBODY FUSION (TLIF) FOR AN INDICATION OF SPINAL STENOSIS. IT WAS REPORTED THAT AFTER THE INTERBODY CAGE WAS PLACED IN ITS FINAL POSITION, THE SURGEON ATTEMPTED TO ROTATE THE CAGE/INTERBODY, BUT THE RED PART OF THE HANDLE ROTATED INSTEAD OF THE CAGE/INTERBODY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629498 ↗
  • MalfunctionNov 24, 2025Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE RED HANDLE OF THE INSERTER ROTATED INSTEAD OF THE CAGE. THE SURGEON HAD SELECTED THE SIZE CAGE TO BE USED, AND UPON ATTEMPTING TO ROTATE IT, ONLY THE HANDLE MOVED WHILE THE CAGE REMAINED STATIONARY. THE CAGE WAS SUCCESSFULLY EXTRACTED, AND AN ALTERNATE ROTATING CAGE FROM ANOTHER MANUFACTURER WAS USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629494 ↗
  • MalfunctionNov 24, 2025Mechanics Altered; Mechanics Altered

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR TRANSFORAMINAL LUMBAR INTERBODY FUSION. IT WAS REPORTED THAT THE INSTRUMENT HANDLES WERE TWISTING DURING ROUTINE CHECKING IN THE STERILE PROCESSING DEPARTMENT, AND THE INSTRUMENT BROKE. THERE WAS NO PATIENT INVOLVED. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629447 ↗
  • MalfunctionNov 24, 2025Material Twisted/Bent; Material Twisted/Bent

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR INTERBODY FUSION (PLIF) L3-4 FOR AN INDICATION OF DEGENERATIVE DISC DISEASE. IT WAS REPORTED THAT THE INSERTER HANDLE PHYSICALLY TWISTED WHILE ATTEMPTING TO ROTATE THE CAGE AND DID NOT ALLOW THE CAGE TO FULLY ROTATE, TWISTING 45 DEGREES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629420 ↗
  • MalfunctionNov 24, 2025Break; Mechanical Problem; Break; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3-S2 SPINAL FUSION PROCEDURE WITH INTERBODIES AT L3-S1. IT WAS REPORTED THAT THE INSERTER COULD NOT HANDLE THE REQUIRED TORQUE AND BROKE. PLIERS WERE USED ON THE TIGHTENING MECHANISM TO COMPLETE THE IMPLANT ROTATION WITHOUT THE HANDLE LOOSENING. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629417 ↗
  • MalfunctionNov 24, 2025Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING THE DEVICES USED FOR TLIF PROCEDURE. IT WAS REPORTED THAT THE HANDLE OF THE INSTRUMENTS WAS BENT AND STRIPPED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629415 ↗
  • MalfunctionNov 24, 2025Break; Mechanical Problem; Break; Mechanical Problem

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3-S2 SPINAL FUSION PROCEDURE WITH INTERBODIES AT L3-S1. IT WAS REPORTED THAT THE INSERTER COULD NOT HANDLE THE REQUIRED TORQUE AND BROKE. PLIERS WERE USED ON THE TIGHTENING MECHANISM TO COMPLETE THE IMPLANT ROTATION WITHOUT THE HANDLE LOOSENING. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629403 ↗
  • MalfunctionNov 24, 2025Material Twisted/Bent; Material Twisted/Bent

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR INTERBODY FUSION (PLIF) L3-4 FOR AN INDICATION OF DEGENERATIVE DISC DISEASE. IT WAS REPORTED THAT THE INSERTER HANDLE PHYSICALLY TWISTED WHILE ATTEMPTING TO ROTATE THE CAGE AND DID NOT ALLOW THE CAGE TO FULLY ROTATE, TWISTING 45 DEGREES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23629399 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K241164. One submission often covers many models, so check the product and lot details in each record.

  • Z-1039-2026 · Oct 29, 2025 · Open, Classified
    Medicrea, IB3D PL Instruments Set, Rx Only REF: SPS03174

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →