Skip to content

screwdriver 264683Recalls & reported problems

Fournitures Hospitalieres Industrie · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03661489646836), received between Dec 14, 2018 and Dec 14, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Accessory Incompatible” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Accessory Incompatible1
  • Compatibility Problem1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 14, 2018Accessory Incompatible; Compatibility Problem; Accessory Incompatible; Compatibility Problem

    WHILE DOING A REVISION RSA USING CUSTOM IMPLANTS AND V12 INSTRUMENT SET, WE WERE NOT ABLE TO GET THE GLENOSPHERE SCREW TO FULLY ENGAGE WITH METAL BACK GLENOID (SIZE 46 LP). GLENOSPHERE AND FH METAL BACK GLENOID HAD TO BE REMOVED AND REPLACED WITH DJO ENCORE BASEPLATE AND GLENOSPHERE. FH STEM STAYED IN PLACE AND CUSTOM HUMERAL HEAD (B)(4) WAS IMPLANTED ((B)(4)) AFTER THE CASE IT WAS DISCOVERED THAT DUE TO V12 SCREWDRIVER'S TAPERED DESIGN, IT IS IMPOSSIBLE FOR THE SCREWDRIVER TO FULLY GO DOWN SCREW SHAFT ON 44 IMPLANT. NO CONSEQUENCE FOR THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8163616 ↗

Other Shoulder Prosthesis, Reverse Configuration devices

All Shoulder Prosthesis, Reverse Configuration devices →

Sources and updates