BI-MENTUM PE Liner BI-MENTUM PE Liner 22,2 / 43Recalls & reported problems

Soc Etudes Recherches Fabrication · BI-MENTUM dual mobility system -BI-MENTUM PE Liner polyethylene liner · Class 2 · Implant

All models of BI-MENTUM PE Liner (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This BI-MENTUM PE Liner is the polyethylene inner liner used in the BI-MENTUM dual mobility hip system. Most of the 31 reports describe the hip joint dislocating or coming apart after surgery, with several patients needing revision operations to fix it; others mention pain, fluid buildup around the hip, and infection. All31 reports describe injuries, with no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
31
Deaths reported
0
Injuries reported
31
Recalls naming it
0

Summary

The FDA's MAUDE database contains 31 adverse event reports that list this exact device (DI 03662200015504), received between Mar 31, 2021 and Mar 10, 2026 — classified as 31 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 6 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 3 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem18
  • Device Dislodged or Dislocated10
  • Appropriate Device Problem Term/Code Not Available3
  • Insufficient Device Problem Information3

Patient problems

  • Joint Dislocation19
  • Unspecified Infection6
  • Pain4
  • Ambulation or Postural Difficulties3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Colitis1
  • Cyst(s)1
  • Distress1

Recent adverse event reports

The 25 most recent of 31 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24558274 ↗
  • InjuryMar 4, 2026Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATION. DOI: (B)(6) 2025 . DOR: (B)(6) 2026. AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 24511329 ↗
  • InjuryJun 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    UPDATED EVENT DESCRIPTION: MEDICAL RECORDS RECEIVED. THE PATIENT PREVIOUSLY UNDERWENT A PRIMARY HIP REPLACEMENT THAT USED A METAL-ON-METAL IMPLANT. OVER TIME, FRICTION BETWEEN THE METAL COMPONENTS CAUSED THE RELEASE OF MICROSCOPIC METAL DEBRIS INTO THE SURROUNDING TISSUES. THIS LED TO THE DEVELOPMENT OF METALLOSIS, AN INFLAMMATORY REACTION TO METAL PARTICLES. THE RESULTING INFLAMMATION PRODUCED A MASS IN THE HIP REGION THAT APPEARED SIMILAR TO A TUMOR ON IMAGING STUDIES, PROMPTING AN EVALUATION FOR POSSIBLE CANCER. FURTHER INVESTIGATION REVEALED THAT THE MASS WAS ACTUALLY A PSEUDOTUMOR¿AN INFLAMMATORY, NON-CANCEROUS GROWTH CAUSED BY THE METAL DEBRIS FROM THE IMPLANT. ADDITIONAL DIAGNOSES AND CONDITION EXPERIENCED FOUND, INFLAMED AND IRRITATED JOINT LINING, WEAR-AND-TEAR CHANGES, EVIDENCE OF OLD JOINT BLEEDING, SOME BONE AREAS WITH CELL DEATH (OSTEONECROSIS), A REMOVED PROSTHETIC JOINT COMPONENT, VESICULAR PALMOPLANTAR ECZEMA, IMPETIGO, SEBACEOUS CYST, TROCHANTERIC BURSITIS, COLITIS, MECHANICAL COMPLICATION OF PROSTHETIC JOINT IMPLANT, AND MECHANICAL COMPLICATION OF PROSTHETIC HIP IMPLANT. MRI ON (B)(6) 2024 SHOWS FINDINGS MOST SUGGESTIVE OF LEFT HIP SYNOVITIS, WITH A FLUID COLLECTI

    Patient problems: Colitis; Cyst(s); Distress; Foreign Body Reaction; Hemorrhage/Blood Loss/Bleeding; Hypersensitivity/Allergic reaction; Inflammation; Necrosis; Pain; Skin Inflammation/ Irritation; Swelling/ Edema; Synovitis

    MAUDE report 22330813 ↗
  • InjuryJun 11, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    PATIENT OCCURED CHRONIC DISLOCATIONS LEADING TO REVISION OF THE IMPLANTS.

    Patient problems: Joint Dislocation

    MAUDE report 22184496 ↗
  • InjuryJun 4, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO CHRONIC DISLOCATIONS. DOI: (B)(6) 2020, DOR: (B)(6) 2025, AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 22143487 ↗
  • InjuryJan 24, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    (B)(4): SERF RC-23-0092 PATIENT WAS REVISED DUE TO PAIN, SERROMA. THERE WAS NO LOOSENING REPORTED.

    Patient problems: Insufficient Information; Pain

    MAUDE report 21230149 ↗
  • InjuryJan 23, 2025Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information

    (B)(4): SERF 02-2023 PATIENT WAS REVISED DUE DISLOCATING. NO LOOSENING AND SURGICAL DELAY WAS THERE.

    Patient problems: Ambulation or Postural Difficulties; Joint Dislocation

    MAUDE report 21226098 ↗
  • InjuryJan 23, 2025Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information

    (B)(4): (B)(6) - PATIENT WAS REVISED DUE TO DISLOCATION. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY.

    Patient problems: Ambulation or Postural Difficulties; Joint Dislocation

    MAUDE report 21225974 ↗
  • InjuryJan 23, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    (B)(4): SERF RC-23-0112, PATIENT WAS REVISED DUE TO DISLOCATION WITH DM WENT BACK IN WITH DIFFERENT LINER AND LONGER HEAD BALL.

    Patient problems: Ambulation or Postural Difficulties; Joint Dislocation

    MAUDE report 21225775 ↗
  • InjuryJan 8, 2025Device Dislodged or Dislocated; Insufficient Device Problem Information; Device Dislodged or Dislocated; Insufficient Device Problem Information

    PATIENT WAS CONTINUING TO DISLOCATE. THE LINER WAS CONVERTED FROM A DUAL MOBILITY TO A CONSTRAINED LINER. RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21100501 ↗
  • InjuryJan 6, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT WAS REVISED DUE TO DISLOCATION. THE LINER WAS CONVERTED FROM A DUAL MOBILITY TO A CONSTRAINED LINER. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 21079596 ↗
  • InjuryAug 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO DISLOCATION WITH DM WENT BACK IN WITH DIFFERENT LINER AND LONGER HEAD BALL. DOI: UNKNOWN, DOR: (B)(6), 2023, AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 17479223 ↗
  • InjuryJul 6, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO PAIN AND SEROMA. THERE WAS NO LOOSENING REPORTED. DOI: (B)(6) 2023. DOR: (B)(6) 2023. AFFECTED SIDE: UNKNOWN.

    Patient problems: Pain; Seroma

    MAUDE report 17267113 ↗
  • InjuryJan 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO DISLOCATION. THERE WAS NO REPORTED LOOSENING. THERE WAS NO REPORTED SURGICAL DELAY. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 16149945 ↗
  • InjuryJan 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT SEEMS TO BE INFECTED. WASHED OUT HIP AND REPLACED DUAL MOBILITY HEAD WITH SAME SIZE. THERE WAS NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 16137423 ↗
  • InjuryJan 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE DISLOCATING. NO LOOSENING AND SURGICAL DELAY WAS THERE. DOI: UNKNOWN. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 16137047 ↗
  • InjuryDec 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PATIENT HAD I&D OF HIP. LINER AND HEAD EXCHANGED. INFECTION WAS CONFIRMED. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 15979193 ↗
  • InjuryOct 12, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CLINICAL ADVERSE EVENT RECEIVED FOR DISLOCATION. EVENT IS SERIOUS AND IS CONSIDERED MILD. EVENT IS POSSIBLY RELATED TO BOTH DEVICE AND PROCEDURE. DATE OF IMPLANT: (B)(6) 2016. DATE OF EVENT: (B)(6) 2022. (LEFT HIP). TREATMENT: CLOSED REDUCTION/MANIPULATION.

    Patient problems: Joint Dislocation

    MAUDE report 15587738 ↗
  • InjuryAug 8, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS REVISED DUE TO INFECTION. NO SURGICAL DELAY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Unspecified Infection

    MAUDE report 15182583 ↗
  • InjuryJun 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO DISLOCATION. THE SURGEON REMOVED CUP, HEAD, AND LINER AND REPLACED WITH STRYKER DUAL MOBILITY. NO SURGICAL DELAY WAS REPORTED. DOI: (B)(6) 2022; DOR: (B)(6) 2022; AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 14685573 ↗
  • InjuryJun 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    PATIENT WAS REVISED DUE TO DISLOCATION OF IMPLANTS. THERE WAS NO LOOSENING AND NO SURGICAL DELAY. DOI: (B)(6) 2021; DOR: (B)(6) 2022; AFFECTED SIDE: UNKNOWN HIP.

    Patient problems: Joint Dislocation

    MAUDE report 14684792 ↗
  • InjuryMay 12, 2022Appropriate Device Problem Term/Code Not Available

    PATIENT WAS REVISED DUE TO PAIN. DOI: UNKNOWN, DOR: (B)(6) 2022, AFFECTED SIDE: LEFT HIP.

    Patient problems: Pain

    MAUDE report 14372587 ↗
  • InjuryMay 11, 2022Device Dislodged or Dislocated

    PATIENT WAS REVISED DUE TO DISLOCATED. THERE WAS NO DELAY IN SURGERY. DOI: (B)(6) 2022. DOR: (B)(6) 2022. AFFECTED SIDE: RIGHT HIP.

    Patient problems: Joint Dislocation

    MAUDE report 14353123 ↗
  • InjuryApr 6, 2022Appropriate Device Problem Term/Code Not Available

    IT WAS REVISED TO ADDRESS WASH OUT (INFECTED). DOI- UNKNOWN, DOR- (B)(6) 2022, AFFECTED SIDE- LEFT HIP.

    Patient problems: Unspecified Infection

    MAUDE report 14021658 ↗
  • InjuryJan 28, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    REVISION COMPLETED DUE TO JOINT INSTABILITY. THE MDM CONSTRUCT WAS REVISED WITH PINNACLE CONSTRAINED. DOI: UNKNOWN. DOR: (B)(6) 2022, AFFECTED SIDE: LEFT HIP.

    Patient problems: Joint Laxity

    MAUDE report 13388731 ↗

Other Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices

All Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented devices →