ROI-C MC1313PRecalls & reported problems
LDR Spine Usa, Inc. · ROI-C IMPLANT 14X14 H7 · Class 2
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03662663016087), received between Oct 19, 2023 and Oct 19, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2023Fracture; Fracture
IT WAS REPORTED THAT A SURGEON HAD MULTIPLE ROI-C CAGES THAT FRACTURED OR CRACKED INTRA-OPERATIVELY. TWO CAGES FRACTURED AT LEVEL C7/T1 AFTER INSERTING THE ANCHORING PLATES. THEY WERE EACH REMOVED AND REPLACED WITH A NEW CAGE. THE THIRD CAGE CRACKED AND THE SURGEON ELECTED TO LEAVE IT IN PLACE. THE SURGEON PROCEEDED TO THE C6/7 LEVEL AND INSTALLED THE FIRST CAGE, BUT IT WAS FOUND TO HAVE FRACTURED AFTER THE FIRST ANCHORING PLATE WAS INSTALLED. THE CAGE WAS AGAIN REMOVED AND REPLACED WITH A NEW CAGE AND THE SURGEON ELECTED NOT TO INSTALL THE ANCHORING PLATES AT THAT LEVEL. THERE WAS A DELAY OF THREE HOURS TO THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS.THIS IS REPORT THREE OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967553 ↗MalfunctionOct 19, 2023Fracture; Fracture
IT WAS REPORTED THAT A SURGEON HAD MULTIPLE ROI-C CAGES THAT FRACTURED OR CRACKED INTRA-OPERATIVELY. TWO CAGES FRACTURED AT LEVEL C7/T1 AFTER INSERTING THE ANCHORING PLATES. THEY WERE EACH REMOVED AND REPLACED WITH A NEW CAGE. THE THIRD CAGE CRACKED AND THE SURGEON ELECTED TO LEAVE IT IN PLACE. THE SURGEON PROCEEDED TO THE C6/7 LEVEL AND INSTALLED THE FIRST CAGE, BUT IT WAS FOUND TO HAVE FRACTURED AFTER THE FIRST ANCHORING PLATE WAS INSTALLED. THE CAGE WAS AGAIN REMOVED AND REPLACED WITH A NEW CAGE AND THE SURGEON ELECTED NOT TO INSTALL THE ANCHORING PLATES AT THAT LEVEL. THERE WAS A DELAY OF THREE HOURS TO THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS. THIS IS REPORT TWO OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967538 ↗MalfunctionOct 19, 2023Fracture; Fracture
IT WAS REPORTED THAT A SURGEON HAD MULTIPLE ROI-C CAGES THAT FRACTURED OR CRACKED INTRA-OPERATIVELY. TWO CAGES FRACTURED AT LEVEL C7/T1 AFTER INSERTING THE ANCHORING PLATES. THEY WERE EACH REMOVED AND REPLACED WITH A NEW CAGE. THE THIRD CAGE CRACKED AND THE SURGEON ELECTED TO LEAVE IT IN PLACE. THE SURGEON PROCEEDED TO THE C6/7 LEVEL AND INSTALLED THE FIRST CAGE, BUT IT WAS FOUND TO HAVE FRACTURED AFTER THE FIRST ANCHORING PLATE WAS INSTALLED. THE CAGE WAS AGAIN REMOVED AND REPLACED WITH A NEW CAGE AND THE SURGEON ELECTED NOT TO INSTALL THE ANCHORING PLATES AT THAT LEVEL. THERE WAS A DELAY OF THREE HOURS TO THE PROCEDURE WITHOUT REPORTED PATIENT IMPACTS.THIS IS REPORT ONE OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967537 ↗
Other Intervertebral Fusion Device With Bone Graft, Cervical devices
- EIT CIF cage CUI8070SEIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Cervical1 recall4 reports
- ROI-C MC1005TLDR Spine Usa, Inc. · Intervertebral Fusion Device With Bone Graft, Cervical54 reports
- EIT CIF cage CUI8060LEIT Emerging Implant Technologies GmbH · Intervertebral Fusion Device With Bone Graft, Cervical1 recall1 report
- CONSTRUX - MINI Ti 37-7206SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-8007SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI PTC 37-4211SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI PTC 37-5211SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
- CONSTRUX - MINI Ti 37-8105SPOrthofix US LLC · Intervertebral fusion device with bone graft, cervical1 recall0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.