Skip to content

Zyston Interbody Spacer System 14-533127Recalls & reported problems

Highridge Medical, LLC · Curve Spacer 7H x 27L 6° · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03662663155205), received between Dec 22, 2025 and May 26, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2026Fracture

    IT WAS REPORTED THAT A PIECE FRACTURED OFF A CURVED PEEK SPACER INTRA-OPERATIVELY. AFTER THE IMPLANT WAS INSTALLED AND THE INSERTER WAS DETACHED, THE SURGEON FURTHER IMPACTED THE SPACER INTO THE DISC SPACE WHEN A PIECE OF THE SPACER BROKE OFF. SINCE THE SPACER NEEDED TO BE INSERTED FURTHER, THE SURGEON ELECTED TO REMOVE AND REPLACE IT WITH A SECOND SPACER. HOWEVER, DURING REMOVAL, THE METAL PIVOTING PIECE OF THE IMPLANT BROKE OUT OF THE SPACER. THE REMAINDER OF THE PEEK PORTION OF THE SPACER WAS JUDGED TO HAVE SUFFICIENT FIXATION AND REMAINS IN-SITU. THERE WAS A DELAY OF 20 MINUTES TO THE PROCEDURE, BUT THERE WERE NO REPORTED PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25284430 ↗
  • MalfunctionDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT WHILE INSERTING THE ZYSTON CAGE AT T11/12 DURING A POSTEROLATERAL LUMBAR FUSION, THE PEEK CHIPPED. THE FRAGMENT WAS REMOVED BUT THE CAGE REMAINED IMPLANTED AND THE SURGERY WAS COMPLETED WITHOUT ANY PATIENT HARM OR DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23868296 ↗

Other Intervertebral fusion device with bone graft, lumbar devices

All Intervertebral fusion device with bone graft, lumbar devices →

Sources and updates