Skip to content

InCompass Spinal Fixation System 2155-4Recalls & reported problems

Highridge Medical, LLC · Universal Driver - Modular · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03662663192361), received between Dec 22, 2025 and Dec 22, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIPS OF THREE SCREW DRIVERS (ONE UNIVERSAL AND TWO UNIVERSAL WITH STANDARD CONNECTIONS) BROKE INTRA-OPERATIVELY WHILE ATTEMPTING TO REMOVE A SCREW THAT HAD BEEN IMPLANTED FOR 16 YEARS. THE SCREW WAS ABLE TO BE REMOVED USING AN ALTERNATIVE METHOD SINCE THE DRIVER ENGAGEMENT FEATURE HAD STRIPPED. THERE WERE NO REPORTED IMPACTS TO THE PATIENT. THIS IS REPORT THREE OF THREE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871747 ↗
  • MalfunctionDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIPS OF THREE SCREW DRIVERS (ONE UNIVERSAL AND TWO UNIVERSAL WITH STANDARD CONNECTIONS) BROKE INTRA-OPERATIVELY WHILE ATTEMPTING TO REMOVE A SCREW THAT HAD BEEN IMPLANTED FOR 16 YEARS. THE SCREW WAS ABLE TO BE REMOVED USING AN ALTERNATIVE METHOD SINCE THE DRIVER ENGAGEMENT FEATURE HAD STRIPPED. THERE WERE NO REPORTED IMPACTS TO THE PATIENT. THIS IS REPORT TWO OF THREE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871677 ↗

Other SCREWDRIVER devices

All SCREWDRIVER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page