Skip to content

Axiom OP52100Recalls & reported problems

Etablissements Anthogyr · AXIOM IMPLANT Ø5.2 x 10.0 Replacement of a missing tooth for placement of a dental restoration · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Anthogyr Axiom dental implant is a root-form implant used to replace a missing tooth. Most of the 849 reports describe implants that did not bond with the bone or later lost that bond, and some mention pain, infection and inflammation. The reports record 782 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
860
Deaths reported
0
Injuries reported
793
Recalls naming it
0

Summary

The FDA's MAUDE database contains 860 adverse event reports that list this exact device (DI 03663394000253), received between Dec 21, 2016 and Sep 28, 2026 — classified as 793 injuries, 67 malfunctions.

The most frequently reported specific device problem is “Failure to Osseointegrate” (693 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 17 reportsDec 2023: 19 reportsJan 2024: 8 reportsFeb 2024: 11 reportsMar 2024: 14 reportsApr 2024: 8 reportsMay 2024: 15 reportsJun 2024: 8 reportsJul 2024: 5 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 6 reportsJan 2025: 2 reportsFeb 2025: 9 reportsMar 2025: 8 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 39 reportsJul 2025: 8 reportsAug 2025: 9 reportsSep 2025: 6 reportsOct 2025: 7 reportsNov 2025: 10 reportsDec 2025: 15 reportsJan 2026: 15 reportsFeb 2026: 12 reportsMar 2026: 14 reportsApr 2026: 7 reportsMay 2026: 9 reportsJun 2026: 6 reportsJul 2026: 4 reportsAug 2026: 12 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate693
  • Loss of Osseointegration153
  • Adverse Event Without Identified Device or Use Problem12
  • Fracture8
  • Separation Failure3
  • Difficult to Insert2

Patient problems

  • No Clinical Signs, Symptoms or Conditions566
  • Inadequate Osseointegration150
  • Fibrosis79
  • Unspecified Infection33
  • Pain31
  • Inflammation28
  • Swelling/ Edema24
  • Hemorrhage/Blood Loss/Bleeding17

Recent adverse event reports

The 25 most recent of 860 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919181 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919023 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919003 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918945 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50824665 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted on (b)(6) 2025 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis and Pain. No further patient complications were reported.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 50824642 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758241 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (B)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758238 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758183 ↗
  • InjurySep 14, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50758178 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414007 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414006 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414005 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414004 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in the patient's mouth. Details of surgery: Primary stability not achieved. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414003 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    Patient problems: Fibrosis; Pain

    MAUDE report 50413986 ↗
  • InjuryAug 24, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2017 in FDI 47. On (b)(6) 2026, loss of osseointegration was verified. No product was returned to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50413960 ↗
  • InjuryAug 17, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 28. Details of surgery: Implant surface not completely covered with bone. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Inflammation. No further patient complications were reported.

    Patient problems: Inflammation

    MAUDE report 50281688 ↗
  • InjuryAug 17, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 16. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50281644 ↗
  • InjuryAug 6, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162296 ↗
  • InjuryAug 6, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted 2025(b)(6) in FDI 17. On 2026(b)(6) non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50162286 ↗
  • InjuryAug 3, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (B)(6) 2026 in FDI 17. On (B)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50118585 ↗
  • InjuryJul 30, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in FDI 16. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50087761 ↗
  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 36. On (b)(6) 2026, non-osseointegration was verified. Patient presented with fair oral hygiene and bruxism. The device was forwarded to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Unspecified Infection

    MAUDE report 50040062 ↗
  • InjuryJul 6, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 46. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25720722 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page