Axiom PX46080Recalls & reported problems
Etablissements Anthogyr · Axiom implant PX Ø4.6 x 8.0, Replacement of a missing tooth for placement of a dental restoration, not usable in the molar area, not recommended in D1 bone · Class 2 · Implant
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Axiom dental implant is a root-form implant placed in the jawbone to hold a replacement tooth. Its 3,192 reports mostly describe the implant failing to fuse with the jawbone, with patients reporting scar tissue, swelling and pain. The reports include no deaths and 3,158 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 3,280
- Deaths reported
- 0
- Injuries reported
- 3,246
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3,280 adverse event reports that list this exact device (DI 03663394001885), received between Jul 20, 2017 and Sep 30, 2026 — classified as 3,246 injuries, 34 malfunctions.
The most frequently reported specific device problem is “Failure to Osseointegrate” (2,803 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Osseointegrate2,803
- Loss of Osseointegration430
- Adverse Event Without Identified Device or Use Problem40
- Fracture26
- Difficult to Insert3
- Use of Device Problem2
- Separation Failure1
Patient problems
- No Clinical Signs, Symptoms or Conditions2,309
- Fibrosis467
- Inadequate Osseointegration257
- Pain240
- Swelling/ Edema160
- Unspecified Infection146
- Inflammation132
- Hemorrhage/Blood Loss/Bleeding51
Recent adverse event reports
The 25 most recent of 3,280 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50950198 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 16. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919258 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in FDI 15. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919225 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 45. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On 2026(b)(6), non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919216 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in FDI 27. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6)2026, non-osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919215 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 17. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919188 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919183 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2026 in FDI 16. On (B)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919182 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2025(b)(6) in FDI 47. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919166 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2025 in FDI 26. On (b)(6)2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919164 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2025 in FDI 26. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919154 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2025 in FDI 17. On (B)(6) 2026 non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919151 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 12. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919137 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2025(b)(6) in FDI 37. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919127 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6) 2026 in FDI 12. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919113 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 15. On 2026-(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919105 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2025(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919069 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (B)(6) 2026 in FDI 26. On (B)(6) 2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919043 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2025(b)(6)in FDI 27. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation; Pain
MAUDE report 50919041 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 47. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation; Pain
MAUDE report 50919040 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted (b)(6)2026 in FDI 37. On (b)(6)2026, non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Pain and Inflammation. No further patient complications were reported.
Patient problems: Inflammation; Pain
MAUDE report 50919039 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6) non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation
MAUDE report 50919033 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2025(b)(6) in FDI 26. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type IV and fair oral hygiene. The device was forwarded to the manufacturer. At the event the patient experienced: Mobility and Inflammation. No further patient complications were reported.
Patient problems: Fibrosis; Inflammation
MAUDE report 50919026 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 16. On 2026(b)(6), non-osseointegration was verified. Patient presented with bone type III and fair oral hygiene. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50919022 ↗InjurySep 28, 2026Failure to Osseointegrate
The clinician reports the implant was inserted 2026(b)(6) in FDI 27. On 2026(b)(6), non-osseointegration was verified. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50918970 ↗
FDA clearance and classification
The FDA regulates product code DZE (implant, endosseous, root-form) under 21 CFR 872.3640 as a Class II device. 21 CFR 872.3640 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Axiom PX46080
Has the Axiom PX46080 been recalled?
No FDA recall record in our data names this exact device identifier (DI 03663394001885), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Axiom PX46080?
The FDA's MAUDE database lists 3,280 adverse event reports for this device, received between Jul 20, 2017 and Sep 30, 2026. In those reports, the most common specific device problems are “Failure to Osseointegrate” (2,803), “Loss of Osseointegration” (430) and “Fracture” (26); the most common patient problems are “Fibrosis” (467), “Inadequate Osseointegration” (257) and “Pain” (240). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Axiom PX46080?
None of the reports are classified as deaths, 3,246 as injuries and 34 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Axiom PX46080 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the Axiom PX46080, and is it still sold?
The FDA's device database lists Etablissements Anthogyr as the company responsible for it (brand Axiom, model PX46080). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Implant, Endosseous, Root-Form devices
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM43,942 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM40,651 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM27,576 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,756 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLA Titanium BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM25,548 reports
- Straumann® Dental Implant System Roxolid® SLA Impl. SLA Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM23,631 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,501 reports
- STRAUMANN BONE LEVEL TAPERED IMPLANT SLActive Roxolid BLInstitut Straumann AG · IMPLANT, ENDOSSEOUS, ROOT-FORM22,423 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record