Skip to content

Axiom PX34080Recalls & reported problems

Etablissements Anthogyr · Axiom implant PX Ø3.4 x 8.0, Replacement of a missing tooth for placement of a dental restoration, not usable in the molar area, not recommended in D1 bone. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Anthogyr Axiom dental implant has 1,433 FDA adverse event reports, with 1,392 injuries, 41 malfunctions and no deaths. The reports mostly describe the implant not bonding with the bone or losing that bond, and some mention gum and bone inflammation around the implant, pain and infection. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,458
Deaths reported
0
Injuries reported
1,417
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,458 adverse event reports that list this exact device (DI 03663394002509), received between Aug 3, 2017 and Sep 28, 2026 — classified as 1,417 injuries, 41 malfunctions.

The most frequently reported specific device problem is “Failure to Osseointegrate” (1,115 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 17 reportsDec 2023: 32 reportsJan 2024: 16 reportsFeb 2024: 15 reportsMar 2024: 4 reportsApr 2024: 14 reportsMay 2024: 20 reportsJun 2024: 8 reportsJul 2024: 16 reportsAug 2024: 11 reportsSep 2024: 16 reportsOct 2024: 18 reportsNov 2024: 17 reportsDec 2024: 12 reportsJan 2025: 18 reportsFeb 2025: 6 reportsMar 2025: 33 reportsApr 2025: 5 reportsMay 2025: 0 reportsJun 2025: 78 reportsJul 2025: 14 reportsAug 2025: 27 reportsSep 2025: 16 reportsOct 2025: 21 reportsNov 2025: 32 reportsDec 2025: 23 reportsJan 2026: 17 reportsFeb 2026: 24 reportsMar 2026: 25 reportsApr 2026: 41 reportsMay 2026: 22 reportsJun 2026: 27 reportsJul 2026: 30 reportsAug 2026: 27 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202678
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate1,115
  • Loss of Osseointegration315
  • Adverse Event Without Identified Device or Use Problem24
  • Fracture3
  • Difficult to Insert2
  • Separation Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions846
  • Fibrosis270
  • Inadequate Osseointegration247
  • Pain127
  • Unspecified Infection79
  • Inflammation63
  • Swelling/ Edema56
  • Hemorrhage/Blood Loss/Bleeding34

Recent adverse event reports

The 25 most recent of 1,458 reports, in the reporter's own words as published by the FDA.

  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919239 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919209 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919192 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919118 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50919076 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918999 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918995 ↗
  • InjurySep 28, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918978 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918923 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918922 ↗
  • InjurySep 28, 2026Loss of Osseointegration

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50918920 ↗
  • InjurySep 18, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824597 ↗
  • InjurySep 11, 2026Loss of Osseointegration

    Patient problems: Fibrosis

    MAUDE report 50737777 ↗
  • InjurySep 11, 2026Failure to Osseointegrate

    Patient problems: Numbness; Unspecified Infection

    MAUDE report 50737758 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50653309 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50653307 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Fibrosis; Pain

    MAUDE report 50653304 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: Pain

    MAUDE report 50653298 ↗
  • InjurySep 7, 2026Failure to Osseointegrate

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50653260 ↗
  • InjuryAug 28, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2025 in FDI 22. On (b)(6) 2026, non-osseointegration was verified. Patient presented with bone type III. No product was returned to the manufacturer. At the event the patient experienced: Peri-implantitis, Mobility and Numbness. No further patient complications were reported.

    Patient problems: Fibrosis; Numbness; Unspecified Infection

    MAUDE report 50503852 ↗
  • InjuryAug 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6)2026 in the patient's mouth. Details of surgery: Implant surface not completely covered with bone and Ridge augmentation. On (b)(6)2026, non-osseointegration was verified. Patient presented with bone type III. The device was forwarded to the manufacturer. There were no reported patient injuries or complications.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480463 ↗
  • InjuryAug 27, 2026Loss of Osseointegration

    The clinician reports the implant was inserted (b)(6) 2024 in FDI 35. Details of surgery: Primary stability not achieved and Implant surface not completely covered with bone. On (b)(6) 2026, loss of osseointegration was verified. Patient presented with fair oral hygiene. The device was forwarded to the manufacturer. There were no patient operative or post-operative complications reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50480451 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50414088 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50414087 ↗
  • InjuryAug 24, 2026Failure to Osseointegrate

    Patient problems: Fibrosis

    MAUDE report 50414086 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →

Sources and updates