Axiom PX40180Recalls & reported problems

Etablissements Anthogyr · Axiom implant PX Ø4.0 x 18.0, Replacement of a missing tooth for placement of a dental restoration, not usable in the molar area, not recommended in D1 bone · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
19
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 03663394002615), received between Jan 5, 2018 and Jul 27, 2026 — classified as 18 injuries, 1 malfunction.

The most frequently reported specific device problem is “Failure to Osseointegrate” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Osseointegrate13
  • Loss of Osseointegration5
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8
  • Inadequate Osseointegration7
  • Inflammation3
  • Pain2
  • Unspecified Infection2
  • Fibrosis1

Recent adverse event reports

The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 27, 2026Failure to Osseointegrate

    The clinician reports the implant was inserted (b)(6) 2026 in FDI 35. On (b)(6) 2026, non-osseointegration was verified. The device was forwarded to the manufacturer. At the event the patient experienced: Pain, Mobility and Inflammation. No further patient complications were reported.

    Patient problems: Fibrosis; Inflammation; Pain

    MAUDE report 50022850 ↗
  • InjuryApr 20, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 25. ON (B)(6) 2026, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24928619 ↗
  • InjuryJan 29, 2026Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN FDI 17. ON (B)(6) 2025, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24212154 ↗
  • InjuryJul 31, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN FDI 47. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17426567 ↗
  • InjuryApr 17, 2023Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2023 IN FDI 25. ON (B)(6) 2023, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16754027 ↗
  • InjuryNov 28, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2022 IN THE PATIENT'S MOUTH. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15871255 ↗
  • InjuryMay 30, 2022Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 23. DETAILS OF SURGERY: IMPLANT SURFACE NOT COMPLETELY COVERED WITH BONE. ON (B)(6) 2022, NON-OSSEOINTEGRATION WAS VERIFIED. PATIENT PRESENTED WITH BONE TYPE III AND FAIR ORAL HYGIENE. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation; Unspecified Infection

    MAUDE report 14541993 ↗
  • InjuryApr 25, 2022Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 13. ON (B)(6) 2022, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFLAMMATION. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Inflammation

    MAUDE report 14195016 ↗
  • InjuryJul 19, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE CLINICIAN REPORTS THAT THE DAY THE IMPLANT WAS PLACED IN THE PATIENT'S MOUTH, FAILURE OCCURRED UPON INSERTION. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12190351 ↗
  • InjuryJul 19, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2020 IN FDI 23. ON (B)(6) 2021, NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. AT THE EVENT THE PATIENT EXPERIENCED: INFECTION AND PAIN. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: Pain; Unspecified Infection

    MAUDE report 12190331 ↗
  • InjuryMay 31, 2021Loss of Osseointegration; Loss of Osseointegration

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2015 IN FDI 15. ON (B)(6) 2021, LOSS OF OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11908067 ↗
  • InjuryApr 16, 2021Failure to Osseointegrate; Failure to Osseointegrate

    THE CLINICIAN REPORTS THE IMPLANT WAS INSERTED (B)(6) 2021 IN FDI 13. ON (B)(6) 2021 , NON-OSSEOINTEGRATION WAS VERIFIED. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. THERE WERE NO REPORTED PATIENT INJURIES OR COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11679857 ↗
  • InjuryFeb 26, 2021Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILS TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 11383472 ↗
  • InjuryOct 2, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILS TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 10618543 ↗
  • InjuryJan 31, 2020Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILS TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 9653275 ↗
  • InjurySep 4, 2019Failure to Osseointegrate; Failure to Osseointegrate

    IMPLANT FAILS TO OSSEOINTEGRATE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8959343 ↗
  • InjuryJan 23, 2019Loss of Osseointegration; Loss of Osseointegration

    IMPLANT IS LOST (LOSS OF OSTEOINTEGRATION) AFTER IMPLANT HAS BEEN RESTORED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8269866 ↗
  • InjuryJan 23, 2019Loss of Osseointegration; Loss of Osseointegration

    IMPLANT IS LOST (LOSS OF OSSEOINTEGRATION) AFTER IMPLANT HAS BEEN RESTORED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 8269864 ↗
  • MalfunctionJan 5, 2018Loss of Osseointegration; Loss of Osseointegration

    IMPLANT IS LOST (LOSS OF OSSEOINTEGRATION) AFTER IMPLANT HAS BEEN RESTORED.

    Patient problems: Inadequate Osseointegration

    MAUDE report 7167802 ↗

Other Implant, Endosseous, Root-Form devices

All Implant, Endosseous, Root-Form devices →