Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03700386955099), received between Sep 17, 2019 and Mar 18, 2025 — classified as 4 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated4
  • Unstable2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • Failure of Implant2
  • Joint Dislocation2
  • Joint Laxity1
  • Loss of Range of Motion1
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 18, 2025Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem; Device Dislodged or Dislocated; Adverse Event Without Identified Device or Use Problem

    PATIENT HAS CONTINUED TO EXPERIENCE LAXITY, OCCASIONAL DISLOCATIONS OF HER AFFECTED SHOULDER JOINT, INCLUDING PAIN, POOR RANGE OF MOTION. SURGEON DECIDED DURING SURGERY TO AGAIN INCREASE GLENOSPHERE DIAMETER AND INCREASE HUMERAL REVERSE TRAY, AND AGAIN IMPLANT A RETENTIVE POLY INSERT. PROSTHESIS IMPLANTED, SHOULDER JOINT REDUCED AND FOUND TO BE STABLE THROUGH RANGE OF MOTION.

    Patient problems: Joint Dislocation; Joint Laxity; Loss of Range of Motion; Pain

    MAUDE report 21633275 ↗
  • InjuryMay 3, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REVERSE SHOULDER REPLACEMENT ON (B)(6) 2021. RECENTLY, THE PATIENT EXPERIENCED INSTABILITY OF THIS PROSTHETIC JOINT. TWICE DISLOCATED IN RECENT FORTNIGHT. SURGEON EXAMINED SHOULDER / MANIPULATION UNDER ANESTHETIC - DEMONSTRATED INSTABILITY. RADIOGRAPHICALLY, THE PROSTHESIS APPEARED WELL POSITIONED.

    Patient problems: Failure of Implant

    MAUDE report 14271699 ↗
  • InjuryMay 3, 2022Unstable; Device Dislodged or Dislocated; Unstable; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE PATIENT UNDERWENT REVERSE SHOULDER REPLACEMENT ON (B)(6) 2021. RECENTLY, THE PATIENT EXPERIENCED INSTABILITY OF THIS PROSTHETIC JOINT. SURGEON EXAMINED SHOULDER / MANIPULATION UNDER ANESTHETIC - DEMONSTRATED INSTABILITY. RADIOGRAPHICALLY, THE PROSTHESIS APPEARED WELL POSITIONED. THE PHYSICIAN THEN OPENED THE SHOULDER - SURGEON REMOVED POLY (6MM REVERSED INSERT, DWF391C), 0MM REVERSED TRAY (DWF510) AND 39MM LAT GLENOSPHERE (DWH905). IMPLANTED A 42MM GLENOSPHERE, +6MM REVERSED TRAY AND AN INSERT.

    Patient problems: Failure of Implant

    MAUDE report 14270259 ↗
  • InjurySep 17, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated

    THE PATIENT UNDERWENT INITIAL RSA ON (B)(6) 2017. THE IMPLANTS USED WERE ASCEND FLEX STD PTC HUMERAL STEM (DWF603B), REVERSE INSERT (DWF362B), REVERSE TRAY (DWF520), REVERSED II DIA. 25 PRESS FIT POST BASE PLATE (DWD170) AND REV. II GLENOID SPHERE DIA. 25 BASE PLATE (DWD180). ON (B)(6) 2017, A DISLOCATION OCCURRED AND REVISION SURGERY WAS PERFORMED. DURING THIS SURGERY, TRAY (DWF520), INSERT (DWF391B), AND SPHERE (DWH905) WERE NEWLY USED. IN 2019 (NO DATE), DISLOCATION OCCURRED AGAIN AND MANUAL REDUCTION WAS PERFORMED, BUT A REVISION SURGERY WAS TO BE PERFORMED AGAIN. ON (B)(6) 2019, REVISION SURGERY WAS PERFORMED: ASCEND FLEX: LONG PTC HUMERAL STEM 4B 132.5 ° 104MM (DWF614B), REVERSED INSERT 42MM (+) 9 B-12.5 (DWF425B), REV. II GLENOID SPHERE DIA. 25 BASEPLATE DIA. 42MM 10 TILTED SPHERE (DWD184), ASCEND FLEX REVERSED TRAY (+) 0 1.5MM (DWF510) WERE NEWLY REPLACED.

    Patient problems: Joint Dislocation

    MAUDE report 9079529 ↗